Probe for EPMA Online Manual


Copyright Dan Kremser, HTML Code Paul Carpenter


Procedures for Routine Tasks on WU JXA-8200 Electron Microprobe


Preparing for an EPMA Run


Outline: Setting Up a Probe for EPMA Run

Generic Microprobe Element Setup: WU Specific Information

  1. Generic Microprobe Setup
  2. Element List and Spectrometer Assignment
  3. Standards for Generic Run

Microprobe Instrument Prerequisites

PFE Run Prerequisites

These tasks should be completed before starting a PFE run, if possible.

If one or more standards to be used in your run have not been added to the standard database, this should be done first. The compositions are entered using standard.exe and the xyz position of the standard is set in the PFE program using the Automate window Digitize button.

  1. Standard database: Standard compositions entered
  2. Standard Positions: Positions digitized using Stage or PFE Automate

Probe for EPMA Setup Procedure

Notes:


Outline

  1. Program Startup
    1. Launch PFE, Connect to Hardware
  2. Open a new PFE run
    1. File--New
      Open New PFE Run Database File
  3. Add Standards to Run
    1. Standards--Add/Remove Standards
  4. Create New Sample template
    1. Acquire: New Sample Create and name new sample, Name: Initial
    2. Summary of Element Setup Methods
      1. Manual Element Entry -- This is the default where elements are added via Acquire: Elements/Cations
      2. Element Setup: How to save and load element settings on a per element basis
      3. Sample Setup: How to save and load Sample Setups (from within the curren run only)
      4. File Setup: How to browse to a pre-existing run and load a Sample Setup from that run
      5. Multiple File Setups: Load multiple Sample Setups from another run (same procedure as File Setup)
  5. Generate Master Element List and Enter Element Parameters
    1. Acquire: Elements/Cations Element parameter entry
    2. Background Measurement Type Off-peak vs. MAN backgrounds
    3. Peak and Background Positions Enter peak positions and background offsets
    4. Bias/Gain Enter bias/gain values
    5. Specified Elements Enter Specified Elements (e.g., oxygen)
    6. Background Interferences Screen potential background interferences
  6. Acquire: Analytical Conditions Set Analytical Conditions for sample
  7. Acquire: Peak/Scan Options Inspect and edit background offsets, wavescan limits, etc.
  8. Acquire: Count Times Inspect and edit count times
  9. Automate: Select Standards, Confirm Standard Positions, Run Confirm standard positions using Automate window
  10. Analyze: Standard Assignments Inspect and edit standard assignments
  11. Analyze: Standard Assignments Declare On-peak interferences via Standard Assignments window
  12. Automate: Select Standards, Peak Spectrometers, Peaking, Run Peak spectrometers using Automate window
    1. Manual: Automate Select Std, Go, Acquire: Peaking Options, Peaking Manual peaking of spectrometers, can include option to acquire bias/gain scan after each peak
    2. Acquire: PHA Conduct Bias/Gain/PHA scan on per element basis manually (if necessary)
  13. Run--Display and Export Scans Inspect peaking record to confirm peaking performed and good centroid acquired
  14. Acquire: Acquisition Options Set Acquisition Options for data acquisition
  15. Acquire: Special Options Set Acquisition Options for Run, Time Dependent Intensity (if beam sensitive materials analyzed)
  16. Analyze--Analytical Options Set Analysis Options and global on/off switches as necessary
  17. Automate Automated Acquisition of Standards(no. points, confirm, analytical mode, etc.)
  18. Automate: Select Standards, Acquire Standard Samples, Run Automated acquisition of Standard Samples
  19. Acquire: Acquire Standard Sample Manual acquisition of standard samples (Automate: select standard, Go, Acquire: New Sample (type standard), Acquire Standard Sample)
  20. Analyze: Select Standard, Analyze Inspect standardization and compare analyses of primary and secondary standards
  21. Analyze: Select Standards, Calculation OptionsSelect standards and assign Calculation Options on per standard basis (stoichiometric oxygen, formula basis, trace element output, etc.)
  22. Analyze: Select Standards, Standard Assignments Select standards and assign primary standards on per standard basis
  23. Run--Display PictureSnap, Picturesnap--Load, Calibrate Set up PictureSnap, load image, calibrate image
  24. Automate: Digitize Digitize Unknown Samples for automated analysis
  25. Automate: Digitize, Single Point Digitize Single Point(s) on Unknown Samples
  26. Automate: Digitize, Line Traverse Digitize Line Traverse on Unknown Samples
  27. Automate: Digitize, (Polygon, Shotgun, Digitize Image, etc.) Digitize (other modes) on Unknown Samples
  28. Automate: Acquire Unknown Samples, Run Acquire Unknown Samples using Automate window (digitized samples)
  29. Acquire: Acquire Unknown Samples Manual acquisition of unknown samples using Acquire window
  30. About Sample Setups
  31. Generate Sample Setups Acquire: New Sample-Unknown, Disable Elements, Edit ALL Settings, Analyze: Elements/Cations, Save Sample Setup, Repeat)
  32. Automated analysis using Sample Setups Automate: Digitize Unknown(s), Use Sample Setups Mode, Assign Sample Setups to digitized positions, Acquire Unknown Samples, Run)
  33. Inspect Analyses during run Analyze: Analyze Select samples, Analyze button, change Calculations Options, Standard Assignments, etc.
  34. Output Data at end of Run Output--User Data Ouput Output Unknowns or Standards, Write to .dat and Excel Files
  35. Backup Digitized Sample Positions Automate: (Unknowns) Select all, Export to .pos File, navigate to user folder, save)
  36. Put microprobe in standby mode Perform sample exchange, filament to standby, as necessary


Probe for EPMA Online Manual Section


  1. Acknowledgements
  2. Disclaimer
  3. Conventions Used in this Guide


Using Standard to Generate, Modify, and Maintain the PFE Standard Database

  1. About the Standard Program
  2. Creating the Default Standard Database File
  3. Creating Standard Position Files


Running PFE in Manual Mode from the Acquire Window: Brass Alloy Example

  1. Brass Alloy Run
  2. Introduction
  3. Opening Probe for EPMA
  4. Creating a New Run
  5. Parameter Initialization
  6. Analytical Standard Selection
  7. Creating a New Sample
  8. Setting Analytical Conditions
  9. Nominal Beam Current Measurement
  10. Element, X-Ray Line and Spectrometer Parameters Selection
    1. Background Offsets
    2. Using Calibrated PHA Settings
  11. Setting Count Time
  12. Manual Peaking using the Acquire! Window
  13. Manual Determination of Pulse Height Analysis Settings
  14. Manual Count Acquisition using the Acquire! Window
  15. Wavescan Acquisition
  16. Off-Peak Adjustments from the Plot! Window
  17. Loading Standard Position Files
  18. Automation Actions
  19. Confirm Standard Positions
  20. Calibrate Peak Position
  21. Acquire Standard Sample
  22. Analyze Standard Data
  23. Unknown Sample Data Collection and Analysis
  24. Output of Analyzed Data
  25. Closing the Current Run and Probe for EPMA


Running PFE in Automated Mode using the Automate Window: Silicate Example

  1. Silicate Sample Run
  2. Introduction
  3. Opening Probe for EPMA
  4. Creating a New Run
  5. Parameter Initialization
  6. Analytical Standard Selection
  7. Creating a New Sample
  8. Setting Analytical Conditions
  9. Nominal Beam Current Measurement
  10. Element, X-Ray Line, and Spectrometer Parameters Selection
  11. Editing Acquisition Options
  12. Modifying Standard Assignments
  13. Setting Count Time
  14. Loading Standard Position Files
  15. Automation Actions
  16. Confirm Standard Positions
  17. Calibrate Peak Position
  18. Acquire Standard Samples
  19. Evaluate Standard Count Data
  20. Assign MAN Background Calibrations
  21. Analyze Standard Samples
  22. Spectral Interference Assignments
  23. Unknown Sample Data Collection and Analysis
  24. Digitized Sample Data Collection and Analysis
  25. Plotting Analysis Data
  26. Closing the Current Run and Probe for EPMA

Probe for EPMA Advanced Topics

  1. Element Setups
  2. Sample Setups
  3. File Setups
  4. Quick Wavescan Acquisition
  5. Calibrated Multi-Element Wavescans
  6. Polygon Gridding – Using Surfer Option
  7. Stage Bit Maps and Picture Snap! Feature
  8. Modal Analysis
  9. Deadtime Calculations
  10. Calculation Options
  11. Linear Calibration - Curve Method
  12. Time-Dependeant Intensity Correction
  13. Advanced Interference Corrections
  14. Light Element Analysis - Empirical APFs
  15. References

WDS Scans and Background Offset Data

  1. Kakanui Hornblende WDS Scans
    1. TAP Full Range Scan
      1. TAP Na Scan
      2. TAP Mg Scan
      3. TAP Al Scan
      4. TAP Si Scan
    2. PET Full Range Scan
      1. PET P Scan
      2. PET S Scan
      3. PET Cl Scan
      4. PET K Scan
      5. PET Ca Scan
    3. LiF Full Range Scan
      1. LiF Ti Scan
      2. LiF Cr Scan
      3. LiF Mn Scan
      4. LiF Fe Scan


Calibration of the Microprobe: The Startwin Program

  1. Startwin Program
  2. Beam and Detector Stability
  3. Determination of WDS Spectrometer Alignment
  4. Determination of WDS Spectrometer Deadtime
  5. Determination of Pulse Height Analysis Parameters
    1. PHA Parameters: Bias Scans (JEOL)
    2. PHA Parameters: Gain Scans (Cameca)
    3. PHA Parameters: SCA Scans


Acknowledgements

This online manual is the result of significant work done by: Dan Kremser, Paul Carpenter, John Donovan, with assistance from Ryan Zeigler, Brad Jolliff, and other PFE users.


Disclaimer

In no event shall Dan Kremser, Paul Carpenter, John Donovan, Probe Software, Washington University in St Louis, or any contributing entity be liable to any party for direct, indirect, special, incidental, or consequential damages, including lost profits, arising out of the use of this documentation. No warranties or fitness for a particular purpose is made. This documentation is provided on an as is basis with not obligations to provide maintenance, support, updates, enhancements, or notification.


Conventions Used in this Guide

The computer operating system employed here has used both Microsoft Windows NT (v. 4.0, service pack 4) and Windows XP Pro.  The software used was both Probe for Windows and Probe for EPMA Enterprise Versions (32 bit).

The following conventions are used in this document:

Several tips for saving time/steps include:


About the Standard Program

Probe for EPMA uses a database of microprobe standards for use in quantitative analysis. This standard database can store up to 32768 standards each with up to 40 elements per standard.  All standard information is stored in a file designated STANDARD.MDB.  The database, like all PFE database files, has an .mdb extension, where mdb is an abbreviation for Microsoft DataBase and represents a Microsoft Access v. 3.5 database file.

Probe for EPMArefers to standards by standard number rather than the name of the standard. It is important to understand the following:

In addition to the default standard database, several other standard databases are supplied as ASCII files. These are:

The DHZ.DAT file is a database of all of the analyses listed in the first edition of “Rock Forming Minerals” by Deer, Howie and Zussman.  The ORE.DAT file is a database composed of sulfide minerals from Dana’s Mineralogy entered in ideal formulas.  The SRM.DAT file is a database of SRM (Standard Reference Materials) alloys and glasses from the NIST SRM catalog.  All of these database files can be used for reference and matching purposes but must first be imported into a Probe for EPMA standard database file (see User’s Guide and Reference documentation) using the File | Import command in STANDARD.


Capabilities of the Standard Program

For more information on the Standard Program see the PFE Help documentation. Here is a summary of what the Standard program is used for.

Capabilities of the Standard program:

Creating the Default Standard Database File

The following procedure illustrates how to create a new default standard database and enter standard compositions into it.  To import standard compositions from 16 bit Probe for EPMA see the User’s Guide and Reference documentation.

From the Desktop, double click on the yellow EPMA Software folder. Then double click on the Standard icon in the EPMA Software group.



This action launches the STANDARD (Compositional Database) program and opens the Open Standard Database File dialog box.


Creating a New Standard Database

You will create a new standard database only once when you are setting up Probe for EPMA as you will use that database for all subsequent work. To create a new standard database, click on the Cancel button to close the Open Old Standard Database File dialog box.


Select File from the menu bar and then click on New from the menu.


This opens the Open New Standard Database File dialog box.


Click the Save button to open a new default standard database (STANDARD.MDB).

The Open New Standard Database File window appears.


Click the Yes button to confirm overwriting the existing default database.  Note: the supplied demonstration files JEOLEL.MDB and JEOLOX.MDB will no longer be usable after this operation.

The File Information window opens.



Enter the relevant information into the User, Title, and other Description text boxes shown in the File Information dialog box displayed below.  Use the tab key to move between text boxes.


When finished, click the OK button.


The user now has an empty database ready to accept standard composition data.



Entering a New Standard into the Standard Database: A Metal Standard

To enter standards into this database, select Standard from the menu bar and click on New from the menu.



This action opens the Standard Composition dialog box.  Type in the appropriate Sample Number, Standard Name, and Standard Description into the text boxes.  The software automatically loads the next available number by default.  Choose standard numbers that will allow grouping of standards into various functional sets.  Standard numbers may range from 1 to 32768, however to avoid conflict with the supplied NIST SRM, DHZ, and Dana ORE sample databases select numbers below 2000.


Note that standard compositions may be entered and displayed using either oxide or element weight percent. These settings are defaults as set in the probewin.ini file and may be changed if the typical analytical project involves metals vs. oxygen-bearing materials.

Also note that compositions may be entered as a formula string by clicking on the Enter Atom Formula Composition button. The composition may be entered as a specific formula stoichiometry such as Mg2SiO4, which would parse to the elements and weight percents for the end-member olivine mineral forsterite. For compounds with non-integer value for the elements it is necessary to convert to integer values. For example, an intermediate olivine with the formula Mg1.84Fe0.16SiO4 needs to have each element multiplied by 100 and entered as Mg184Fe16Si100O400.

This shortcut can also be used to populate the elements without initially entering a composition by simply typing MgSiO into the field followed by editing the weight percent values for each element.

The first example will illustrate the entry of an elemental metal standard.

Click the Elemental Percent and Elemental Standard buttons under Enter Composition In and Display Composition As respectively, as necessary.  All standard compositions are saved to the standard database as elemental concentrations.  If oxygen is present in the standard then the user must enter oxygen as an element and its concentration into the standard entry.  See the silicate example in this manual for details.

To enter the composition for each element individually, click on any empty row in the spreadsheet.  This opens the Element Properties dialog box.  In the Element field either type in the first element in the standard or use the drop-down list box to select the element symbol.  Continue by choosing the correct X-Ray line, Cations, and Oxygens.  The X-Ray line is used for modeling purposes only.  When entering properties and concentrations for elements in elemental mode, the program grays out the Cations and Oxygens text boxes, no editing of these text boxes are necessary.



Enter the elemental weight percent for copper into the Enter Composition In Elemental text box.  Finish by clicking the Ok button of the Element Properties dialog box.


The program returns to the Standard Composition dialog box.



If there are more elements (compound standards) in the standard, click the next empty Element row and repeat the data entry process.  When all elements are entered, click the OK button on the Standard Composition dialog box.  This concludes the entry of a standard into the standard database and results in the following log window output.



Entering a New Standard into the Standard Database: A Silicate Standard

Many standards contain oxygen in their compositions.  Since all standard compositions are saved to the standard database as elemental concentrations, it is necessary to enter the oxygen concentration if oxygen is present in the compound.  This applies to all standards, even those that are entered and/or displayed as oxide concentrations.  The following example illustrates a silicate (oxygen bearing) standard entry into the database.

From the main STANDARD log window, select Standard from the menu bar and click on New from the menu choices.  This action opens the Standard Composition dialog box.  Type in the appropriate Sample Number, Standard Name, and Standard Description into the text boxes Click the Oxide Percent and Oxide Standard buttons under the Enter Composition In and Display Composition As boxes.


Click on any empty row in the spreadsheet.

This opens the Element Properties dialog box.  In the Element field either type in the first element in the standard or use the drop-down list box to select the element symbol.  Continue by choosing the correct X-Ray line, Cations, and Oxygens.  Finally, enter the weight percent for SiO2 into the Enter Composition In Oxide Weight Percent text box.


Finish by clicking the OK button of the Element Properties dialog box.  This results in the following Standard Composition dialog box.


Note: to facilitate the data entry for the oxygen concentration of standard compositions which are entered as oxide concentrations, the program will display a running total in the text box designated Total Oxygen From Cations.

Continue the data entry process for the remaining elements (as oxides).



To complete the standard entry into the standard database, enter oxygen as the last element in the standard.  Click on any empty row in the spreadsheet.  This opens the Element Properties dialog box.  In the Element field type in the element symbol for oxygen.  Check for the appropriate X-Ray line, Cations, and Oxygens.  Finally, enter the running total from the Total Oxygen From Cations text box into the Enter Composition in Oxide Weight Percent text box.


Click the OK button of the Element Properties dialog box.

The following Standard Composition dialog box illustrates the completed five-element silicate standard, Albite.



The compositional data of any standard entered into the standard database may be reviewed by simply double-clicking on the standard of interest from the scrollable Standards list box.  The following window contains two standards with the compositional data of Albite displayed in the log window in oxide form.


To modify a particular standard, select the standard in the Standards list box.  Click Standard from the menu bar and select Modify from the menu.  Edit the appropriate fields in the Standard Composition window as described previously.

After entering all of the standard compositions in your standard collection, save this important file (STANDARD.MDB) to another directory on the hard disk and likewise to another storage media for archival purposes.

Note: the takeoff, kilovolt, x-ray and cation ratio parameters displayed here are used only for nominal calculations of the k-factors and ZAF corrections within the program STANDARD.  The Probe for EPMA quantitative analysis will calculate the quantitative standard k-factors based on the actual conditions.


Creating Standard Position Files

Program STAGE.EXE is used to digitize your standard mounts to create pre-digitized standard coordinate files.  These files are necessary for automated acquisition and standardization.  The standard coordinates are digitized in three dimensions (X, Y, and Z) as well as the W stage position (multi-position specimen stages only) and are typically referenced to three physical fiducial marks on the standard mount surface.  These coordinate files should be digitized with the standard mount located in the position where it is typically found.

The following procedure illustrates how to create a new standard position file.  In this example, four carbonate standards will be digitized.  These standards must already be entered into the standard database, using program STANDARD.

When creating digitized standard files for standard mounts containing more than 48 standards, a slightly different procedure than outlined below must be followed.  Concise instructions on how to bypass the current 48 standard limit in the STAGE digitize feature are outlined in the reference documentation.  To find these instructions, open the PROBEWIN.HLP program from the EPMA Software folder.  Click the Help Topics button and type in digitize in the text box.  Highlight the topic entitled Digitizing Standard Mounts with More than 48 Standards and click the Display button.


Open STAGE (Stage Control and Automation) by double clicking on the Stage icon in the EPMA Software group.



The following window does not appear on JEOL and Cameca PFE interfaces

This starts the STAGE program and brings up the Confirm Motor and Crystal Positions dialog box.  Confirm that all of the motors (stage and spectrometer positions) and crystal designations are correctly calibrated.  If there is disagreement between the mechanical positions (actual) and the software values, adjust the software values.  Use the tab key to move between the Stage and Spectrometer Positions text boxes.  Click the OK button to close the Confirm Motor and Crystal Positions dialog box when done.



The following display illustrates the STAGE log window.



Select Standard from the menu bar and click on Add/Remove Standards To/From Run from the menu choices.


This action opens the Add Standards to Run dialog box.  Click on the name of each of the standards in the standard block to be digitized from the Available Standards in Database list box.



Click the Add Standard to Run >> button to move these standards into the current run.  Standards maybe added one at a time or the user may multi-select standards by holding down the Ctrl button on the keyboard as standards are selected.  Double clicking each entry will also send the standard to the other column in the dialog window.


Click the OK button of the Add Standards to Run dialog box when finished.


Select Window from the menu bar and choose Digitize Positions from the menu choices.


If this is the first time you have digitized positions, then the following message will be displayed.


Click the OK button.


This action opens the Digitize! dialog box.


Click the Fiducials button.


This opens the Select Fiducial Set window.


Click the New button.  This opens the Modify Fiducial Positions window.  The current stage coordinates are loaded by default.



Type in a Fiducial Description.  Enter the nominal coordinates or move to each of the three fiducial marks on the standard mount, determining their approximate coordinates, and enter those values into the appropriate fields.  On JEOL 733 microprobes, the W stage position needs to be recorded as well.  The following window results.


Click the OK button when done.  This creates a new entry in the Select Fiducial Set list box as shown below.


Select (highlight) the new fiducial set and click the Confirm button to initiate a precise centering of the three fiducial marks.


The Modify Fiducial Positions window opens displaying the originally entered fiducial coordinates.  Click the OK button to initiate the centering process.


The computer then drives to each fiducial mark and displays the FiducialVerifyFiducial window.  Adjust the stage position to center the fiducial mark and click the OK button.


After centering the third fiducial mark and clicking the OK button, the FiducialVerifyFiducials window opens to display the specimen tilt in radians and degrees.  A warning will be given if the sample is tilted at more than 0.5 degrees.


Click this OK button.


Closing the FiducialVerifyFiducials window returns to the Select Fiducial Set dialog box.


Finally, click the OK button on the Select Fiducial Set dialog box.  This opens the FiducialSaveSelect window to confirm the currently selected fiducial set.


Click the OK button of the FiducialSaveSelect window.


The fiducial coordinate positions are recorded to disk and the Digitize! dialog box returns.



The position of each of the standards in this standard mount must now be digitized.  Move to the first standard; either by turning the motor controls manually or using the joystick via the JOYWIN (Joystick Control for Stage and Spectrometers) program or use the Move button in the Digitize! window.  Clicking the Move button opens the Move Motors and Crystals dialog box.

Type in the appropriate target coordinates in the Stage Target Positions boxes for the first standard.  Use the tab key to move between entries.


Click Go or press Enter on the keyboard, this will drive the stage to the target positions.  Check the position and optical focus.


Click the Digitize button of the Digitize! dialog box.  This activates the Digitize Sample Positions dialog box.  The Standard Position Samples list box contains the standards already added to the run.



Select (highlight) the first standard to digitize from the Standard Position Samples list box.  The standard will be added automatically to the Digitize! Position List.  If a BMP electron image file exists for the standard, then it too will be displayed. This can be used to check whether you are on the correct standard.  Further, the image could be annotated to suggest places or grains to avoid during the standard digitization or acquisition process.


To digitize a random point on this standard, click the Random Point button of the Digitize Sample Positions dialog box to record the current coordinates (X, Y, Z, and W) for this grain.


The coordinates of this standard in the Digitize! dialog box, are seen below.


Note, that although only one position per standard need be digitized, if additional points are digitized, Probe for EPMA will automatically utilize them.  Otherwise, Probe for EPMA will simply increment the stage X position for each additional acquisition required.


Move to the next standard, select the standard from the list box in the Digitize Sample Position window and click the Random Point button again.  The standard position will be digitized.  Continue until all of the remaining standards in the standard block are digitized.  In this example the Digitize! and Digitize Sample Positions dialog boxes would appear as follows.


Close the Digitize Sample Positions dialog box by clicking the Close button in the upper right corner.


Finally, store the new pre-digitized standard coordinates to disk as an ASCII position file (.POS).  Select all of the standards using the Select Stds button of the Digitize! dialog box.


Click the Export Selected Samples button.


This action opens the Open File To Export Position Data To window.  The default Save in: location is specified by the StandardPOSFileDirectory keyword in the PROBEWIN.INI file.


Type in an appropriate File name:


Click the Save button of the Open File To Export Position Data To window.


After the positions are written to disk, the AutomateExportPositions window appears.


Click the OK button to confirm the exported position coordinate data to disk in the AutomateExportPositions window.

Close the Digitize! dialog box by clicking the Close button.


Close STAGE by clicking the File | Exit menu.


After digitizing all of the standards on the standard mounts and creating various STDPOSx.POS files, save these files to another directory and to a backup media.



Generic Microprobe Setup

On the WU JEOL 8200 electron microprobe there are 5 wavelength-dispersive spectrometers, and we have a basic assignment of elements to spectrometers for routine analyses. You will likely do the same on your instrument as a function of the number of spectrometers, the available scanning range for use, and the diffracting crystal inventory on each spectometer.

See the full writeup on the WU JEOL 8200 for more information.

The crystal inventory on the WU JEOL 8200 is shown in this table:

Washington University JEOL JXA-8200

Spectrometer

Sp1

Sp2

Sp3

Sp4

Sp5

Crystal 1
PETJ
TAP
PETJ
LiF
PETH
Crystal 2
TAP
LDE2
LiF
PETJ
LiFH
Crystal 3
LDEB
none
none
none
none
Crystal 4
LDE1
none
none
none
none
Spect Range, mm
71 - 258 mm
62 - 258 mm
62 - 258 mm
62 - 258 mm
88 - 235 mm
Counter Type
Flow, P-10 Ar-CH4
Flow, P-10 Ar-CH4
Sealed, Xe
Sealed, Xe
Sealed, Xe

Element Ranges for Diffracting Crystals

Analyzing Crystal

Element Range

LDE1
Kα lines of C, N, O, and F
LDE2
Kα lines of B, C, and N

LDEB

Kα lines of Be and B

PETJ, PETH

Kα lines Si - Cr
Lα lines Kr-Eu
Mα lines Lu - Bi and Th - U
PETJ is a high reflectivity crystal
PETH is on H-type spectrometer

TAP

Kα lines O - Si
Lα lines Cr - Zr
Mα lines La - Pt

LIF, LIFH

Kα lines of Ca - Rb
Lα lines of Sb - U
LIFH is on H-type spectrometer


Element List and Spectrometer Assignment

Default Element Setup for Washington University JEOL JXA-8200

Spectrometer

Sp1

Sp2

Sp3

Sp4

Sp5

Crystal
TAP
TAP
LiF
PET
LiFH
Set 1
Na Kα
Al Kα
Mn Kα
K Kα
Ti Kα
Set 2
Mg Kα
Si Kα
Fe Kα
Ca Kα
Cr Kα


Standards for Generic Run

Default Primary and Secondary Standards Used for Generic Run

Primary standards are used to peak the WDS and to acquire the X-ray intensity for calibration. Secondary standards are analyzed as calibration standards in the pfe run and may be used for primary calibration upon demand.

For more information about standards, including the Smithsonian Microbeam Standards, see the following links:
Microanalysis Standards
http://mineralsciences.si.edu/facilities/standards.htm

Element(s)

Standard (Nominal Composition)

Na, Si
Amelia Albite NaAlSi3O8
Mg
Shankland Forsterite Mg2SiO4, synthetic

Al

Alaska Anorthite, CaAl2Si2O8

K

Microcline or Madagascar Orthoclase, KAlSi3O8

Ca

Wollastonite (Gates), CaSiO3

Ti

Rutile, TiO2, synthetic

Cr

Cr2O3, synthetic

Mn

Mn-olivine, Mn2SiO4, synthetic

Fe

Fayalite, Fe2SiO4, synthetic

Various Elements

Other Primary Standards -- Carpenter S1 and S2 Mounts:
  • Ni-olivine (Ni)
  • Elba Hematite (Fe, O)
  • Durango Apatite (P, F)
  • Corning 95-IRV (trace K, Ti, Cr, Fe, Ce, Hf)
  • Corning 95-IRW (trace V, Mn, Co, Cu, Cs, Ba, La, Th)
  • Corning 95-IRX (trace Ni, Zn, Rb, Sr, Y, Zr, Pb, U)
  • YAG (Y)
  • F-Phlogopite (F)
  • Gahnite (Zn)
  • SrBaNb-oxide (Sr, Ba, Nb)
  • RbTiOPO4 (Rb)
  • DJ35 glass (Na)
  • Tugtupite (Cl)
  • Benitoite (Ba)

Taylor Mount:

  • Coming soon...

Various Elements

Primary and Secondary Standards -- Carpenter S1 and S2 Mounts
  • Kakanui Hornblende
  • Basalt Glass VG-2
  • Kyanite
  • San Carlos Olivine
  • Springwater Olivine
  • Kakanui Augite
  • NBS K411 Glass
  • NBS K412 Glass
  • Others...

Brass Alloy Run



Introduction

This document illustrates step by step how to set up a new Probe for EPMA quantitative run and how to analyze an unknown two-element alloy sample.  This documentation was produced on a three spectrometer JEOL 733 electron microprobe. Your particular run may look very different depending on the specific configuration of your microprobe.  This document should be used in conjunction with the User’s Guide and Reference documentation, on-line help and the PFWUSERWIZARD program.

This run will demonstrate some of the basic features of the Probe for EPMA program.  These include the use of manual and automated spectrometer peaking, manual and automated standard count acquisition and manual unknown sample acquisition.  The use of pre-digitized standard positions, the unique wavescan option, off-peak adjustment capabilities and data output methods will be illustrated.



Opening Probe for EPMA

From the Desktop, double-click on the yellow EPMA Software folder opening the EPMA Software group.  Double click on the Probe for EPMA icon.



Upon launching PROBEWIN (Probe for EPMA), the main log window appears along with the RealTimeInitInterface window as illustrated below.  To collect real time data click the Yes button.  The program can also be run off-line without the microprobe interface to re-process previously acquired data or on another computer.



This action causes the Confirm Motor and Crystal Positions dialog box to open.  Confirm that all of the motors (stage and spectrometer positions) and crystal designations are correctly calibrated.  If there is disagreement between the mechanical positions (actual) and the software values, adjust the software values.  Use the tab key to move between the Stage and Spectrometer Positions text boxes.  Click the OK button after you have finished to close the Confirm Motor and Crystal Positions dialog box.



The main Probe for EPMA log window is now visible as seen below.




Creating a New Run

To create a new sample run, select File from the menu bar and click New from the menu.


The Open New Probe Database File dialog box opens.


Change the Save in: location (directory) if desired and type an appropriate run name into the File name: text box.

The initial Save in: location is specified by the UserDataDirectory keyword in the PROBEWIN.INI file.  File names longer than 8 characters are now supported.


In this example, a new file designated BRASS01.MDB will be created in the Kremser directory.  Any existing old runs maybe re-opened to acquire additional data or used as a “setup” file for starting a new run.  This will be the first .mdb file in this folder.


Close the Open New Probe Database File window by clicking the Save button.

First time users will see the creation of a new User database as shown below.


Clicking the OK button opens the File Information dialog box.


Enter the relevant information for the new run into the User, Title, and other Description text boxes.  Use the tab key to move between text boxes.  When finished, click the OK button.


This returns the program to the main Probe for EPMA log window. Now the four main Probe buttons: Acquire!, Analyze!, Automate!, and Plot! become active.




Parameter Initialization


Analytical Standard Selection

Select the analytical standards to be used in the new probe run.  From the main Probe for EPMA log window, click Standard from the menu bar and select Add/Remove Standards To/From Run from the menu.



This opens the Add Standards to Run dialog box.


All previously entered standards in the default standard database are accessible.  Scroll through the Available Standards in Database list box to find the copper and zinc metal standards to be used in this run.  Select each and click the Add Standard To Run >> button to add each to the Current Standards in Run list box.


Click the OK button of the Add Standards to Run window when finished selecting standards.  This returns the program to the main log window.



Creating a New Sample

Click the Acquire! button in the main Probe for EPMA log window.  This action opens the Acquire! dialog box.  Note, not all buttons are active.


Click the New Sample button of the Acquire! dialog box.


This opens the New Sample dialog box.


Options for Entering Elements into a Sample

There are several options to set up the element list for analysis. The possibilities are as follows:

Here we will be setting up a sample setup from the beginning.

Select Unknown from the New Sample Typebuttons.  Type an appropriate sample name and description into the New Sample Name and New Sample Description text boxes.  This first sample will be used as a “template”, only to establish the analysis parameters.


Click the OK button of the New Sample dialog box.


The program returns to the Acquire! window.  Notice that the first sample designated Un   1  * template is now listed in the Current Sample text box.  The * symbol indicates that no data has been collected for this sample yet.  Note, all of the buttons in Acquire! are now available.




Setting Analytical Conditions

Click the Analytical Conditions button to open the Analytical Conditions dialog box.  Enter the appropriate numbers into the Kilovolts, Beam Current, and Beam Size text boxes for the currently Selected Sample.  The Kilovolts, Beam Current, and Beam Size will need to be manually adjusted if a column digital interface is not present (all parameters other than kilovolts are simply available for documentation purposes only.  If a hardware interface is supported, the user may specify a column condition string to indicate the desired analytical conditions of the instrument.


Click the OK button when done.


The Analytical Conditions dialog box closes, returning to the Acquire! window.




Nominal Beam Current

For modern electron microprobes the nominal beam current is not formally used. Each analysis includes a measurement of the probe current, which is used for calculation of cps/nA data and time-dependent drift correction of the probe current.

There are two settings for the nominal beam current that are used for PFE intensity output:

Open the Count Times window from the Acquire! button.


Any value desired may be directly entered into the Nominal Beam text box (25 nA value is stored in our PROBEWIN.INI file) or the user may measure the present beam current by clicking the Nominal Beam button.


The AcquireCheckNormal dialog box appears, choose the Yes button to measure the present beam current for use in the beam drift correction.



The current value of the faraday beam is measured and reported to both the Acquire! window and the Nominal Beam text box in the Count Times window as seen below.


Close the Count Times window by clicking the OK button.



Element, X-Ray Line and Spectrometer Parameters Selection

Next, the user specifies the elements to be analyzed.  Click the Elements/Cations button.



This action opens the Acquired and Specified Elements dialog box.  Click on any empty row in the spreadsheet to enter the first element to analyze.  The user may enter the analyzed elements in any order however, the analysis output will follow this order.



This opens the Element Properties dialog box.



In the Element field either type in the first element to analyze or use the drop-down menu to select the element symbol.  Certain default values listed in this window are based on parameters entered into the previously established configuration files.



Under the Enter Element Properties For: section (top of the Element Properties dialog box) choose the correct X-Ray Line, Cations, and Oxygens for the first element.  Both alpha and beta lines are now supported as well as the ability of running the same element on all relevant spectrometers.

Background Method Selection


There are several sophisticated methods for measurement or empirical calculation of background intensity in PFE. There are as follows, and are selected from the Parameters section of the Element Properties window:

  1. Off Peak: This is for conventional detuning of the spectrometer to background positions away ("off") from the x-ray peak. The values for the nominal high position are set in the High Off-Peak text box, and for the low position entry is made in the Low Off-Peak text box.
    1. The use of these nominal off peak backgrounds depends on the button selections in the Off Peak Correction Type part of the window:
    2. Linear: A linear fit to the two background measurements is made via y = mx + b.
    3. Average: The simple average of the two background measurements is used.
    4. High Only: Only the high background measurement is used.
    5. Low Only: Only the low background measurement is used.
    6. Exponential: An exponential (e-x) is used to model concave-up background shape. Use Plot window with wavescan sample to perform background model adjustment.
    7. Polynomial: A polynomial fit is made to the measured background points and used for calculation of the background at the peak position. Use Plot window with wavescan sample to perform background model adjustment.
  2. MAN: Mean atomic number empirical background method. Select MAN on a per element basis. Appropriate standards for MAN calibration need to be added to the run. Use Acquire--Acquisition Options--Element measurement order to also assign MAN for unknowns and/or standards.

Here we are using conventional off-peak backgrounds.

Continue by selecting the Off Peak Correction Type and Background Type.  Two background correction methods are available to the user; off-peak and the MAN (mean atomic number) method (see the User’s Guide and Reference documentation for a complete discussion of both types).

Setting Peak and Background Positions

Next, click the text box under Spectrometer and enter the appropriate spectrometer number that will be used to analyze the first element. The drop-down menu may also be used to select the spectrometer number. Choosing a spectrometer number for the chosen element uses the default diffracting crystal on that spectrometer for this element and also loads the theoretical peak position into the peak position field. This theoretical peak position will be modified when spectrometer peaking is performed or you can hand edit this value to correspond to a known position.

The off-peak background values that are shown are nominal offsets calculated by the PFE program. You should understand that these nominal off-peak positions have no correspondence to correct positions, and you need to correct them to known valid background offsets that have been determined by prior experience or by acquiring wavelength scans on samples, standards, or both. The nominal offsets can be used temporarily but must be corrected prior to acquisition of standards during the calibration run.

Bias and Gain assignment

The assignment of gain and bias settings is made after the spectrometer and crystal have been selected. If empirical bias (for Jeol) or gain (for Cameca) measurements have been made and the emppha.dat file has been created for your probe and enabled via the probewin.ini flag, then one can click on the Use Calibrated button in order to look up a calibrated bias or gain value for the spectrometer-crystal pair.

Even if the empirical pha file is not being used, you should have a basic familiarity with the gain and bias values used for spectrometer-crystal settings on your instrument. These can be assigned here and adjusted later after using the Acquire--PHA button to perform the necessary scans.

The next screen shows the edited Element Properties dialog box for copper metal.


Click the OK button of the Element Properties dialog box to accept these element parameters for copper.

The program returns to the Acquired and Specified Elements window with copper now entered into the Element/Cations Parameters table.


Enter the next element in the run by clicking on any empty row of the Acquired and Specified Elements window.  This opens the Element Properties dialog box again.  Enter the appropriate Element, Spectrometer, and adjust all other text boxes and buttons.


The completed Element Properties window for zinc is shown below.


Click the OK button of the Element Properties to enter zinc into the Element/Cations Parameters table of the Acquired and Specified Elements window.

Entering Specified Elements

Any elements that will not be formally analyzed can be entered as specified from the Acquired and Specified Elements window. For example, oxygen is typically determined by stoichiometry. Other examples are stoichiometry determination of carbon as CO2, or any list of elements that make up the matrix composition of samples for which a subset will be analyzed and the remainder will be provided as fixed weight percent. These options are handled in the Analyze--Calculation Options window.

Enter all specified elements but choose the last entry from the x-ray line pull down menu to leave the x-ray field blank. This sets the flag indicating that element will be handled by some specified method which will be determined on a per sample basis from the Analyze window.

Identifying Background Interferences

PFE can be used to list possible background interferences. After all elements in the sample setup have been entered, you can go back to the individual element properties windows and generate the low off-peak and high off-peak interference list. This list by default searches only the elements in the sample setup. You can also search against all elements, which will surely find some potential background interference. Note that this function does not list on-peak interferences.

For a complete listing of background and peak interferences use the Standard program (Option window).

Click the OK button of the Acquired and Specified Elements window when done entering elements in the run.


The GetElmLoadDefaultStds window opens to inform the user that default standard assignments have been made based on elemental concentrations.  Standard assignments may be edited via the Analyze--Standard Assignments button.


Click OK to return to the main Acquire! window.



Peak and Scan Options

Click the Peak/Scan Options button of the Acquire! window.  This opens the Peak and Scan dialog box.  This window allows you to view and edit various background, peaking, and wavescan parameters.  The Display radio buttons allow you to view:

Click on the row of an element to edit any of these parameters (regardless of the radio button state). This opens the Peak and Scan Properties window which has fields which can be edited for each parameter. Note that these values are set on a per-element basis and apply to the current sample (and subsequent samples until you make a change to a parameter).

This window is where full-range wavelength scans are set using a dummy element. For example, using Si on TAP one can acquire a full-range spectrometer scan that nominally is labelled as a Si wavescan but includes the full spectrometer range during scanning.

Edit each element as necessary using the Peak and Scan window. Note that this window duplicates the background offsets that are set using Elements/Cations.

Use ROM Based Spectrometer Scanning is a global switch that uses a continuous step scan on the JEOL and Cameca platforms during wavescan acquisition. If this check box is set then a faster scan is acquired; if the check box is not set the wavescan is acquired using a step scan mode. It is suggested toi use the ROM mode for rapid scans and the non-ROM (i.e., step scan) for slower scans.

Increment Stage During Peakscan/Wavescan or Peaking (X and Y Axis) allows you to increment the stage position on presumably beam-sensitive materials when performing either a peaking or a wavescan acquisition. Stage movement is enabled using the checkboxes for a wavescan (Scanning) or a peaking routine (Peaking), the X and/or Y stage increment value, and the time interval in seconds that is used to increment the stage.


Setting Count Times

Click the Count Times button of the Acquire! window.  This opens the Count Times dialog box.  Here various parameters relating to counting times can be adjusted.  The On-Peak count time is set for 30 seconds and both Hi-Peak and Lo-Peak times are set for 15 seconds. These defaults are based on the probewin.ini configuration file defaults (the image here was captured on another probe).


Note: Real time spectrometer motion and acquisition time is graphically displayed.

Click the OK button of the Count Times dialog box to accept these count times and return to the Acquire! window.

This completes the initial parameter setup phase for this new run.



Manual Peaking using the Acquire! Window

The user may now manually peak center the copper and zinc peak positions from the Acquire! window.

Move to the copper standard by clicking the Move button.  This opens the Move Motors and Crystals dialog box.  Enter the coordinates of copper metal standard into the Stage Target Positions text boxes.  Click the Go button.


The stage motors will move the stage to the expected position of the copper metal standard.  Inspect the final X, Y location of the standard, adjust if necessary and check the focus.

Inspect the spectrometer crystal type and position text boxes, edit if required.  The user may also select the element and x-ray line from the Periodic Table function.  Click on the Periodic Table button.


The Select Element and Xray for Spectrometer window opens, click on Cu in the periodic table.


Click the OK button to return to the Move Motors and Crystals window.


The Move Motors and Crystals window appears as below.


Send the spectrometer directly to the theoretical position by clicking the Go button.

The user could peak spectrometers as well as adjust the PHA parameters (baseline, window, gain, and bias) from the Move Motors and Crystals window.


Click the Peaking Options button of the Acquire! window.


This opens the Peak Center dialog box.



From the Peak Center Method group, choose Interval Halving (see User’s Guide and Reference documentation for discussion of various Peak Center methods) and click Display Spectrometer Pre-Scan for Confirmation from the Peak Center Options choices.  Finally, select the Cu kα Spec  2 LiF (107.220) selection under the Elements to Peak list box.  The Peak Center window should appear as follows.


Click the OK button to close the Peak Center dialog box.


Click the Start Peaking button in the Acquire! window.


This action opens the Wavescan Acquisition window.  The spectrometer then performs a spectrometer peak pre-scan (40 step, user defined parameter) on spectrometer 2 in the copper Kα region.


Upon completion of the spectrometer pre-scan the Peak Center Start Position Selection window opens.



Slide the scroll bar to move the vertical (red) peak line to match the actual x-ray maximum position.  This selects a starting peak center position for the peaking routine.

Click the OK button when manually centered.  This initiates a peak center routine to locate the precise peak center for copper Kα x-rays.  The Peak/ROM/PHA Scan Acquisition window opens and real time peaking can be viewed.


The results for copper appear in the main log window, displayed below.



Zinc Kα may also be peaked in this manner.  First click the Move button of the Acquire! window to access the Move Motors and Crystals dialog box.  Enter the appropriate stage target positions for the zinc metal standard and click the Go button.


When the stage motors stop moving, inspect the location, and adjust the optical focus if necessary.  Close the Move Motors and Crystals window.  Then click the Peaking Options as in the copper example.  Remember to highlight only Zn kα Spec  3 LIF (99.8904) this time.  Close the Peak Center window and finally click the Start Peaking button of the Acquire! window.


The Wavescan Acquisition window appears as below, adjust the slider bar (not shown) to manually adjust the peak position in the Peak Center Start Position Selection window.


Upon completion of the peak centering the results for zinc Kα are displayed in the log window.




Manual Determination of Pulse Height Analysis Settings

Pulse Height Analysis (PHA) refers to the proper detection of X-ray pulses as measured by the X-ray detector of the wavelength-dispersive spectrometer. Energy-dispersive PHA electronics are sophisticated and continuously monitor the zero-energy pulse and also have digital electronics that resolve the wide array of X-ray pulses that enter the EDS system and must be properly measured. WDS detector electronics are not as sophisticated since the spectrometer is tuned to a specific X-ray peak position and only a very narrow range of pulse energies are therefore counted by the detector electronics. It is important for the analyst to ensure that the pulse height analysis system is set properly.

The proper PHA settings for WDS analysis require that the effective gain is sufficient to cleanly separate the X-ray pulses from the baseline noise of the detector system. All WDS counting systems are operated at a detector voltage, typically referred to as the bias, which is typically 1500-1900 volts, and results in the detector operating in the proportional counter range, where a photon produces a signal that is proportional to the X-ray energy.

The effect of increasing the detector bias is to increase the effective resolution of the X-ray pulse from the baseline noise, so that on an SCA plot the X-ray pulse moves to a higher voltage value. The effect of increasing the detector gain is to also increase the resolution and also moves the pulse to a higher SCA voltage. In practice, there are equivalent sets of bias and gain settings that should result in identical counting performance as measured by P/B and total count rate. Given the sets of equivalent bias and gain values, it is prudent to use a higher gain coupled with a lower detector bias in order to maximize the life of the detector. This implicitly assumes that modern counting systems do not exhibit increased noise at higher gain settings, which appears to be the case. You should demonstrate to your own satisfaction that these relationships are observed on your system.

Two different approaches are used for JEOL and Cameca microprobes which result in similar discrimination of pulses from baseline noise. On the JEOL microprobe, the detector gain is chosen from integer values of (typically) 16x, 32x, 64x, or 128x, and the bias is adjusted relative to the chosen gain value in order to achieve the proper PHA settings.

On the Cameca microprobe, the gain is continuously variable and the procedure is to choose a detector bias and adjust the gain value again in order to achieve the proper PHA settings. Experience with WDS counting systems has revealed that there is a dependance of the deadtime of the counting electronics on X-ray energy, which can be minimized by always applying the same effective gain to the X-rays being measured. This is the reason for adjusting the bias and gain in order to produce a consistent single channel analyzer (SCA) plot of the observed voltage distribution returned from the counting electronics. On a JEOL microprobe the bias and gain are set to obtain a ~4V SCA pulse, and on a Cameca microprobe the bias and gain are set to obtain a ~2.5V SCA pulse distribution. These relationships are summarized below:

JEOL Microprobe PHA Method

For a given spectrometer, generally two gain values can be used that cover all X-ray energies for the crystals on that spectrometer:

Cameca Microprobe PHA Method

After the peak position has been determined, it is important to verify that the Pulse Height Analysis (PHA) settings are correct for this element. The spectrometer peak position can be determined with PHA settings that are only approximately correct since the peak intensity is usually obvious. The reverse approach is not practical as adjusting the PHA settings requires that the spectrometer is already tuned to the peak position.


Manual Count Acquisition using the Acquire! Window

To acquire a single point of x-ray count data for a standard proceed as follows.  From the Acquire! dialog box click the New Sample button.  This opens the familiar New Sample window.  Click on Standard from the New Sample Type buttons.  This allows the user to specify a standard from the list now active at the bottom of the New Sample dialog box.  Click 530 Zinc Taylor, its name now appears under New Sample Name.  Click the OK button when done.


Check the optical focus on the zinc standard and click the Start Standard or Unknown Acquisition button of the Acquire! window.  Notice that the current sample is displayed in the Acquire! window.  The progress of all data acquisition may be viewed in the Acquire! window.


Clicking the Start Standard or Unknown Acquisition button initiates the data acquisition.  The spectrometers for copper and zinc move to their respective peak positions and count on peak and off peak (both sides in this example) for times specified earlier in the Count Times window.  The Faraday cup is also measured.

Both the Acquire! window (screen capture during acquisition) and the main Probe for EPMA log window (after acquisition) appear as below.  Faraday cup counts (BEAM) are reported in total counts in the Acquire! window and in nanoamps in the main log window.




Repeated clicking of the Start Standard or Unknown Acquisition button acquires additional intensity data.  The following log window illustrates the acquisition of five data points on the zinc metal standard.


Similarly, x-ray counts can be acquired on the copper metal standard.  Move back to the copper standard position via the Move button and inspect the location and focus.  Click the New Sample button, select the copper standard from the standard list, click the OK button when done.


Start collecting counts by clicking the Start Standard or Unknown Acquisition button.  Repeating five times as with the zinc standard gives the following main log window output.


Inspection of the copper data reveals an interesting feature.  All of the zinc off-peak corrected counts on the pure copper standard are very negative suggesting that a background position may be incorrectly set.  This can be easily checked by moving to an alloy sample containing both elements of interest and performing a wavescan.  A Cartridge Brass standard (NIST SRM 478) containing both elements may be used.



Wavescan Acquisition

To perform a wavescan acquisition on the Cartridge Brass standard, click the Move button in the Acquire! window and move to this standard.

Click the New Sample button.  Select Wavescan under New Sample Type, edit the New Sample Name and New Sample Description text boxes, as desired.


Click OK when done.


Click the Start Wavescan button of the Acquire! window.


This action opens the Wavescan Acquisition window and automatically initiates a 100 step (user defined) wavelength scan for all of the elements entered into the current sample.  In this example, simply copper and zinc.  Graphical output of the completed scan via the Wavescan Acquisition window can be seen below.



The wavescan labels appear in the main Probe for EPMA log window.


The wavescan positions and counts may be displayed in the main log window by clicking the Analyze! button opening the Analyze! window.  Select the Wavescans button, then click the Select All button and finally clicking the Data button.


A portion of a very long list of data is illustrated below.


A more complete graphical display of these wavescans may be accomplished using the Plot! window.



Off-Peak Adjustments from the Plot! Window

Click on the Plot! button under the main Probe for EPMA log window.  The Plot! dialog box opens.


Under Sample List the Wavescans button should be clicked.  Use the mouse to select both wavescan data samples.  Since each sample can only accommodate 50 data points, a complete wavescan of 100 points is continued in two sample numbers.


Click on Cu (2) Spectrometer from the X-Axis list and Cu (2) Wavescan Counts from the YAxis list selections.  The number following the element label (2, in this case) designates which spectrometer collected the data.  The same element maybe run on multiple spectrometers (see User’s Guide and Reference documentation for additional details).  Choose a Graph Type, click the Line button and an Output Target of Send Data to Plot Window.  Finally, click the Output button to view the graph.



The program then loads the selected data into the Plot Graph Data window.


The Plot Graph Data module allows a more robust treatment of the wavescan data.  The plot of Cu (2) Spectrometer position versus Cu (2) Wavescan Counts (labeled as LIF cps) is graphed as well as the locations of the on-peak and both off-peaks (green vertical lines).  Various options are available for evaluation of the data.  Besides click and drag Zoom capabilities, a large selection of KLM Markers options may be enabled.  Further, a model background option is available (see User’s Guide and Reference documentation for a complete discussion of this feature).


With the Zoom On button active, simply click and drag the mouse over the region the user wishes to magnify.  The Analyzed Elements button of the KLM Markers may be selected, painting the various x-ray line positions for all analyzed elements into this plot.


The default choices for both copper background positions (green vertical lines) appear sound as no analyzed element lies nearby and the background counts near these peaks are low.  Click the OK button of the Plot Graph Data window to return to the Plot! dialog box.


Next, the user evaluates the same data set for zinc.  From the Plot! dialog box, select Zn (3) Spectrometer from the X-Axis list and Zn (3) Wavescan Counts from the Y-Axis list selections.  Click the Output button.



The Plot Graph Data window for the zinc data set is shown below.  Two peaks are visible and the user observes that the low background position lies on top of the smaller unknown peak.


Use either the KLM Markers or the NIST x-ray database via the Load Xray Database button to evaluate the unknown peak.  To open the NIST x-ray database, first click the User Selected Lines button, this causes the Load Xray Database button to read Choose Selected Lines.  Click this button to open the NIST x-ray line catalog.


The Xray Database window opens and the user may select or multi-select any x-ray line to plot in the Plot Graph Data window, simply highlight a line(s) and click the Graph Selected button.


Close the Xray Database window when finished.


Examination of the data suggests that the second peak is due to the copper Kα line.  It is apparent that the low background position for zinc needs to be adjusted away from the copper Kβ line.


The user adjusts the low off-peak position by clicking the Low button, creating a crosshair on the Plot Graph Data window.  At any place on the plot the Zn Spectrometer position and Zn Wavescan Counts may be read, the values appear in the box below the two-way Zoom On/ Hot Hit On button.  Move the crosshair to a new low background position and click the mouse.  A new vertical line appears.


In the screen capture below the low background position (green line) has been moved to a lower L value, away from interfering copper Kβ line.


Click the OK button to update this background position in the run and close the Graph Data window.  The GetPeakSave window appears and the user is notified that new parameters (off-peak position) will take effect on the next new sample.


Click this OK button, returning to the Plot! window.

Finally, click the Close button to exit the Plot! window returning to the main PROBE FOR WINDOW log window.



Loading Standard Position Files

To run analytical standards using automation, requires that the computer know the physical location of all the standards for this run.  Click the Automate! button from the main Probe for EPMA log window.



This opens the Automate! dialog box shown below.


The last set of digitized standards used is visible in the Position List list box of the Automate! window.  Currently, the four carbonate standards digitized for an earlier section of this manual are listed.  These will be deleted and replaced by the appropriate standard position file(s).  Normally, after initial setup, several sets of digitized standard positions would be visible in this list.  Typically, the user would not delete these but rather append other position files to the list.


Click the Delete All button.  This opens the AutomateDeleteAll window, seen below.  Click the Yes button of the AutomateDeleteAll window to clear the Position List list box of all displayed position samples.


The FiducialDeleteUnreferenced window opens.


Click the Yes button to clear the fiducial coordinate set from the position database.

Click the Import from ASCII File button of the Automate! dialog box to import position samples from a previously saved ASCII file.



This action opens the Open File To Import Position Data From window.  It is assumed that the user has previously digitized all standard blocks and created STDPOS*.POS files.  The metal standards to be used in the brass analysis are digitized in STDPOS2_TAYLOR.POS.


The *.POS files are located in the directory specified in the probewin.ini file. This will be similar to C:\User Data\PFE Position Files directory.  This is the same location as the BMP electron image files for the standards.


Type in the appropriate file name in the Filename: text box or simply highlight the file in the list and click the Open button.


This action opens the FiducialLoad window.  Click the Yes button to do a fiducial transformation on this pre-digitized standard block to obtain an accurate set of standard positions.


The Modify Fiducial Positions window opens.  Normally the user would simply accept the defaults or edit the position text boxes for each point, including the appropriate stage location number (JEOL 733 use appropriate W stage position).  When done, click the OK button.



This action causes the stage motors to drive to the first fiducial coordinate in its lookup table.  The FiducialVerifyFiducial window appears.  Adjust the stage motors to center the first fiducial mark, click the OK button.


The computer will drive to each of the three fiducial marks for centering.  Clicking the OK button after the third fiducial mark opens the FiducialVerifyFiducials window.  Click this OK button.


The program then imports and updates the position coordinates of all of the standards in the pre-digitized standard position file.  The AutomateImportPositions window opens. Click the OK button returning to the Automate! window.



The Automate! window would appear as below.  The currently transformed standard position file is listed in the Position List list box.  In this example, the copper standard, number 529, has been highlighted and its coordinates are visible.  If an additional standard position file (standard block) is required for use in the automation, the same procedure would be followed.



All of the standards listed in the Position List list box may now be accessed by the program during any automation action.  For instance, it is now possible to have the computer drive to any standard on this block.  The user may click the Move button of the Automate! window opening the Move Motors and Crystals dialog box.  Then, click the Positions button.



This opens the Position Database dialog box.  From here any sample that has been digitized can be located by simply selecting it and clicking the Go button.


Once the stage motors drive the stage to the chosen standard, exit the Position Database by clicking the Close button.  Likewise, the user may close the Move Motors and Crystals window by clicking its Close button, returning to the Automate! window.



Automation Actions


Confirm Standard Positions

All of the basic peak centering and x-ray count acquisition procedures may be automated.  This is accomplished via the Automate! window.

Click the Select Stds button of the Automate! dialog box.  All standards that have been added to the current run will be highlighted in the Position List list box. In this example, the two standards copper (529) and zinc (530) metal are highlighted.



The user might start by checking the location and focus of each standard selected for the automated analysis.  Click the box for Confirm Standard Positions under Automation Actions.


Click the Run Selected Samples button.


The AutomateConfirmSelected window opens informing the user that two standards were chosen and asks if the user wants to run these automated samples, click Yes.


The program then sends the stage motors to the fiducial transformed coordinates for the first selected standard and opens the Confirm Positions window.  Clicking the two-way Pause/Continue button suspends the 10 second countdown (user defined in the PROBEWIN.INI file).  Adjust the stage motors (X , Y, and Z) to a new, clean analysis position.  Click the OK button of the Confirm Positions window when done, sending the stage to the next standard to confirm its position.  Again, the Confirm Positions window opens, allowing the user to pause the countdown and adjust the sample position.

If more than one position is digitized, the software moves to the first position and updates all positions for that sample by the same X, Y, and Z offset.


After the final standard is confirmed, the AcquireStop window appears.  Click this OK button returns to the Automate! dialog box.



Calibrate Peak Position

X-ray peaking may be automated from the Automate! window as follows.  Under Automation Actions click only the Peak Spectrometers box.  Under Automation Options click the Peak on Assigned Standards box.  This option causes the program to attempt a peak center on a standard position sample if the standard is assigned as the primary standard for that element.  If the element has no assigned standard, then the program will attempt to assign one automatically based on the highest concentration of the elements present among the standards in the run.

Next, click the Peaking button to open the Peak Center dialog box.


In the Peak Center dialog box, highlight (select) all of the elements in the Elements to Peak list box, and click on a Peak Center Method.  A spectrometer pre-scan is useful if that element has not been run recently or if maintenance has occurred on the spectrometer.  Click the OK button of the Peak Center window.

Click the Run Selected Samples button from the Automate! window.

This opens the AutomateConfirmSelected window.  To run these automated samples, click Yes.


The stage motors move to the position coordinates of the first standard in the Position List list box.  If the Use Confirm During Acquisition box under Automation Options is checked then the computer automation will pause at each standard (Confirm Positions window will open) for some user defined amount of time to allow the operator to adjust the stage position and focus.  The spectrometers go through the peaking routine to peak center the spectrometer position to the intensity maximum for all the elements assigned to that standard.  After finding a new peak position and reporting the results to the main log window, the stage motors move on to the coordinates of the next standard highlighted in the Position List list box.  Once situated on this standard, the spectrometers peak center those elements assigned to it.  This procedure continues until all standards are done.  When all automation action is complete, the AcquireStop window appears and requests the user to click the OK button.


The following summary of the peak automation for the two standards is found in the main log window.




Acquire Standard Samples

The next step is to calibrate the analytical standards in preparation for unknown samples.  The user may automate the entire acquisition of x-ray counts on all standards as follows.

From the Automate! dialog box and under Automation Actions, click only on the Acquire Standard Samples box.  Under Automation Options, select the number of Standard Points To Acquire and whether to use the Confirm During Acquisition feature.  In this example, five standard points are entered along with a Standard X Increment of 10 um, as well as the Confirm During Acquisition option.


Click the Run Selected Samples button.


The AutomateConfirmSelected window opens again, informing the user that two standards are chosen and asks if you want to run these automated samples, click Yes.

The stage moves to the coordinates of the first standard in the Position List list box, the Confirm Positions window opens, allowing a readjustment of the stage position and optical focus.  A complete analysis (all elements in the current sample) are measured, x-rays are counted on peak and at both background positions for times specified in the Count Times window.  Finally, the Faraday cup is measured.  The stage jogs 10 um in the X direction and this procedure is repeated for the number of points specified in the Automation Options section of the Automate! dialog box.  After completing data collection on the first standard, the stage travels to the next standard in the list and acquires five complete analyses on that standard.  After finishing the automation schedule the familiar AcquireStop window opens and requires the user to click the OK button thereby returning to the Automate! window.


The log window result for the copper standard x-ray count acquisition is seen below.



The log window result for the zinc standard x-ray count acquisition is displayed below.




Analyze Standard Data

After standard data is acquired it is useful to analyze the data to check for agreement among standards and for possible interferences.  Click the Analyze! button in the main Probe for EPMA log window.



This opens the Analyze! dialog box.


The Sample List list box contains the standards acquired so far.  To examine the data acquired on the two standards run under automation, first choose the copper metal, selecting St  529 Set   2 JEOL Copper and click the Analyze! button.


The results for the five automated standard analyses of the copper metal are shown below.  Each individual line (111G to 115G) is illustrated along with the Average, Std Dev and a variety of other statistical parameters for the acquired points (see User’s Guide and Reference documentation for additional details).


If the sample that has been run is a standard, the program will show a Published line as well in the analysis output.  This is the weight percent value for the element as entered in the standard database.  If an element is not found in the standard database it is shown as n.a. or not analyzed.


Selecting St  530 Set   2 Zinc Taylor and clicking the Analyze button calculates the five individual data points (116G to 120G) as seen in the Analyze! window.


Both standards look ok, the user may move on to analyze unknowns.



Unknown Sample Data Collection and Analysis

To collect x-ray data on an unknown sample, minimize the Analyze! window and/or bring forward the Acquire! dialog box to start a new sample.

Click the Move button on the Acquire! window.  Either enter the coordinates of the first unknown sample and click Go or click the Positions button to open the Position Database window.


Select the sample of interest and double click the Row number to automatically drive the stage to the coordinates of the first unknown sample.  Minimize or click the Close button, return to the Acquire! window.


Click the New Sample button to activate the New Sample dialog box.  Check that the Unknown button under New Sample Type is checked.  Enter an appropriate sample name and description into the New Sample Name and New Sample Description text boxes.


Click the OK button.


To start acquiring x-ray counts on the first unknown sample (NIST 478), simply click the Start Standard or Unknown Acquisition button of the Acquire! window.  Progress of the analysis can be followed graphically (far right of Acquire! window, yellow bar or by watching the scaler values.


The Start Standard or Unknown Acquisition button is clicked five times, to acquire five data points at various sample locations.


Next, the Analyze! dialog box is reopened or simply brought forward.  Click the Unknowns button under the Sample List buttons and highlight (select) Un   2 NIST 478.



Clicking the Analyze button calculated the results for these five points and those values are viewed below.


The certified values for NIST 478 are 72.85 wt% copper and 27.10 wt% zinc.  This sample can also be used as a secondary standard to check the quality of the standardization.  If the sample is a standard sample, the analysis printout will list several other output lines (see page 134, zinc metal).  The “Published“ weight percent value for the element as entered in the default standard database and the “Std Err” which is the standard deviation divided by the square root of the number of data points.  After this is a line labeled “%Rel SD” is given, which is the standard deviation divided by the average times 100.  These parameters give the user an easy way to evaluate the quality of the standardization.


Output of Analyzed Data


Before closing the run, the user decides to output the data to an ASCII file for importing to another application such as Excel and to the laser printer for a hardcopy.  From the main Probe for EPMA log window, select Output from the menu bar and click Save to Disk Log from the menu.


This opens the Open File To Output Probe Data To dialog box.  Edit the Save in: location and type in a File name.  This output file has the extension .OUT.  Note that all raw data is always automatically saved in the .MDB run file for future re-calculation and /or output.


Click the Save button.

Select the Analyze! button in the main Probe for EPMA log window, to bring forward the Analyze! dialog box.


Select from the Sample List of the Analyze! window the samples to be output to this file.


Click the Analyze button to reanalyze the samples.


When the program finishes recalculating data to the file, return to the main Probe for EPMA log window.  Select Output from the menu bar again and click View Disk Log from the menu.


This opens the file editor.  A number of text file editors may be used.  To utilize a specific editor such as Notepad or Textpad, edit the FileViewer keyword in the PROBEWIN.INI file.


This example utilizes the Programmer’s File Editor, seen below.


The user might note that the output file is designated TEMP.OUT rather than the BRASS01.OUT specified earlier in the Open File To Output Probe Data To window.  When the user wishes to view the disk log, the program closes the active log, creates a copy called TEMP.OUT, reopening the original (BRASS01.OUT) for further log writing, and allows the user to view the copy.  Each time the user reopens the disk log and assuming new information has been appended, the editor will prompt that the TEMP.OUT file has been altered on disk by another application.  Do you want to load this new version?  Select the Yes button to view the additional material.  Also, if the user types a comment (to annotate output) into the main Probe for EPMA log window, this comment string is echoed into the active disk log.  Be aware that the comments entered will be placed at the current end position of the active log window.

The user may now direct the log data to a laser printer for hard copy viewing by selecting File from the Programmer’s File Editor menu bar and clicking on Print in the drop-down menu.

Exit the file editor and return to the main Probe for EPMA log window.


Closing the Current Run and Probe for EPMA

The user ends the analysis session from the main Probe for EPMA log window.  Select File from the menu bar and click Close from the menu selections.


This opens the ProbFormCloseFile window, click Yes to close this file.


Close Probe for EPMA by selecting File from the menu bar and clicking Exit.



Silicate Sample Run


Introduction

This document illustrates step by step how to set up a new Probe for EPMA quantitative run and how to analyze an unknown ten-element silicate sample.  This documentation was produced on a three spectrometer JEOL 733 electron microprobe. Your particular run may look very different depending on the specific configuration of your microprobe.  This document should be used in conjunction with the User’s Guide and Reference documentation, on-line help and the PFWUSERWIZARD program.

This run will demonstrate some of the powerful features of the Probe for EPMA program.  These include the use of pre-digitized standard mounts, automated spectrometer peaking, non-linear MAN (mean atomic number) background corrections, automated spectral interference corrections, automated standard acquisitions and digitizing unknown sample acquisitions.



Opening Probe for EPMA

From the Desktop, double-click on the yellow pfe-E Software folder opening the EPMA Software group.  Double click on the Probe for EPMA icon.



Upon launching PROBEWIN (Probe for EPMA), the main log window appears along with the RealTimeInitInterface window as illustrated below.  To collect real time data click the Yes button.  The program can also be run off-line without the microprobe interface to re-process previously acquired data or on another computer.



This action causes the Confirm Motor and Crystal Positions dialog box to open.  Confirm that all of the motors (stage and spectrometer positions) and crystal designations are correctly calibrated.  If there is disagreement between the mechanical positions (actual) and the software values, adjust the software values.  Use the tab key to move between the Stage and Spectrometer Positions text boxes.


Click the OK button after you have finished to close the Confirm Motor and Crystal Positions dialog box.

The main Probe for EPMA log window is now visible as seen below.




Creating a New Run

To create a new sample run, select File from the menu bar and click New from the menu.


The Open New Probe Database File dialog box opens.


Change the Save in: location (directory) and type in an appropriate run name into the File name: text box.

The initial Save in: location is specified by the UserDataDirectory keyword in the PROBEWIN.INI file.  File names longer than 8 characters are now supported.


The screen capture of the first window in this section indicates that other probe runs are already established.  Any of the existing old runs maybe re-opened to acquire additional data or used as a “setup” file for starting a new run.  In this example, a new file designated PYROXENE01.MDB will be created in the Kremser directory.


Close the Open New Probe Database File window by clicking the Save button.  This action opens the File Information dialog box.

Enter the relevant information for the new run into the User, Title, and other Description text boxes.  Use the tab key to move between text boxes.  When finished, click the OK button.


This returns the program to the main Probe for EPMA log window. Now the four main Probe buttons: Acquire!, Analyze!, Automate!, and Plot! become active.




Parameter Initialization


Analytical Standard Selection

Select the analytical standards to be used in the new probe run.  From the main Probe for EPMA log window click Standard from the menu bar and select Add/Remove Standards To/From Run from the menu.



This opens the Add Standards to Run dialog box.


All previously entered standards in the default standard database are accessible.  Scroll through the Available Standards in Database list box to find the standards to be used in this run.  Select both primary analytical standards and the MAN background standards.  Some standards may be run as both.  Select each and click the Add Standard To Run >> button to add each to the Current Standards in Run list box.


Click the OK button of the Add Standards to Run window when finished selecting standards.  This returns the program to the main log window.



Creating a New Sample

Click the Acquire! button in the main Probe for EPMA log window.  This action opens the Acquire! dialog box.  Note, not all buttons are active.


Click the New Sample button of the Acquire! dialog box.


This opens the New Sample dialog box.



Select Unknown from the New Sample Type buttons.  Type an appropriate sample name and description into the New Sample Name and New Sample Description text boxes.  This first sample will be used only to establish the analysis parameters.


Click the OK button of the New Sample dialog box.


The program returns to the Acquire! window.  Notice that the first sample designated

Un   1  * template is now listed in the Current Sample text box.  The * symbol indicates that no data has been collected for this sample yet.




Setting Analytical Conditions

Click the Analytical Conditions button to open the Analytical Conditions dialog box.  Enter the appropriate numbers into the Kilovolts, Beam Current, and Beam Size text boxes for the currently Selected Sample.  The Kilovolts, Beam Current, and Beam Size will need to be manually adjusted if a column digital interface is not present (all parameters other than kilovolts are simply available for documentation purposes only.  If a hardware interface is supported, the user may specify a column condition string to indicate the desired analytical conditions of the instrument.


Click the OK button when done, returning to the Acquire! window.



Nominal Beam Current Measurement

The nominal beam current may be adjusted from the Count Times dialog box.  Here, the nominal beam current is not the actual measured beam current but a close approximation that is used to calibrate the magnitude of the beam drift correction.  If the nominal beam current is close to the actual measured beam current then the correction is close to 1.0 and the beam drift corrected counts displayed in the main log window will be close in magnitude to the counts displayed on the screen scalers.  The nominal beam can be adjusted in several ways.

Open the Count Times window from the Acquire! button.


Any value desired may be directly entered into the Nominal Beam text box (30 nA value is stored in the PROBEWIN.INI file) or the user may measure the present beam current by clicking the Nominal Beam button.


The AcquireCheckNominal dialog box appears, choose the Yes button to measure the present beam current for use in the beam drift correction.



The current value of the faraday beam is measured and reported to both the Acquire! window and the Nominal Beam text box in the Count Times window as seen below.


Close the Count Times window by clicking the OK button.



Element, X-Ray Line, and Spectrometer Parameters Selection

Next, the user specifies the elements to be analyzed.  Click the Elements/Cations button.



This action opens the Acquired and Specified Elements dialog box.  Click on the first empty row under the Element column to enter the first element to analyze.  The user may enter the analyzed elements in any order however, the analysis output will follow this order.



This opens the Element Properties dialog box.  In the Element field either type in the first element to analyze or use the drop-down menu to select the element symbol.  Certain default values listed in this window are based on parameters entered into the previously established configuration files.



Under the Enter Element Properties For: section (top of the Element Properties dialog box) choose the correct X-Ray Line, Cations, and Oxygens for the first element.  Both alpha and beta lines are now supported as well as the ability of running the same element on all relevant spectrometers.  Note: the Disable Acq and Disable Quant check boxes below the Delete button (see User’s Guide and Reference documentation for details).



There are two common methods for performing a background correction on wavelength dispersive x-ray data; off-peak backgrounds and MAN (mean atomic number) background corrections.  The off-peak method entails measuring the background on each element in the sample of interest with the spectrometer adjusted to a position, typically on each side of the analytical peak.  This method while somewhat time-consuming can accurately determine the background contribution for major, minor and trace element concentrations.  Sophisticated modeling routines are available for precisely fitting backgrounds around analytical peaks (see User’s Guide and Reference documentation for details).

The MAN method relies on the fact that most of the background (continuum) production in the sample is directly proportional to the average atomic number of the sample.  The MAN correction is an empirical calibration curve method involving the measurement of standards of known composition (hence average atomic number).  If many samples are to be analyzed for their major and minor element concentrations then substantial time may be saved using the MAN method.  However, if the user is required to measure high atomic number samples and/or trace concentrations, more accurate data may be obtained with off-peak background corrections.

Continue by selecting MAN for the Background Type.  Selecting MAN deactivates the Off Peak Correction Type buttons as well as the High and Low Off-Peak boxes.  Next, click the text box under Spectrometer and enter the appropriate spectrometer number that will be used to analyze the first element.  Choosing a Spectrometer and Crystal loads various parameters from the configuration files.  Each of these parameters in this window should be inspected and edited as needed (use the tab key to move between boxes).


The next screen shows the edited Element Properties dialog box for silicon.


Click the OK button of the Element Properties dialog box to accept these element parameters for silicon.


The program returns to the Acquired and Specified Elements window with silicon now entered into the Element/Cations Parameters table.



Enter the next element in the run by clicking on the next empty Element row of the Acquired and Specified Elements window.  This opens the Element Properties dialog box again.  Enter the appropriate Element, Spectrometer, Crystal and adjust all other text boxes and buttons.  Repeat the element entry process until all of the elements are listed in the Acquired and Specified Elements window.  The remaining nine element entries are not shown here to save space.  Finally, oxygen is added to the element list as a not analyzed element for subsequent formula calculations.  This is done by entering o (for oxygen) in the Element text box and leaving the X-Ray Line text box empty (see User’s Guide and Reference documentation for more details).


Click the OK button of the Acquired and Specified Elements window when done entering elements in the run.


The GetElmLoadDefaultStds window opens to inform the user that standard assignments have been made based on the highest concentration of the element in the standard.  The user will edit these choices shortly.


Click OK to return to the main Acquire! window.



Editing Acquisition Options

The user may change the element acquisition order of the spectrometers by clicking the Acquisition Options button in the Acquire! dialog box.



This opens the Acquisition Options dialog box.


To change the order that the spectrometer measures an element, select the User Defined Order Number button under Acquisition Order and click the row of the element to edit.


This opens the Acquisition Properties dialog box, seen below.  Here, the user will re-define sodium (Na) to be counted on the first spectrometer pass due to its susceptibility to being volatilized by long exposure to the electron beam.  In samples containing volatile elements the user may wish to consider running the volatile element calibration routine (see User’s Guide and Reference documentation and/or Advanced Topics manual).

Edit the Spectrometer Order Number for all elements to change the acquisition order.  Further, the user wishes to use the same background correction method for both standards and unknowns, edit the Background Type for Standards to MAN for each element.  Click the OK button returning to the Acquisition Options window.  For spectrometer efficiency and element volatilization issues the user redefines the acquisition order as seen below.


Click the OK button of the Acquisition Options window to return to the Acquire! dialog box.


Modifying Standard Assignments

The standard assignments chosen by Probe for EPMA may be inspected and edited by clicking the Analyze! button from the main log window.


The program automatically wraps element data output to eight elements per line.  If the extended format menu is checked (activated from the Output menu) then the data is written out (in log window and to disk file, if enabled) as far as necessary to the right.


This opens the Analyze! dialog box.


Click the Standard Assignments button.


The Standard and Interference Assignments dialog box opens.


Click the row of the element that the user wishes to change the standard assignment for.


This opens the Assignment Properties dialog box.  The default standard assignments are based on the highest concentration of the element in the standards currently in the run.  In addition to standard assignments, the user may assign spectral interference corrections and volatile element calibrations from this window.



Click the Assigned Standard menu box.  A scrollable list of all standards added to the current run are displayed.  Select a new standard for element si.


Click the OK button returning to the Standard and Interference Assignments dialog box.


Repeat these editing steps until all necessary element standard assignments have been modified.  In this example, the standard assignments for si, al, and mg are edited, resulting in the following window.


Click the OK button of the Standard and Interference Assignments dialog box returning to the Analyze! window.



Setting Count Times

Click the Count Times button of the Acquire! window.


This opens the Count Times dialog box.  Here various parameters relating to counting times can be adjusted.  Initially On-Peak count time is set for 10 seconds based on the configuration file defaults.  Note: Real time spectrometer motion and acquisition time is graphically displayed.


To edit the count times for any element click that row in the spreadsheet.  This opens the Count Time Properties dialog box.


Edit the Count Time text boxes with new times.  To adjust the count times on unknowns, change the Unknown Count Time Factor.  This is the multiplicity factor for acquiring unknown sample elements relative to the count times specified for the standards.

The Unknown Maximum Count text box is used to specify a statistics based count time.  This feature is most useful if the user wishes to count for 30 seconds or 40000 counts whichever comes first.  For samples with high count rate elements, the actual analysis time would be shorter.

Click the OK button of the Count Time Properties window.

Finally, click the OK button of the Count Times dialog box to accept any modified count times and return to the Acquire! window.



Loading Standard Position Files


To run analytical standards using automation, requires that the computer know the physical location of all the standards for this run.  Click the Automate! button from the main Probe for EPMA log window.


This opens the Automate! dialog box shown below.


The last set of digitized standards used is visible in the Position List list box of the Automate! window.  Currently, the standard block for the brass alloy run digitized previously are listed.  These will be deleted and replaced by the appropriate standard position file(s).


Click the Delete All button.  This opens the AutomateDeleteAll window.  Click the Yes button of the AutomateDeleteAll window to clear the Position List list box of all displayed position samples.


The FiducialDeleteUnreferenced window opens.  Click the Yes button to clear the fiducial coordinate set from the position database.



Click the Import from ASCII File (*.POS File) button of the Automate! dialog box to import position samples from a previously saved ASCII file.



This action opens the Open File To Import Position Data From window.  The user has previously digitized all standard blocks and created STDPOS*.POS files.  Three STDPOS*.POS files are typically loaded for silicate runs; STDPOS1_RECTA.POS, STDPOS2_TAYLOR.POS, and STDPOS3_RECTB.POS.


The default location for *.POS files is defined in the probewin.ini file, and is typically C:\User Data\PFE Position Files.  Edit the Look in: text box if necessary.


Type in the appropriate file name in the File name: text box or simply highlight the file in the list and click the Open button.


This action opens the FiducialLoad window.  Click the Yes button to do a fiducial transformation on this pre-digitized standard block to obtain an accurate set of standard positions.


The Modify Fiducial Positions window opens.  Normally the user would simply accept the defaults or edit the position text boxes for each point, including the appropriate stage location number (JEOL 733 use appropriate W stage position).  When done, click the OK button.



This action causes the stage motors to drive to the first fiducial coordinate in its lookup table.  The FiducialVerifyFiducial window appears.  Adjust the stage motors to center the first fiducial mark, click the OK button.


The computer will drive to each of the three fiducial marks for centering.  Clicking the OK button after the third fiducial mark opens the FiducialsVerifyFiducials window.  Click this OK button.


The program then imports and updates the position coordinates of all of the standards in the pre-digitized standard position file.  The AutomateImportPositions window opens.


Click the OK button returning to the Automate! window.


The Automate! window would appear as below.  The currently transformed standard position file is listed in the Position List list box.


Repeat the same loading procedure for the other two standard position files required for use in the automation.  After clicking the Import from ASCII File button, the AutomateImportFile window opens.


Typically, when using more than one standard mount, the user would not delete all positions in the Position List, instead appending the additional position files to the first file.  Select No and import additional standards.

All of the standards loaded are listed in the Position List list box of the Automate! window.  These may now be accessed by the program during any automation action.  For instance, it is now possible to have the computer drive to any standard located on the three blocks. The user may click the Move button of the Automate! window opening the Move Motors and Crystals dialog box.  Then, click the Positions button.



This opens the Position Database dialog box.  From here, any sample that has been digitized may be located by simply selecting it and clicking the Go button.


Once the stage motors drive the stage to the chosen standard, exit the Position Database by clicking the Close button.  Likewise, the user may close the Move Motors and Crystals window by clicking its Close button, returning to the Automate! window.

Note: Double clicking from any position list will also drive the stage, highlight the row number of interest and double click.

This concludes the initial parameter setup portion of Probe for EPMA.



Automation Actions


Confirm Standard Positions

All of the basic peak centering and x-ray count acquisition procedures may be automated.  This is accomplished via the Automate! window.

Click the Select Stds button of the Automate! dialog box.  All standards that have been added to the current run will now be highlighted in the Position List list box.



The user might start by checking the location and focus of each standard selected for the automated analysis.  Click the box for Confirm Standard Positions under Automation Actions. Click the Run Selected Samples button.



The AutomateConfirmSelected window opens informing the user that thirteen standards were chosen and asks if you want to run these automated samples, click Yes.


The program then sends the stage motors to the fiducial transformed coordinates for the first selected standard and opens the Confirm Positions window.  Clicking the two-way Pause/Continue button suspends the 10 second countdown (user defined in the PROBEWIN.INI file).  Adjust the stage motors (X, Y, and Z) to a new, clean analysis position.  Click the OK button of the Confirm Positions window when done, sending the stage to the next standard to confirm its position.  Again, the Confirm Positions window opens, allowing the user to pause the countdown and adjust the sample position.


After the final standard is confirmed, the AcquireStop window appears.  In this example standards on several standard blocks are located and confirmed.  Click this OK button returning to the Automate! dialog box.




Calibrate Peak Position

X-ray peaking may be automated from the Automate! window as follows.

The Select Stds button from the previous step highlighted all of the standards added to the current run.  Presently the Position List list box in the Automate! window contains both the analytical and MAN background standards for the current run.  Since x-ray peak centering is only done on the primary analytical standards, either re-select the primary analytical standards or de-select the additional MAN background standards from the Position List list box.  Under Automation Actions click only the Peak Spectrometers box.  Under Automation Options click the Peak on Assigned Standards and Use Confirm During Acquisition boxes.  Finally, click the Peaking button to open the Peak Center dialog box.



In the Peak Center dialog box, select all the elements from the Elements to Peak list box, next click on a Peak Center Method.  Click the OK button of the Peak Center window.

Click the Run Selected Samples button from the Automate! window.

The AutomateSave window opens, asking if you want to do a PRE-scan on each element.


The user selects Yes to do a PRE-scan.

This opens the AutomateConfirmSelected window.  To run these automated samples, click Yes.


The stage motors move to the position coordinates of the first standard highlighted in the Position List list box and the Confirm Positions window opens.  This window allows the user to readjust if necessary the stage motors (X, Y, and Z) to a new, clean analysis position.  Click the OK button of the Confirm Positions window when done and the spectrometers go through the peaking routine to peak center the spectrometer position to the x-ray maximum for all the elements assigned to that standard.  After finding a new peak position and reporting the results to the main log window, the stage motors move on to the coordinates of the next standard highlighted in the Position List list box.  Once situated on this standard, the spectrometers peak center those elements assigned to it.  This procedure continues until all standards are done.  When all automation action is complete, the AcquireStop window appears and requests the user to click the OK button.


The following summary of the peak center automation for the primary standards is found in the main log window.


All elements were peak centered using the Parabolic fit method.  The new peak locations (OnPeak) along with the start and stop intensities in counts per second and

peak-to-backgrounds are listed.  The final on-peak intensities (StopI) are valuable for adjusting count time parameters for your standardizations to improve statistics.



Acquire Standard Samples

The next step is to calibrate the analytical and MAN background standards in preparation for unknown samples.  The user may choose to run both types of standards together or separate them.  In the latter case, the MAN background standards would normally be acquired first since backgrounds drift less than peak intensities.

Here, the user will automate the entire acquisition of x-ray counts on all standards.  Click the Select Stds button in the Automate! dialog box.  This selects all current standards in the run, highlighting them in the Position List list box.  Next, under Automation Actions, click only on the Acquire Standard Samples box.  From the Automation Options choices select the number of Standard Points To Acquire and whether to Use Confirm During Acquisition.  In this example, four standard points are chosen along with a Standard X Increment of 10 um.  Finally, click the Run Selected Samples button.


The familiar AutomateConfirmSelected window opens again, informing the user that thirteen standards are chosen and asks if you want to run these automated samples, click Yes.


The stage moves to the coordinates of the first highlighted standard in the Position List list box.  If the Use Confirm During Acquisition box is checked then the Confirm Positions window will open.  A complete analysis is acquired on all elements in the current sample, x-rays are counted on peak only for times specified in the Count Times window.  Progress of the acquisition can be followed from within the Acquire! Window (far right).


Finally, the Faraday cup is measured.  The stage jogs 10 um in the X direction and this procedure is repeated for the number of points specified in the Standard Points To Acquire text box of the Automate! dialog box.  After completing data collection on the first standard, the stage travels to the next highlighted standard in the list box and acquires four complete analyses on that standard.  This procedure is repeated for all selected standards.  After finishing the automation schedule the AcquireStop window opens and requires the user to click the OK button thereby returning to the Automate! window.


The following log window illustrates typical on-peak x-ray count data (in cps) for the Taylor Quartz standard.


In addition to the four individual lines of count data, the AVER, SDEV, 1SIG, SERR, and %RSD are calculated.  The AVER (average) is the average intensity reading of each element column.  The SDEV (standard deviation) is the range of these results, 1SIG (one sigma) is the predicted standard deviation, and the SERR (standard error) is essentially the precision of the average.  The %RSD number is the SDEV divided by the AVER times 100.  See the User’s Guide and Reference documentation for exact equations.  The output of the raw data counts for the remaining twelve standards are not shown here to save space.



Evaluate Standard Count Data

After all the standard data is acquired it is useful to examine the raw on-peak counts to check for and delete any obviously bad data points.  Click the Analyze! button in the main Probe for EPMA log window.



This opens the Analyze! dialog box.


The Sample List list box contains the list of the standards that data has been acquired on.  To examine the raw count data acquired on any standard run under automation, first select the standard of interest and click the Data button.


The raw count data for the four automated standard analyses of the SiO2 Quartz Taylor standard are shown below.  Each individual line (49 G to 52 G) is illustrated along with the Average, Std Dev, OneSigma, Std Err, %Rel SD, Minimum and Maximum of the acquired points.  This count data is also printed to the log window.



Examine the raw count data for each standard.  If more than one sample/standard is selected for analysis, select the Pause Between Samples check box.  When this box is checked, the program will automatically pause after displaying each analysis until the user clicks the Cancel or Next (red flashing) buttons on that are located at the bottom of the log window.  If there are any bad data points, use the Delete Selected Line(s) button to flag a line of data as bad.  In the SiO2 Quartz Taylor standard, seen below, line 52 G (good) is deemed a bad data point since its cps value is low compared to the other three lines.  Click on the line number, highlighting the line.  Next click the Delete Selected Line(s) button.



This opens the SampleDeleteLines window.


Click the Yes button.  The computer will flag this line with a B (bad) and ignore this data for any subsequent calculations.

Click the Data button again to re-analyze the remaining data lines for statistical parameters.  Remember one can always undelete data lines with the Undelete Selected Line(s) button.


At this point, the user has collected all standardization data and is ready to make MAN background assignments.



Assign MAN Background Calibrations

From the main Probe for EPMA log window, select Analytical from the menu bar and click Assign MAN Fits from the menu choices.


This opens the MANLoadNewElements window.


Click the OK button.


This opens the MAN Assignment and Fit dialog box.  The second element, aluminum (al) in the list is shown below.


From this dialog box, the user may display and modify the MAN background assignments and fits used for the background correction of all elements in the current run.  The advantage of this method is that it requires only a simple calibration of the analyzing channel over a range of atomic number.  Substantial time may be saved when many samples are to be analyzed.  However, if measuring high atomic number samples and/or trace concentrations, the off-peak background correction technique is usually superior.


For each element, select standards from the Standards list box that do not contain the element itself.  In this way the measured background counts can be plotted as a function of the average atomic number (Z-bar).  Choose at least five standards per element and compute a second-order polynomial or force a straight line fit (if deemed appropriate) between background counts and MAN for each.  For further details and suggestions, see the User’s Guide and Reference documentation.  Several fits (sodium, iron, and vanadium) are illustrated respectively, below.



The vanadium plot above illustrates another effect in WDS analysis; spectral interferences.  The well-known transition metal interferences are easily visible in these types of plots.  The Kβ x-ray line for the element of atomic number Z interfers with the Kα Z-ray line of element Z+1 (Ti with V, V with Cr, Cr with Mn and Mn with Fe). Above, standard 212 is pure TiO2 with no V2O3 but an apparent vanadium x-ray signal is seen.


The 212 standard is removed and the MAN background fit updated by clicking the Update Fit button.  All of these interferences will be examined shortly.


When done adjusting individual elements, click the OK button to store the updated MAN background corrections.



Analyze Standard Samples

The user will now analyze all of the standard data re-calculating the x-ray counts to compositions in oxide weight percent.  Since the program treats all samples as unknowns, the results of the standards provide a valuable check on the quality of the analysis.

Click the Analyze button in the main Probe for EPMA log window.  This opens the Analyze! dialog box.



Under Sample List select the All Samples button.  Click the Select All button highlighting all standards.


Click the Calculation Options button in the Analyze! window.


This action opens the Calculation Options dialog box.  Under Calculations Options click the Display Results As Oxides and the radio button Calculate with Stoichiometric Oxygen.  Elemental results are always calculated and output to the log window.


Click the OK button to output data in oxide form.


Analyzing all of the data on the standards will create a large amount of output, possibly overflowing the log window buffer, depending on the value specified in the LogWindowBufferSize parameter in the PROBEWIN.INI file.  The size of the log window buffer is limited only by the amount of memory available.  Setting this parameter to 512000 bytes is roughly equivalent to 300 pages of average density text.  In some cases saving all log window output to a user specified text file for viewing with a text editor or printing to a laser printer may be best.

Select Output from the menu bar in the main log window and click Save to Disk Log.



This opens the Open File To Output Probe Data To dialog box.  The Save in: location will be the directory specified for the original file name (PYROXENE01.MDB).  All subsequent files created by the user will use this location.  Edit the File name if desired.  The default output file has the extension .OUT.  Note that the raw data is always saved in the .MDB run file for future re-calculation and /or output.  Click Save when finished.


Select the Analyze! button in the main Probe for EPMA log window, to bring forward the Analyze! dialog box.  Click the Select All button highlighting all standards again.  Then click the Analyze button.  This will analyze all selected standard data into the specified text file.


To view this data return to the main Probe for EPMA log window and select Output from the menu bar again and click View Disk Log from the menu.



This opens the file editor.  This example utilizes the Programmer’s File Editor, seen below.  A number of text file viewers may be used.  To utilize a specific editor such as Textpad or Word, edit the FileViewer keyword in the PROBEWIN.INI file.


The user may now scroll through the analyzed standards using the text editor or may direct the file data to a laser printer by selecting File from the Programmer’s File Editor menu bar and clicking on Print in the drop-down menu.


Since all elements were acquired on all standards, examination of the oxide weight percents will provide a check on the quality of the calibration.  Several of the standard compositions will be displayed.  The first example is the Orthopyroxene standard displayed in the Analyze! window below.  This is the primary standard for magnesium and silicon.  The average values for both elements show excellent agreement with the published standard database values.



The analysis of the Rutile standard reveals several interesting points; 1) the TiO2 concentration is close to the published value of 99.26 and 2) an apparent 1.6 weight percent concentration of V2O3 is found!  This sample has no vanadium, here the user sees the notorious Ti-V spectral interference. This interference overestimates the amount of V2O3 in the sample resulting in the total exceeding 100%.  This will be corrected for (shortly) using the automatic interference correction routine.


All of the data lines gathered on the standards are examined and appear close to their standard database values.  To save space they will not be reproduced here.



Spectral Interference Assignments

Probe for EPMA allows the user to select a fully quantitative correction for spectral interferences.  The program can only correct for interferences if both the interfered and interfering elements are analyzed for.  Further, data for an interference calibration standard must be acquired that contains a major concentration of the interfering element and none of the interfered element or any other elements that interfere with the interfered element.

Select the Standards button in the Sample List and click the Select All button in the Analyze! window.  Next, click the Standard Assignments button.


Clicking this button opens the Standard and Interference Assignments dialog box.


Click on the element row to edit the Interference Assignments.


The Assignment Properties dialog box opens.  Select the first interference element for this element and the corresponding standard that contains a known amount of the interfering element but none of the interfered element.


Click the OK button when finished.

The Standard and Interference Assignments window will appear as below.



Repeat these editing steps for all of the other element interferences, resulting in the following Standard and Interference Assignments window.


Click the OK button when finished returning to the Analyze! window.


Next, the user might check the analysis options that are currently assigned.  From the main Probe for EPMA log window, select Analytical from the menu bar and click Analysis Options from the menu choices.



This opens the Analysis Calculation Options window.  Check that the appropriate boxes are marked.


Click the OK button returning to the main log window.


The user then reanalyzes the standards (Analyze button in the Analyze! window), utilizing the spectral interference correction routine.  The results for the Rutile standard are dramatic; the apparent 1.6 wt% V2O3 concentration has been replaced with an average 0.02 wt% content (which is below the detection limit).


The user is ready to move on to unknown samples.



Unknown Sample Data Collection and Analysis

To collect x-ray data on an unknown sample, minimize the Analyze! window and/or bring forward the Acquire! dialog box to start a new sample.

Click the Move button on the Acquire! window to drive the stage to the coordinates of the first unknown sample.

Click the New Sample button to activate the New Sample dialog box.  Check that the Unknown button under New Sample Type is marked.  Enter an appropriate sample name and description into the New Sample Name and New Sample Description text boxes.  Finally, click the OK button.



To start acquiring x-ray counts on the first unknown sample, simply click the Start Standard or Unknown Acquisition button of the Acquire! window.



Pyroxene #164, a chromium augite, is run once to obtain representative count rate information for the adjustment of element count times using the Count Times button to improve statistics and lower detection limits.

The Unknown Count Time Factor (Factor) in the Count Times window may be modified.  This factor is a simple multiplication of the default count times (peak and backgrounds) on the standards.  Further, the graphical spectrometer motion and acquisition time bars indicate the total time for an analysis and how efficient your usage of the spectrometers are.  Here, a fourth spectrometer with an LIF crystal would substantially shorten the analytical time.



Four random spots are then acquired (a New Sample is started) on the same pyroxene.


The Start Standard or Unknown Acquisition button in the Acquire! window is clicked four times, to acquire four data points with improved count times.


Next, the Analyze! dialog box is reopened or simply brought forward.  Click the Unknowns button and select Un    3 Pyroxene #164.


Click the Calculation Options button.


This opens the Calculation Options dialog box.

Make the following changes; under Calculations Options check Display Results as Oxides, Calculate Detection Limits and Homogeneity, and Calculate with Stoichiometric Oxygen.  Under Formula and Mineral Calculations check the Calculate Formula Based On box.  Select Pyroxene and enter 6 Atoms of Oxygen in the other two text boxes.


Click the OK button closing the Calculation Options window, returning to the Analyze! dialog box.

Clicking the Analyze button calculated the results for these four points and those values are viewed below, as copied from the text editor.


  Un    3 Pyroxene #164
  TakeOff = 40.0  KiloVolt = 15.0  Beam Current = 40.0  Beam Size =   10
  (Magnification =    2000.)                    Beam Mode = Analog  Spot
  Four random spots
  Number of Data Lines:  4               Number of 'Good' Data Lines:  4
  First/Last Date-Time: 09/23/2006 11:44:50 PM to 09/24/2006 12:03:54 AM
  Average Total Oxygen:       43.823     Average Total Weight%:   99.294
  Average Calculated Oxygen:  43.823     Average Atomic Number:   12.365
  Average Excess Oxygen:        .000     Average Atomic Weight:   21.725
  Average ZAF Iteration:        3.00     Average Quant Iterate:     4.00
  Oxygen Calculated by Cation Stoichiometry and Included in the Matrix Correction

  Results in Elemental Weight Percents
  SPEC:        O
  TYPE:     CALC
  AVER:   43.823
  SDEV:     .172
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na
  BGDS:      MAN     MAN     MAN     MAN     MAN     MAN     MAN     MAN     MAN     MAN
  ABS%:   -20.88  -28.35   -3.13   -2.04   -1.42    -.52    -.82  -32.24   -3.22  -46.88
  TIME:    20.00   20.00   30.00   30.00   30.00   30.00   30.00   40.00   40.00   30.00
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na   SUM
      54  23.298   3.942    .309    .015    .584   3.667    .115  10.410  12.302    .617  98.857
      55  23.521   3.901    .290    .011    .597   3.675    .124  10.488  12.355    .647  99.504
      56  23.637   3.908    .317    .022    .567   3.695    .105  10.480  12.286    .632  99.651
      57  23.520   3.895    .266    .024    .620   3.549    .127  10.465  12.273    .626  99.165
  AVER:   23.494   3.912    .296    .018    .592   3.647    .118  10.461  12.304    .630  99.294
  SDEV:     .142    .021    .022    .006    .022    .066    .010    .035    .036    .013
  SERR:     .071    .010    .011    .003    .011    .033    .005    .017    .018    .006
  %RSD:       .6      .5     7.6    32.8     3.8     1.8     8.5      .3      .3     2.0
  STDS:      206     207     212     211     224     203     205     206     210      81
  STKF:    .2112   .2706   .5519   .5083   .6408   .4982   .4894   .1774   .3205   .0500
  STCT:   4597.1 20306.5   917.8  1665.7  2830.4  3132.6  2529.2 10126.2 10692.2  1635.7
  UNKF:    .1834   .0273   .0025   .0002   .0050   .0306   .0010   .0707   .1132   .0033
  UNCT:   3992.1  2045.4     4.1      .5    22.2   192.5     5.0  4039.4  3777.3   106.9
  UNBG:     12.8    30.0     1.1     2.2     3.5     6.9     3.8    18.7    32.2     8.6
  ZCOR:   1.2810  1.4351  1.1938  1.2087  1.1797  1.1910  1.2113  1.4786  1.0867  1.9308
  KRAW:    .8684   .1007   .0045   .0003   .0078   .0615   .0020   .3989   .3533   .0654
  PKBG:   312.32   69.09    4.80    1.23    7.27   29.06    2.31  217.21  118.38   13.42
  INT%:      .00     .00     .00     .00     .00     .00     .00     .00     .00     .00

  Results in Oxide Weight Percents
  SPEC:        O
  TYPE:     CALC
  AVER:     .000
  SDEV:     .000
  ELEM:     SiO2   Al2O3    TiO2    V2O3   Cr2O3     FeO     MnO     MgO     CaO    Na2O   SUM
      54  49.844   7.448    .516    .022    .853   4.718    .148  17.264  17.213    .831  98.857
      55  50.319   7.372    .484    .017    .873   4.727    .160  17.392  17.287    .872  99.504
      56  50.568   7.383    .528    .033    .829   4.754    .135  17.378  17.191    .851  99.651
      57  50.318   7.360    .444    .035    .906   4.566    .164  17.354  17.172    .844  99.165
  AVER:   50.262   7.391    .493    .027    .865   4.691    .152  17.347  17.216    .850  99.294
  SDEV:     .303    .040    .037    .009    .033    .085    .013    .058    .050    .017
  SERR:     .151    .020    .019    .004    .016    .042    .006    .029    .025    .009
  %RSD:       .6      .5     7.6    32.8     3.8     1.8     8.5      .3      .3     2.0

  Results Based on 6 Atoms of o
  SPEC:        O
  TYPE:     CALC
  AVER:    6.000
  SDEV:     .000
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na   SUM
      54   1.827    .322    .014    .001    .025    .145    .005    .943    .676    .059  10.015
      55   1.832    .316    .013    .000    .025    .144    .005    .944    .674    .062  10.015
      56   1.836    .316    .014    .001    .024    .144    .004    .941    .669    .060  10.009
      57   1.836    .316    .012    .001    .026    .139    .005    .944    .671    .060  10.010
  AVER:    1.832    .318    .014    .001    .025    .143    .005    .943    .673    .060  10.012
  SDEV:     .004    .003    .001    .000    .001    .003    .000    .001    .003    .001
  SERR:     .002    .001    .001    .000    .000    .001    .000    .001    .002    .001
  %RSD:       .2      .9     7.6    32.7     3.9     1.8     8.6      .2      .5     1.8

  Pyroxene Mineral End-Member Calculations
              Wo      En      Fs
      54    38.3    53.5     8.2
      55    38.3    53.6     8.2
      56    38.1    53.6     8.2
      57    38.3    53.8     7.9
  AVER:     38.2    53.6     8.1
  SDEV:       .1      .1      .1

  Detection limit at 99 % Confidence in Elemental Weight Percent (Single Line):
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na
      54    .014    .007    .041    .030    .028    .027    .025    .005    .009    .009
      55    .014    .007    .041    .030    .028    .027    .025    .005    .009    .009
      56    .014    .007    .041    .030    .028    .027    .025    .005    .009    .009
      57    .014    .007    .041    .030    .028    .027    .025    .005    .009    .009
  AVER:     .014    .007    .041    .030    .028    .027    .025    .005    .009    .009
  SDEV:     .000    .000    .000    .000    .000    .000    .000    .000    .000    .000
  SERR:     .000    .000    .000    .000    .000    .000    .000    .000    .000    .000

  Percent Analytical Error (Single Line):
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na
      54      .4      .5    10.8    97.5     4.5     1.4    13.2      .3      .3     1.9
      55      .4      .5    11.2   127.2     4.4     1.4    12.4      .2      .3     1.9
      56      .4      .5    10.6    66.4     4.6     1.4    14.2      .2      .3     1.9
      57      .4      .5    11.9    62.9     4.3     1.4    12.2      .2      .3     1.9
  AVER:       .4      .5    11.1    88.5     4.5     1.4    13.0      .2      .3     1.9
  SDEV:       .0      .0      .6    30.1      .1      .0      .9      .0      .0      .0
  SERR:       .0      .0      .3    15.1      .1      .0      .5      .0      .0      .0

  Range of Homogeneity in +/- Elemental Weight Percent (Average of Sample):
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na
    60ci    .072    .010    .009    .001    .009    .031    .003    .017    .018    .006
    80ci    .121    .017    .015    .001    .016    .052    .005    .029    .030    .010
    90ci    .173    .024    .021    .001    .023    .075    .007    .042    .044    .014
    95ci    .235    .032    .028    .002    .031    .102    .009    .056    .059    .019
    99ci    .430    .059    .052    .003    .056    .187    .017    .103    .108    .035

  Test of Homogeneity at 1.0 % Precision (Average of Sample):
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na
    60ci     yes     yes      no      no      no     yes      no     yes     yes     yes
    80ci     yes     yes      no      no      no      no      no     yes     yes      no
    90ci     yes     yes      no      no      no      no      no     yes     yes      no
    95ci     yes     yes      no      no      no      no      no     yes     yes      no
    99ci      no      no      no      no      no      no      no     yes     yes      no

  Level of Homogeneity in +/- Percent (Average of Sample):
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na
    60ci      .3      .3     2.9     3.0     1.6      .9     2.4      .2      .1      .9
    80ci      .5      .4     4.9     4.9     2.7     1.4     3.9      .3      .2     1.5
    90ci      .7      .6     7.1     7.1     3.8     2.1     5.7      .4      .4     2.2
    95ci     1.0      .8     9.5     9.6     5.2     2.8     7.7      .5      .5     3.0
    99ci     1.8     1.5    17.5    17.7     9.5     5.1    14.1     1.0      .9     5.5

  Detection Limit in Elemental Weight Percent (Average of Sample):
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na
    60ci     ---     ---    .014    .004    .014     ---    .006     ---     ---    .008
    80ci     ---     ---    .023    .006    .023     ---    .011     ---     ---    .013
    90ci     ---     ---    .034    .009    .033     ---    .015     ---     ---    .019
    95ci     ---     ---    .046    .012    .045     ---    .020     ---     ---    .026
    99ci     ---     ---    .084    .022    .083     ---    .038     ---     ---    .048

  Projected Detection Limits (99% CI) in Elemental Weight Percent (Average of Sample):
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na
  TIME:      .31     .31     .47     .47     .47     .47     .47     .63     .63     .47
  PROJ:      ---     ---    .670    .178    .665     ---    .301     ---     ---    .382
  TIME:      .63     .63     .94     .94     .94     .94     .94    1.25    1.25     .94
  PROJ:      ---     ---    .474    .126    .470     ---    .213     ---     ---    .270
  TIME:     1.25    1.25    1.88    1.88    1.88    1.88    1.88    2.50    2.50    1.88
  PROJ:      ---     ---    .335    .089    .332     ---    .150     ---     ---    .191
  TIME:     2.50    2.50    3.75    3.75    3.75    3.75    3.75    5.00    5.00    3.75
  PROJ:      ---     ---    .237    .063    .235     ---    .106     ---     ---    .135
  TIME:     5.00    5.00    7.50    7.50    7.50    7.50    7.50   10.00   10.00    7.50
  PROJ:      ---     ---    .168    .044    .166     ---    .075     ---     ---    .096
  TIME:    10.00   10.00   15.00   15.00   15.00   15.00   15.00   20.00   20.00   15.00
  PROJ:      ---     ---    .118    .031    .118     ---    .053     ---     ---    .068
  TIME:    20.00   20.00   30.00   30.00   30.00   30.00   30.00   40.00   40.00   30.00
  PROJ:      ---     ---    .084    .022    .083     ---    .038     ---     ---    .048
  TIME:    40.00   40.00   60.00   60.00   60.00   60.00   60.00   80.00   80.00   60.00
  PROJ:      ---     ---    .059    .016    .059     ---    .027     ---     ---    .034
  TIME:    80.00   80.00  120.00  120.00  120.00  120.00  120.00  160.00  160.00  120.00
  PROJ:      ---     ---    .042    .011    .042     ---    .019     ---     ---    .024
  TIME:   160.00  160.00  240.00  240.00  240.00  240.00  240.00  320.00  320.00  240.00
  PROJ:      ---     ---    .030    .008    .029     ---    .013     ---     ---    .017
  TIME:   320.00  320.00  480.00  480.00  480.00  480.00  480.00  640.00  640.00  480.00
  PROJ:      ---     ---    .021    .006    .021     ---    .009     ---     ---    .012
  TIME:   640.00  640.00  960.00  960.00  960.00  960.00  960.00 1280.00 1280.00  960.00
  PROJ:      ---     ---    .015    .004    .015     ---    .007     ---     ---    .008
  TIME:  1280.00 1280.00 1920.00 1920.00 1920.00 1920.00 1920.00 2560.00 2560.00 1920.00
  PROJ:      ---     ---    .010    .003    .010     ---    .005     ---     ---    .006

  Analytical Sensitivity in Elemental Weight Percent (Average of Sample):
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na
    60ci    .102    .014    .012    .001    .013    .044    .004    .025    .026    .008
    80ci    .171    .023    .021    .001    .022    .074    .007    .041    .043    .014
    90ci    .245    .034    .030    .002    .032    .107    .009    .059    .062    .020
    95ci    .332    .045    .040    .002    .043    .144    .013    .080    .083    .027
    99ci    .609    .083    .073    .005    .079    .264    .023    .146    .153    .049
  

The user may obtain a large amount of information besides elemental and oxide weight percent data; these expanded capabilities include formula and mineral end member calculations, an extended set of detection limit and statistics including homogeneity and analytical sensitivity.  See the User’s Guide and Reference documentation for calculation details.


Digitized Sample Data Collection and Analysis

Next the user will perform a digitized traverse across an unknown pyroxene grain.  The user can digitize standards, unknowns or wavescan positions based on random points, linear traverse or rectangular or polygon gridded areas.  Check that the Unknowns button is clicked in the Automate! window.


Click the Digitize button in the Automate! window.


This opens the Digitize Sample Positions dialog box.



To create an unknown digitized sample click Unknown under Sample Type and enter a sample name in the Unknown or Wavescan Position Samples text box.  Next, click the Create New Unknown or Wavescan button.  The unknown sample will now appear in the Position List list box of the Automate! window.


Finally, click the Linear Traverse button to create a traverse of digitized points.  Other options are rectangular and polygon grids as well as digitize clusters of random points.


The Linear Traverse Parameters dialog box opens.


Move to the start position of the linear traverse, and click the Update Start button.  Move to the stop position and click the Update Stop button.  The total distance is displayed.


Select the Use Number of Points Per Traverse or Use Step Size in Microns Per Step radio button and adjust the text boxes appropriately.


Click the OK button returning to the Automate! window.


Now all of the calculated analysis positions have been digitized and listed.  Under Automation Actions click the Acquire Unknown Samples button.


Click Run Selected Samples button to initiate the traverse.


The AutomateConfirmSelected window opens, click Yes.


When the traverse is completed the familiar AcquireStop window appears.


Click the OK button returning the user to the Automate! dialog box.


To analyze the data obtained from the traverse, the user opens the Analyze! window and selects the Un    4 Pyroxene Traverse unknown sample in the Sample List.


Again, save the log window output to the text editor.  Click the Analyze button to calculate compositions and finally view the disk log in the text editor.  A portion is shown below.


  Un    4 Pyroxene Traverse
  TakeOff = 40.0  KiloVolt = 15.0  Beam Current = 40.0  Beam Size =   10
  (Magnification =    2000.)                    Beam Mode = Analog  Spot
  Number of Data Lines: 10               Number of 'Good' Data Lines: 10
  First/Last Date-Time: 09/24/2006 12:16:49 AM to 09/24/2006 12:50:11 AM
  Average Total Oxygen:       43.864     Average Total Weight%:   99.347
  Average Calculated Oxygen:  43.864     Average Atomic Number:   12.362
  Average Excess Oxygen:        .000     Average Atomic Weight:   21.721
  Average ZAF Iteration:        3.00     Average Quant Iterate:     3.90
  Oxygen Calculated by Cation Stoichiometry and Included in the Matrix Correction

  Results in Elemental Weight Percents
  SPEC:        O
  TYPE:     CALC
  AVER:   43.864
  SDEV:     .186
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na
  BGDS:      MAN     MAN     MAN     MAN     MAN     MAN     MAN     MAN     MAN     MAN
  ABS%:   -20.86  -28.33   -3.13   -2.04   -1.42    -.52    -.82  -32.23   -3.22  -46.87
  TIME:    20.00   20.00   30.00   30.00   30.00   30.00   30.00   40.00   40.00   30.00
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na   SUM
      58  23.592   3.885    .278    .032    .581   3.656    .133  10.432  12.290    .645  99.412
      59  23.527   3.844    .345    .016    .577   3.666    .125  10.405  12.209    .638  99.115
      60  23.579   3.894    .343    .029    .597   3.731    .133  10.398  12.226    .635  99.465
      61  23.437   3.893    .248    .031    .544   3.558    .120  10.457  12.365    .631  98.975
      62  23.282   3.861    .314    .004    .571   3.639    .120  10.437  12.221    .634  98.569
      63  23.631   3.916    .309    .031    .579   3.632    .141  10.496  12.309    .642  99.710
      64  23.641   3.888    .288    .019    .545   3.648    .100  10.402  12.261    .643  99.322
      65  23.670   3.889    .336    .030    .581   3.652    .145  10.473  12.240    .631  99.667
      66  23.701   3.923    .360    .034    .534   3.649    .110  10.519  12.298    .659  99.924
      67  23.535   3.894    .290    .050    .548   3.619    .095  10.465  12.334    .634  99.310
  AVER:   23.559   3.889    .311    .028    .566   3.645    .122  10.448  12.275    .639  99.347
  SDEV:     .124    .023    .035    .012    .021    .043    .017    .041    .052    .008
  SERR:     .039    .007    .011    .004    .007    .014    .005    .013    .017    .003
  %RSD:       .5      .6    11.4    44.1     3.7     1.2    13.6      .4      .4     1.3
  STDS:      206     207     212     211     224     203     205     206     210      81
  STKF:    .2112   .2706   .5519   .5083   .6408   .4982   .4894   .1774   .3205   .0500
  STCT:   4597.1 20306.5   917.8  1665.7  2830.4  3132.6  2529.2 10126.2 10692.2  1635.7
  UNKF:    .1840   .0271   .0026   .0002   .0048   .0306   .0010   .0707   .1130   .0033
  UNCT:   4004.4  2033.9     4.3      .7    21.2   192.5     5.2  4034.7  3768.2   108.4
  UNBG:     12.8    30.0     1.1     2.2     3.5     6.9     3.8    18.7    32.2     8.6
  ZCOR:   1.2806  1.4349  1.1938  1.2087  1.1797  1.1910  1.2114  1.4786  1.0867  1.9305
  KRAW:    .8711   .1002   .0047   .0004   .0075   .0614   .0021   .3984   .3524   .0663
  PKBG:   313.41   68.70    4.99    1.35    6.99   29.05    2.36  216.96  118.13   13.60
  INT%:      .00     .00     .00     .00     .00     .00     .00     .00     .00     .00

  Results in Oxide Weight Percents
  SPEC:        O
  TYPE:     CALC
  AVER:     .000
  SDEV:     .000
  ELEM:     SiO2   Al2O3    TiO2    V2O3   Cr2O3     FeO     MnO     MgO     CaO    Na2O   SUM
      58  50.471   7.340    .464    .047    .850   4.704    .172  17.299  17.196    .869  99.412
      59  50.332   7.262    .576    .024    .844   4.717    .162  17.255  17.083    .860  99.115
      60  50.443   7.357    .572    .042    .873   4.800    .172  17.243  17.106    .856  99.465
      61  50.140   7.357    .413    .046    .795   4.577    .155  17.342  17.301    .851  98.975
      62  49.809   7.296    .524    .006    .834   4.681    .156  17.308  17.100    .855  98.569
      63  50.556   7.400    .516    .046    .846   4.672    .182  17.405  17.222    .865  99.710
      64  50.576   7.346    .480    .028    .797   4.694    .129  17.249  17.155    .867  99.322
      65  50.638   7.349    .560    .044    .850   4.698    .187  17.367  17.126    .850  99.667
      66  50.704   7.413    .600    .049    .781   4.695    .142  17.444  17.208    .888  99.924
      67  50.350   7.358    .484    .073    .801   4.656    .122  17.354  17.257    .854  99.310
  AVER:   50.402   7.348    .519    .041    .827   4.689    .158  17.327  17.175    .861  99.347
  SDEV:     .266    .044    .059    .018    .031    .055    .021    .068    .073    .011
  SERR:     .084    .014    .019    .006    .010    .017    .007    .022    .023    .004
  %RSD:       .5      .6    11.4    44.1     3.7     1.2    13.6      .4      .4     1.3

  Results Based on 6 Atoms of o
  SPEC:        O
  TYPE:     CALC
  AVER:    6.000
  SDEV:     .000
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na   SUM
      58   1.837    .315    .013    .001    .024    .143    .005    .939    .671    .061  10.010
      59   1.838    .312    .016    .001    .024    .144    .005    .939    .668    .061  10.008
      60   1.836    .316    .016    .001    .025    .146    .005    .936    .667    .060  10.008
      61   1.834    .317    .011    .001    .023    .140    .005    .945    .678    .060  10.015
      62   1.830    .316    .014    .000    .024    .144    .005    .948    .673    .061  10.016
      63   1.835    .316    .014    .001    .024    .142    .006    .942    .670    .061  10.011
      64   1.841    .315    .013    .001    .023    .143    .004    .936    .669    .061  10.007
      65   1.838    .314    .015    .001    .024    .143    .006    .940    .666    .060  10.007
      66   1.835    .316    .016    .001    .022    .142    .004    .941    .667    .062  10.009
      67   1.835    .316    .013    .002    .023    .142    .004    .943    .674    .060  10.012
  AVER:    1.836    .315    .014    .001    .024    .143    .005    .941    .670    .061  10.010
  SDEV:     .003    .001    .002    .001    .001    .002    .001    .004    .004    .001
  SERR:     .001    .000    .001    .000    .000    .001    .000    .001    .001    .000
  %RSD:       .2      .4    11.2    44.1     3.8     1.1    13.5      .4      .5     1.1

  Pyroxene Mineral End-Member Calculations
              Wo      En      Fs
      58    38.3    53.6     8.2
      59    38.2    53.6     8.2
      60    38.1    53.5     8.4
      61    38.4    53.6     7.9
      62    38.1    53.7     8.1
      63    38.2    53.7     8.1
      64    38.3    53.5     8.2
      65    38.1    53.7     8.2
      66    38.1    53.8     8.1
      67    38.3    53.6     8.1
  AVER:     38.2    53.6     8.1
  SDEV:       .1      .1      .1

  Detection limit at 99 % Confidence in Elemental Weight Percent (Single Line):
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na
      58    .014    .007    .041    .030    .028    .027    .025    .005    .009    .009
      59    .014    .007    .041    .030    .028    .027    .025    .005    .009    .009
      60    .014    .007    .041    .030    .028    .027    .025    .005    .009    .009
      61    .014    .007    .041    .030    .028    .027    .025    .005    .009    .009
      62    .014    .007    .041    .030    .028    .027    .025    .005    .009    .009
      63    .014    .007    .041    .030    .028    .027    .025    .005    .009    .009
      64    .014    .007    .041    .030    .027    .027    .025    .005    .009    .009
      65    .014    .007    .041    .030    .027    .027    .025    .005    .009    .009
      66    .014    .007    .041    .030    .027    .027    .025    .005    .009    .009
      67    .014    .007    .041    .030    .028    .027    .025    .005    .009    .009
  AVER:     .014    .007    .041    .030    .028    .027    .025    .005    .009    .009
  SDEV:     .000    .000    .000    .000    .000    .000    .000    .000    .000    .000
  SERR:     .000    .000    .000    .000    .000    .000    .000    .000    .000    .000

  Percent Analytical Error (Single Line):
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na
      58      .4      .5    11.6    47.5     4.5     1.4    11.7      .3      .3     1.9
      59      .4      .5    10.0    89.7     4.5     1.4    12.3      .3      .3     1.9
      60      .4      .5    10.1    53.2     4.4     1.3    11.7      .3      .3     1.9
      61      .4      .5    12.6    48.9     4.7     1.4    12.8      .2      .3     1.9
      62      .4      .5    10.7   349.9     4.6     1.4    12.7      .3      .3     1.9
      63      .4      .5    10.8    49.1     4.5     1.4    11.2      .2      .3     1.9
      64      .4      .5    11.3    78.2     4.7     1.4    14.7      .3      .3     1.9
      65      .4      .5    10.2    51.0     4.5     1.4    11.0      .2      .3     1.9
      66      .4      .5     9.8    45.7     4.7     1.4    13.6      .2      .3     1.9
      67      .4      .5    11.3    32.3     4.7     1.4    15.4      .2      .3     1.9
  AVER:       .4      .5    10.8    84.6     4.6     1.4    12.7      .2      .3     1.9
  SDEV:       .0      .0      .9    94.7      .1      .0     1.5      .0      .0      .0
  SERR:       .0      .0      .3    30.0      .0      .0      .5      .0      .0      .0

  Range of Homogeneity in +/- Elemental Weight Percent (Average of Sample):
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na
    60ci    .036    .007    .008    .001    .005    .012    .003    .012    .014    .002
    80ci    .056    .010    .012    .001    .008    .018    .004    .019    .022    .003
    90ci    .074    .014    .016    .002    .011    .024    .006    .025    .029    .005
    95ci    .091    .017    .020    .002    .013    .030    .007    .031    .036    .006
    99ci    .131    .024    .029    .003    .019    .043    .010    .045    .052    .008

  Test of Homogeneity at 1.0 % Precision (Average of Sample):
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na
    60ci     yes     yes      no      no     yes     yes      no     yes     yes     yes
    80ci     yes     yes      no      no      no     yes      no     yes     yes     yes
    90ci     yes     yes      no      no      no     yes      no     yes     yes     yes
    95ci     yes     yes      no      no      no     yes      no     yes     yes     yes
    99ci     yes     yes      no      no      no      no      no     yes     yes      no

  Level of Homogeneity in +/- Percent (Average of Sample):
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na
    60ci      .2      .2     2.5     3.2      .9      .3     2.2      .1      .1      .3
    80ci      .2      .3     4.0     5.0     1.4      .5     3.4      .2      .2      .5
    90ci      .3      .3     5.3     6.6     1.9      .7     4.5      .2      .2      .7
    95ci      .4      .4     6.5     8.1     2.3      .8     5.6      .3      .3      .9
    99ci      .6      .6     9.4    11.6     3.3     1.2     8.1      .4      .4     1.3

  Detection Limit in Elemental Weight Percent (Average of Sample):
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na
    60ci     ---     ---    .013    .004    .008     ---    .006     ---     ---    .003
    80ci     ---     ---    .020    .007    .012     ---    .009     ---     ---    .005
    90ci     ---     ---    .026    .009    .016     ---    .012     ---     ---    .006
    95ci     ---     ---    .032    .011    .019     ---    .015     ---     ---    .008
    99ci     ---     ---    .047    .016    .028     ---    .022     ---     ---    .011

  Projected Detection Limits (99% CI) in Elemental Weight Percent (Average of Sample):
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na
  TIME:      .31     .31     .47     .47     .47     .47     .47     .63     .63     .47
  PROJ:      ---     ---    .372    .129    .224     ---    .176     ---     ---    .089
  TIME:      .63     .63     .94     .94     .94     .94     .94    1.25    1.25     .94
  PROJ:      ---     ---    .263    .091    .158     ---    .125     ---     ---    .063
  TIME:     1.25    1.25    1.88    1.88    1.88    1.88    1.88    2.50    2.50    1.88
  PROJ:      ---     ---    .186    .064    .112     ---    .088     ---     ---    .045
  TIME:     2.50    2.50    3.75    3.75    3.75    3.75    3.75    5.00    5.00    3.75
  PROJ:      ---     ---    .132    .045    .079     ---    .062     ---     ---    .032
  TIME:     5.00    5.00    7.50    7.50    7.50    7.50    7.50   10.00   10.00    7.50
  PROJ:      ---     ---    .093    .032    .056     ---    .044     ---     ---    .022
  TIME:    10.00   10.00   15.00   15.00   15.00   15.00   15.00   20.00   20.00   15.00
  PROJ:      ---     ---    .066    .023    .040     ---    .031     ---     ---    .016
  TIME:    20.00   20.00   30.00   30.00   30.00   30.00   30.00   40.00   40.00   30.00
  PROJ:      ---     ---    .047    .016    .028     ---    .022     ---     ---    .011
  TIME:    40.00   40.00   60.00   60.00   60.00   60.00   60.00   80.00   80.00   60.00
  PROJ:      ---     ---    .033    .011    .020     ---    .016     ---     ---    .008
  TIME:    80.00   80.00  120.00  120.00  120.00  120.00  120.00  160.00  160.00  120.00
  PROJ:      ---     ---    .023    .008    .014     ---    .011     ---     ---    .006
  TIME:   160.00  160.00  240.00  240.00  240.00  240.00  240.00  320.00  320.00  240.00
  PROJ:      ---     ---    .016    .006    .010     ---    .008     ---     ---    .004
  TIME:   320.00  320.00  480.00  480.00  480.00  480.00  480.00  640.00  640.00  480.00
  PROJ:      ---     ---    .012    .004    .007     ---    .006     ---     ---    .003
  TIME:   640.00  640.00  960.00  960.00  960.00  960.00  960.00 1280.00 1280.00  960.00
  PROJ:      ---     ---    .008    .003    .005     ---    .004     ---     ---    .002
  TIME:  1280.00 1280.00 1920.00 1920.00 1920.00 1920.00 1920.00 2560.00 2560.00 1920.00
  PROJ:      ---     ---    .006    .002    .004     ---    .003     ---     ---    .001

  Analytical Sensitivity in Elemental Weight Percent (Average of Sample):
  ELEM:       Si      Al      Ti       V      Cr      Fe      Mn      Mg      Ca      Na
    60ci    .051    .009    .011    .001    .007    .017    .004    .017    .020    .003
    80ci    .079    .014    .018    .002    .011    .026    .006    .027    .031    .005
    90ci    .105    .019    .023    .003    .015    .034    .008    .036    .041    .006
    95ci    .129    .024    .029    .003    .018    .042    .010    .044    .051    .008
    99ci    .186    .034    .041    .005    .027    .061    .014    .063    .073    .011
  


Plotting Analysis Data

The use may wish to examine the traverse data in a graphical presentation.  Click the Plot! button in the main Probe for EPMA log window.  This opens the Plot! dialog box.



First, choose the pertinent samples from the Sample List list box.  Select the required X-Axis, and Y-Axis items from the axis lists.  Choose a Graph Type and the button Send Data to Plot Window.  Finally, click the Output button.



Here the user selects the X-Stage Coordinates for the X-Axis and multi-selects the TiO2, V2O3, Cr2O3 and MnO Oxide Percents for the Y-Axis.  This graph is displayed below.  Furthermore, the weight percent concentration of any point may be read directly off the plot using the two-way Hot Hit On/Zoom On button.  Any graph maybe directly output using the Print button.




Closing the Current Run and Closing Probe for EPMA

The user ends the analysis session from the main Probe for EPMA log window.  Select File from the menu bar and click Close from the menu selections.


This opens the ProbFormCloseFile window, click Yes to close this file.



Close Probe for EPMA by selecting File from the menu bar and clicking Exit.



Advanced Topics Manual



Introduction

Probe for EPMA is a very versatile and powerful 32-bit acquisition, automation and analysis package for WDS and EDS electron microprobe analysis.  This software (Probe for EPMA) is running under the Windows NT 4.0 (service pack 4) operating system on a Pentium PC platform.

One of the strengths of Probe for EPMA is the wide variety of options and features for many different tasks that are available to the probe operator.  The aim of this manual then is to document some of the more advanced features usually skipped over in an introductory text.  And as always, the path taken to cover a feature may not be the only avenue to approach the subject.

This manual was produced on the Washington University (Earth and Planetary Sciences) JEOL 733 Superprobe equipped with three wavelength dispersive spectrometers and using Probe for EPMA in demo mode.


Element Setups

Individual element analytical configurations for a specific element, x-ray, spectrometer, and reflecting crystal may be saved to the SETUP.MDB database for use in creating new sample setups within a probe run, for use in future runs or for documentation and performance evaluation purposes.  The example below will illustrate how to create element setups from within a typical eight-element olivine routine and store them in a new SETUP.MDB database.

Open a new Probe for EPMA run in the usual manner.  From the Acquire! window, create a new unknown sample from the New Sample dialog box, then click the Elements/Cations button.  Next, enter the elements of interest into the Acquired and Specified Elements window in the usual manner.  Below is the completed Acquired and Specified Elements window after the entry of all eight elements plus oxygen.

Go through the calibration process; find new peak positions and standardize to acquire intensity data on each standard.  Normally one should save the element setup of an element that is assigned as the standard for that element.  This is done because in that case the x-ray intensity data, P/B data, PHA parameters, and other information will also be saved in the SETUP.MDB database.  This information is very useful for documentation and evaluation purposes.

After completing the calibration and standardization process, open the Analyze! window.  Choose the element setup to be stored and highlight the standard (iron in fayalite, in this example).

Click the Elements/Cations button.

This opens the Analyzed and Specified Elements window.

Click the Save Element Setup button.

The SetupOpenNewFile window will appear if no existing SETUP.MDB database exists.

Click the OK button.

The Element Setup Database opens.

Highlight the specific element to save and click the Add To Database >> button.  In this standard, the iron intensity is to be archived, select Fe kα Spectro 3 LIF (134.724).

Record number 1 has been stored as illustrated below.  Note that the standard x-ray intensity and wavescan parameter data are stored as well.

Click the Close button.  The Analyzed and Specified Elements window reappears.  Click the OK button.  The Analyze! window returns.

The other element setups from this calibrated and standardized run or other probe runs may be entered into the database in a similar manner for future use.

To recall an element setup from the SETUP.MDB database for a new sample setup follow the procedure outlined below.  Open a new PROBE FOR WINDOWS run.  This process will also be applicable if the user simply wants to add an element to an existing sample setup. This example will illustrate recalling elements from the database for the setup of a new pyroxene run.

From the Acquire! window, click the New Sample button.

This opens the New Sample window.  Edit the New Sample Name text field.

Click the Load Element Setup button in the New Sample window.

This opens the Element Setup Database.

Scroll through the list of elements (records) and find the desired element and setup using the data cursor.  Use the left, right arrows (top, center) to move through the database.  To see all the setups for a particular element, enter the element symbol into the Search Element text field and use the arrow keys as before.  To see all element setups again, simply clear the Search Element text field.  To view the most recent addition(s) to the SETUP database, click the button on the data cursor.

Here, the user browses through the records and selects the appropriate silicon (si) entry as the first element setup to load.  The output list order of elements will follow this list.

Click the << Add to Sample button to add the element setup to the current sample.

The Element Properties window for silicon appears.

Edit if required, then click the OK button to accept these values.

The silicon record is then listed in the text field under the previously defined sample name.

Continue browsing the element setup database and add all required element setups desired to the sample.

A typical pyroxene element setup list is shown below.

Click the Close button when finished.  The New Sample window reappears.

Click the OK button, returning to the Acquire! window.  Don’t forget to add oxygen as a specified element to the list for stoichiometry and other calculations.  Note specified elements cannot be saved to the SETUP.MDB database.


Sample Setups

Normally, Probe for EPMA uses the sample setup from the last unknown (or standard if there are no unknown samples) to create the next new sample setup.  Sample setups on the other hand are designed to allow the user to easily recall a previous sample setup within a current run.  This allows the user to create and re-use multiple setups comprised of different groups of elements within a single run.  In the example below, sample setups for pyroxene and olivine will be created, each with a different set of elements and conditions, that may be recalled at anytime during the current probe run.

The saving of a sample setup actually saves only a pointer to the sample selected.  All of this sample’s acquisition and calculation options, elements/cations, standard assignments, etc will be utilized when a new sample is created based on this sample setup.  However, because counting time and the associated unknown count factor is treated by Probe for EPMA as data as opposed to setup information, it is necessary (if the user wants this information to be carried over) to acquire at least one data point with the sample setup prior to saving it as a sample setup.

A new Probe for EPMA run is opened in the usual manner. Ten elements and appropriate standards for pyroxenes are loaded from the SETUP.MDB database and the STANDARD.MDB database, respectively.  Each element is then calibrated and standardized.  Count times, acquisition and calculation options are adjusted to optimize the analyses and output requirements.  And finally, a pyroxene standard is run once to update count time information.

To save the just calibrated pyroxene sample as a sample setup, start by clicking the Elements/Cations button from the Analyze! window.

The Analyzed and Specified Elements dialog box appears.

Click the Save Sample Setup button.

The Sample Setup Selection window opens.

Edit the Sample Setup Description text box as desired.

Click the << Add To Sample Setups button.

The Sample Setup Selection window appears as below.  Note: the first number (12) represents the sample’s row number and can be seen listed using the Run | List Sample Names menu.

Click the OK button, returning to the Analyzed and Specified Elements window.

Click the OK button of the Analyzed and Specified Elements window returning to the Analyze! window.

Return to the Acquire! window to create a new sample.

Click the New Sample button.

Edit the New Sample Name text field.  Here, the user will establish an olivine sample setup.

Several paths may be taken from here to load new elements for the olivine sample.  To enter an entirely new list of elements and parameters it might be easier to click the OK button and follow the Element/Cations button of the Acquire! window to the Element Properties dialog.

If (as in this example) only minor changes to a sample are required then from the New Sample window, click the Load Element Setup button.

The Element Setup Database opens.

Edit the previous pyroxene list, in this example chromium, sodium, potassium are eliminated from the list by highlighting each element and clicking the Delete from Sample button.  If additional elements are required, recall them at this time (nickel is added in this example).

After editing, the window appears as below.

Click the Close button of the Element Setup Database, returning to the New Sample window.

Click the OK button in the New Sample window, returning to the Acquire! window.

If different standard choices are required they should be added from the STANDARD.MDB database at this point.  Use the Standard | Add/Remove Standards To/From Run menu in the main Probe for EPMA log window.

Recalibrate and standardize all new elements and adjust count times, acquisition, and calculation options to optimize for olivine analysis. Run an olivine standard to obtain data and check the calibration.

From the Analyze! window, click the Elements/Cations button.

The Analyzed and Specified Elements window opens, click the Save Sample Setup button.

The Sample Setup Selection window appears. Edit the Sample Setup Description text field and click the << Add To Sample Setups button, storing the Olivine setup along with the previously stored Pyroxene setup.

Click the OK button returning to the Analyzed and Specified Elements window.

Click the OK button to go back to the Analyze! window.

Any number of sample setups can be created as described above.

The user now has two calibrated sample setups available to analyze any pyroxene or olivine in the samples supplied for microprobe analysis.  The olivine setup (last) is currently active however to recall any other sample setup, follow the steps outlined below.

Bring forward the Acquire! window.  Move to the next unknown analysis spot, in this example the user wishes to analyze several pyroxene grains.

Click the New Sample button.

The New Sample window opens.  Enter the appropriate text into the New Sample Name and New Sample Description fields.

Click the Load Sample Setup button.

This opens the Sample Setup Selection window.

Select the Pyroxene setup, highlighting it allows the operator to view the element list.

Click the OK button of the Sample Setup Selection window to load the sample setup.

The program returns to the New Sample window.  Click the OK button.

The Acquire! window reappears.

Double check your spot selection and focus and click the Start Standard or Unknown Acquisition button to initiate data acquisition.

The availability of multiple sample setups during the course of automated unknown analysis gives the user tremendous flexibility.  Upon activation of the Use Digitized Sample Setups button in the Automate! window, each unknown analysis may be based on a different sample setup that was specified when the unknown sample position was digitized.  See the User’s Guide and Reference documentation for more details.


File Setups

To load any sample setup from a previous probe run file, the file setup option is provided.  These file setups are old Probe database files that contain old sample setups and may or may not contain standardization count intensity data.

The example below will illustrate how to use the file setup option to easily import two different (an olivine and a pyroxene) sample setups into the current new probe run.  Open a new Probe for EPMA run and click the New Sample button from the Acquire! window.

The New Sample window appears.  Edit the New Sample Name text box.

Click the Load File Setup button.

The Load File Setup window opens and will list all available Probe for EPMA files that can be loaded.  The initial available Probe Run Files directory pointer is the location specified when opening a new probe database file earlier.  Move to another directory location if necessary. The last file listed in the available Probe Run Files along with the last entry in the Samples List will be shown by default.

In this example, the file STILLWATER PY-OL RUN1.MDB was newly created and the user will load both an olivine and a pyroxene sample setup into this file.

Scroll through the available Probe Run Files list and highlight the file to load from.  The last sample setup will be displayed in the Samples List and Element List text field.  Next, select the sample setup that you wish to load into the new probe run.  All of the run parameters and options for that sample setup will be loaded.  The only parameters not loaded are the nominal beam current and the volatile element assignments since they are unknown sample specific.

Click the OK button to load in this sample setup of interest.

The NewFileSetupLoadStandardData window appears next, asking whether the user wants the previous standard intensity data to be loaded as well.

Selecting Yes would load the old standard intensity data from the file setup into this new run.  Depending on the stability of your instrument, it may or may not be necessary to re-standardize some or all of the standards.  In this case, the user chooses to load the standard intensity data, selecting the Yes button.

The NewFileSetupLoadStandardData window appears.

Click the OK button.

The New Sample dialog box reappears.

Click the OK button to complete the loading of the olivine sample setup from the old probe run.

The program now returns to the fully active Acquire! window.

The user then opens the Analyze! window to save this olivine setup as a sample setup in this current probe run.

Click the Elements/Cations button, opening the Analyzed and Specified Elements window.

Click the Save Sample Setups button.

The Sample Setup Selection window opens.  Edit the Sample Setup Description text box and click the << Add To Sample Setups button, resulting in the following window.

Click the OK button, returning to the Analyzed and Specified Elements window.  Click this OK button to return to the Analyze! window.

If another, previously created sample setup is needed for this current probe run, open the New Sample window and follow the instructions of the past eight pages.

Remember to save each sample setup in the Sample Setup Selection window as described above.

Returning to the Acquire! window, the user can now employ either sample setup for probe work.


Quick Wavescans

This feature is useful if an EDS detector is not available or WDS resolution over the entire spectrometer range is required.  The program will move each spectrometer currently assigned to it’s upper limit and then continuously scan each spectrometer to it’s lower travel limit while acquiring simultaneous count data.  The count time used for the Quick Wavescan Acquisition is specified in the Count Times dialog box, opened from the Acquire! window.  The current sample setup specifies which spectrometer and reflecting crystal to use.  The program uses the spectrometer calibration of the first acquired element (order = 1) in the sample.

From an open Probe for EPMA run, containing an unknown sample and the appropriate unknown under the crosshairs, click the Special Options button from the Acquire! window.

This opens the Volatile Calibration and Quick Wavescan Samples window.  Note the default acquisition option is Normal Acquisition.

Select the Quick Wavescan Acquisition dialog button. Enter a Quick Wavescan Sample Name and Quick Scan Speed into the text fields.  The smaller the scan speed percentage the slower the spectrometer will travel per second and of course each instrument would require different settings.

Click the OK button to return to the Acquire! window.

To initiate the quick wavescan acquisition, click the Start Wavescan button in the Acquire! window.  A new wavescan sample is automatically started using the sample name just supplied.  The spectrometers move to their respective upper limits and proceed with the wavescan.  The Wavescan Acquisition window opens and real time data display is viewable.  A completed three-spectrometer Wavescan Acquisition window appears below.

The size of each graph maybe expanded (as shown below) by clicking on the relevant wavescan.

Upon completion of the quick wavescan, the data may be exported via the Export Data button to an ASCII file or examined in more detail along with KLM marker overlay capabilities from the Plot! window.  Printing of the quick wavescan is possible by selecting the Print option under the Graph Data window (see next section for a specific example).


Calibrated Multi-Element Wavescans

Another unique feature of Probe for EPMA is the ability to acquire calibrated multi-element wavescans.  This provides an easy and rapid method to scan all elements in a sample for off-peak interferences.  The example below will illustrate calibrated wavescans on a ten-element pyroxene sample and the adjustment of off-peak background positions.

Open a new Probe for EPMA run in the usual manner. Confirm motor and crystal positions in the Move Motors and Crystals window.  Click the Acquire! button and allow the program to acquire a nominal beam current (in a new run).  Click the New Sample button and create a sample using the elements of interest.  Next, re-peak the elements using either manual or automatic peaking on the appropriate standards.  This calibrates the spectrometer motors.  And finally, move to the sample to perform the calibrated wavescan.

From the Acquire! window, click the New Sample button to create a wavescan sample.

The New Sample window opens.  Select the Wavescan check button as the New Sample Type.  Edit the New Sample Name and New Sample Description text fields.

Click the OK button.

The program returns to the Acquire! window.

To modify the wavescan range and/or number of data points to be collected, click on the Peak/Scan Options button in the Acquire! window.

Select the Wave Scan Limits check button under Display: and click on the appropriate element row to edit the parameters.  The stage may also be moved (incremented) during the acquisition using the Stage Step During Peakscan/Wavescan or Peaking (X and Y Axis) check box and Increment Size (microns) text field.

The user wishes to adjust the spectrometer start and stop values for k ka, click on the row of that element.

This opens the Peak and Scan Properties window. Adjust the appropriate values.

Click the OK button of the Peak and Scan Properties window when done editing.

Then click the OK of Peak and Scan Properties to close.

Wavescan count times for each element are adjusted via the Count Times button in the Acquire! window.

Click on the appropriate element row to edit the wavescan time.  Edit the Count Time Properties dialog box and then click it’s OK button to close.

Click the OK button to close the Count Times window.

Click the Start Wavescan button in the Acquire! window to initiate the calibrated multi-element wavescan. The Wavescan Acquisition window opens.  The program will automatically start acquiring the wavescan ranges selected.  If more than one element is assigned to a given spectrometer, the program will automatically go to the next element’s wavescan range after the previous wavescan element range is completed.  The order of acquisition is defined in the Acquisition Options window.  Below illustrates a completed wavescan acquisition.

As the wavescan is acquiring data, the wavescan graph may be viewed in greater detail by clicking on the graph to toggle/expand the display size.

The top portion of the following screen capture illustrates an expanded view of the manganese x-ray wavescan on spectrometer 3.

The Position (spectrometer units), Angstroms, and Counts in any channel may be read by placing the cursor on the graph.  Selecting the Load Xray check box and clicking the graph, loads the NIST x-ray database (also seen below).  The location of the manganese K alpha x-ray lines are highlighted.

After all wavescans have been acquired on the sample, the user would typically inspect off-peak interferences and background locations by using the Plot! window.  Note that if more than 50 points were acquired in a wavescan be sure to highlight all of the “continued” samples associated with the wavescan.

Select an X-Axis parameter (normally a specific spectrometer) and a Y-Axis parameter (normally the associated wavescan counts).  The number (X) after the element in each List designates the spectrometer employed to collect the data.  Finally, click the Line check button under Graph Type.

Click the Output button to graph the wavescan.

The Graph Data window opens displaying the plotted components.  The currently selected off-peak positions for background measurements are also indicated (green).

To evaluate potential interferences select a KLM Markers option (Analyzed Elements check button, for instance) to view the KLM markers or use the Load Xray Database button.

Click and drag the mouse to Zoom in on any portion of the graph.  The Graph Data window open below illustrates this powerful feature and the identification of the small x-ray peaks (satellite lines) to the high-energy side of the main silicon x-ray peak.

The off-peak positions for background determinations for quantitative samples are adjusted with the Low and High buttons (located lower right of Graph Data window).

In the next screen capture, detailing K (1) Wavescan Counts (PET cps) versus K (1) Spectrometer, the user will note that the high background position falls on top of the signal from Fe Kα1,2 second order.  This could give an anonymously high background reading on this sample, if iron is present.  Therefore, the user chooses to move the high background slightly to a lower spectrometer position.

Click the High button, move the mouse cursor (note it appears as a cross on the graph) to an appropriate new position and click the graph.

The graph will update the new off-peak position (purple is the old position, green the new position).

Click the OK button of the Graph Data window. The GetPeakSave window appears.

Click the OK button to accept and store the new high background position for K Kα x-rays.


Polygon Gridding – Using Surfer Option

Another useful feature of Probe for EPMA is the ability to perform automated polygon gridded analyses of unknowns.  After acquiring the digitized data set, Probe for EPMA can create a script file (if the SURFER.BAS file option is selected in the Plot! window) for use with SURFER for Windows to automatically generate contour, surface and *.GRD concentration files of your data.  These *.GRD files can be imported into MicroImage for viewing in false color.  The images will be quantitatively registered during the import process so that color represents elemental or oxide concentration.

In this example an unknown and complexly exsolved pyroxene (see image below) will be gridded and digitized, then run quantitatively.  Move to the unknown grain location.


Click the Digitize button of the Automate! window.

The Digitize Sample Positions window opens.

Select the Unknown check button from the Sample Type choices.

Enter a new sample name in the Unknown or Wavescan Position Samples text field, and click the Create New Unknown or Wavescan button.

A digitized polygon area grid will now be setup on the unknown grain.

Click the Polygon Grid button at the bottom of the Digitize Sample Positions window.

The Polygon Grid Parameters window opens.

The user will outline the perimeter of the grain to be gridded.  An easy way to accomplish this is to image the grain with backscattered electrons, at any magnification, and trace around the grain boundary.  Start in one corner and on a recognizable feature, click the Add Polygon Boundary Coordinate button and then move linearly toward another feature or edge, clicking the Add Polygon Boundary Coordinate button to outline this portion of the grain.  Continue to trace line segments around the grain, clicking the Add Polygon Boundary Coordinate button to enclose another portion of the grain.  Eventually, returning to the starting point, completing the enclosure.

In this example, twenty line segments were used to enclose the grain of interest.  Each end point is listed in the Polygon Grid Stage Coordinate Boundary Points text box.  If a mistake is made or you simply wish to remove the previous boundary point, click the Remove Last Polygon Position button.

Click the Plot Polygon Boundary button to inspect the perimeter just drawn.

To start over and re-draw the perimeter outline again, click the Close button on the Polygon Plot window, click the Cancel button of the Polygon Grid Parameters window, and the click the Polygon Grid button of the Digitize Sample Positions window.

When satisfied with the outline of the grid, click the Close button of the Polygon Plot window.

Enter Grid Stage Step Sizes (in microns) for both X and Y.

Click the Calculate Number of Points in Polygon> button to determine how many data points will be digitized.  Readjust the X and Y Grid Step Sizes if necessary.  Select a method of Z determination from the two option buttons under Polygon Grid Z Interpolation Options.

The DigitizeSaveGrid window appears with the number of points in an ideal rectangular grid.

Click the Yes button to calculate the total number of points.

When the appropriate gridding parameters have been set, click the OK button, closing the Polygon Grid Parameters window.

The DigitizeSaveGrid window re-appears, click the Yes button.

The program automatically digitizes each of the number of points in the polygon and returns to the Automate! window.

Click the Plot button in the Automate! dialog box to open the Position Plot window and view the locations of all of the digitized points in this sample.  In this example, the 20 micron spacing creates too many points to be individually visible on this view.

The user may click and drag the mouse to zoom on the plot to expand the scale.

Click the Close button of the Position Plot window to return to the Automate! dialog box.

The user should proceed with calibration and standardization of the elements in the probe run and checking the accuracy of the standardization.

Then, to run the just digitized polygon grid sample from the Automate! window, highlight it in the Position List.  Under the Automation Actions, click the Acquire Unknown Samples check box.  Finally, click the Run Selected Samples button.

The AutomateConfirmSelected window opens and the user clicks the Yes button to activate the acquisition.

Note, acquisition time is now calculated.

Upon completion of the data acquisition, open the Plot! window.

Select (highlight) all of the unknown digitized points (Un6-35 in this example).  Click the Minimum Total check box to skip low points (analyses in holes, etc).  Select the 3-D check button under Graph Type.

Click the Send Data to ASCII File check button. This activates the other check boxes listed here.  Click the SURFER.BAS File check box and select an X-Axis, Y-Axis and Z-Axis (multi-select) parameters to plot.

Click the Output button.

The Open File To Save ASCII Data To window opens. Adjust the Save in: location if required.  Enter a File name: in the text field provided.

Click the Save button.

The PlotDoFile window opens, click the OK button.

Another PlotDoFile window appears.

Click the OK button to create these files.

The SW123.BAS script file created above contains the OLE code for generating contour and surface plots of the digitized probe data.

To run the SW123.BAS script file, double click on the GS Scripter32 icon in the EPMA Software folder on the desktop.  Select the File | Open menu.

The Open Document window appears.  Edit the Look in: directory to identify the location of the SW123.BAS file.

Click the Open button.

GS Scripter now details the open SW123.BAS file, of which a portion is illustrated below.

The default output mode of the script file is "TEST", which will only output the plots to the screen.  To produce output to the default printer, comment out the line OutputType$ = "TEST" by placing a single quote in front of the line and uncomment the line OutputType$ = "SURF" by removing the single quote in front of it (highlighted in next screen capture).

Click Run | Start menu to begin the automated plotting.

Basic contour and surface maps will be output to the printer.  Raw data concentration (*.GRD) files will also be created; these may be opened in SURFER for further modification and output.

An example of a basic contour map for calcium is shown below.  The perimeter of the pyroxene grain is visible.  Regions of higher calcium concentrations appear dark in this view.


The next screen capture illustrates a 3-D surface map for iron in the pyroxene.  Here, the image of iron concentration (vertical scale) has been rotated and tilted slightly.


Stage Bit Maps and Picture Snap! Feature

Unknown or standard samples loaded into the electron microprobe can present some difficulty to the user in terms of rapid and precise positioning or the location of small phases or specific areas of interest to analyze upon a large sample. On the JEOL 733 microprobe the user has several options for searching for analysis or standard locations.  An optical image (reflected or transmitted light) or a video feed of the same image is available but at only one relatively high magnification, about 400 times.  Additionally, one can search for the area of interest utilizing the secondary or backscattered detectors on the microprobe at variable magnifications (ranging as low as 40 times), but this can be time consuming. Still the entire sample may not be in one field of view upon observation in the chamber.

Another device employed to aid in feature location and rapid positioning is a gridding device that holds a sample mounted in a standard holder under a moveable grid system.  The rough coordinates of a region on the sample may be read off and used to effectively narrow the search for the analysis position.

Now, navigation around and exact positioning is easily accomplished using the stage bit map and Picture Snap! features in PROBE FOR WINDOWS.  The Stage Bit Map feature will be discussed first.  The Stage button is located in the Move Motors and Crystals window.

Clicking the Stage button opens the Stage Map window.  Two different maps are displayed below.

To select another Holder Selection image, simply select the file from the drop-down list box.  Image files (windows metafiles (*.WMF)) and coordinate limits are specified in the Standards section of the PROBEWIN.INI file.  The entire map maybe reduced or enlarged retaining scale using the x0.8 or x1.2 buttons or to re-size the Stage Bit Map window simply drag any corner of the window to the desired size and shape.  The minus and plus button (upper left) minimizes the stage bitmap selection and cursor position display.

The current position is indicated as a small red-purple circle on the map.  To move from one location to another, simply double-click on the spot you wish the stage to travel to.  The current position (X and Y stage coordinates) is displayed above the Remove Faraday/ Insert Faraday button.  Digitized positions of various samples can also be viewed by selecting the appropriate radio button.

To create stage drawing maps of your standard holders, for instance, use a vector based drawing program (Micrografx Designer or the shareware program Metafile Companion1) and the exact dimensions of your holders to build dimensionally correct drawings.  These can be exported as windows metafiles and directly loaded into the graphical stage move feature in Probe for EPMA.

1Mention of specific third party software products does not imply their endorsement.

Each stage map must be calibrated in coordinate space for accurate movement to features on the map.  Typically two diagonally located points near the edge of the map are chosen for calibration.  Initiate the calibration routine by clicking the @ button (upper right) in the Stage Map window.  The MoveStageMapCalibrate window appears.

Click the Yes button to open the Calibrate Stage Bit Map Min and Max window for calibration.

Click the top Select button, opening the Stage Select window.

Click on the unique position on the stage map to identify the stage coordinates.

These values appear next to the Select button chosen.

Click the lower Select button and repeat the process.  Click on the second position on the image.  The Calibrate Stage Bit Map Min and Max window will appear as below.

Next, enter the stage coordinate information (manually or by using the appropriate Update button) from the optical scanner or microscope coordinate system.

Click the Calculate Corners button to obtain the correct corner values to calibrate your Stage Map.  These values min and max values are entered into the Standards section of the PROBEWIN.INI file.

The new Picture Snap! feature allows the user to incorporate images of your unknown thin section or polished mounts into PROBE FOR WINDOWS to aid in navigation and the digitizing of analysis locations. Images (BMP, JPEG, GRD) taken with a flatbed scanner or other camera system can be entered into Picture Snap!, then calibrated and used for analysis.

Picture Snap! dialog can be accessed from the Digitize Sample Positions window.

Picture Snap! can also be accessed from the STAGE program by accessing the Window | Picture Snap! menu.

The main Picture Snap! window appears.

Select the File menu and open the appropriate image file.

The Open File To Input Bitmap Image From window opens.

Select the appropriate directory and file to open and click the Open button.

The image is displayed in the Picture Snap! window.

Select the Window | Calibrate menu.

The Picture Snap Calibration window appears.

Image calibration is accomplished using a two point method for rectangular mounts or a three point calibration when importing round images subject to rotation.  Click the Point #1 Calibration Pick Pixel Coordinate on Picture button  The Picture Select Point window appears, select the first unique point on the image.

The X,Y Pixel Coordinates are entered.

The values shown in the X, Y Stage Coordinates text boxes are the current stage location.  Drive the stage to the same unique location and click the Read Current Stage Coordinate button.

The stage location for the first calibration point is entered.

Repeat these steps for the second calibration point, resulting in the following window.

Finally, click the Calibrate Picture button opening the PictureSnapSaveCalibration window.

Click the OK button to save the picture calibration.

The operator can digitize analysis locations for later unattended work.  Click the Digitize button.

This opens the familiar Digitize Sample Positions window.

Create a new unknown.  Double click on the spot for the first analysis point on the just calibrated image to drive the stage to those coordinates.  Click the Random Point(s) button to digitize that sample location. Additional points maybe saved.  All analysis locations can be viewed from the Display | Unknown Position Samples menu.  All analysis locations can be acquired via the Automate! Window.


Advanced Topics Section 2


Modal Analysis

Modal analysis is a statement of the composition of a sample expressed in terms of the relative amounts of phases or minerals present. These volumetric proportions can be estimated from quantitative measurements made on the specimen by point counting analysis.  This quantitative modal analysis on unknown compositions is based on a defined set of modal phases, selected from a standard database.  Any database of standard compositions may be used to define the phases.

There are three basic steps involved in the modal analysis routine.  This procedure involves initially the acquisition of a large set of compositional data acquired using either multiple traverses or large area gridding.  It is assumed that this data set is statistically representative of the sample.  In the example illustrated below, a large representative area of a fine-grained sandstone thin section was gridded and some 324 quantitative analysis points were collected.

The second step involves the creation of an input file to load into STANDARD FOR WINDOWS for the actual modal analysis calculation.  The simplest method of generating this input file is to use the Plot! window in Probe for EPMA to output a *.DAT file of the elemental or oxide weight percent compositions to disk.

After data collection has been completed, open the Plot! window.

Highlight the compositional dataset in the Sample List.

Select the first oxide for the X-Axis and the remainder in the Y-Axis (multi-select) range.

Activate the Send Data to ASCII File (X, Y, (Z)…) check button and the ASCII File Column Labels check box.  The labels are required so that the modal analysis routine can identify the elements in the input file.

The input file can come from any source as long as the element or oxide symbols are in the first line, enclosed within double quotes, and the data is in weight percent.  The weight percent data can be in any format.  Do not include a totals column.

Click the Output button.

The Open File To Save ASCII Data To window appears. Locate the appropriate directory under Save in: and type in a File name: in the text field provided.

Click the Save button.

The PlotDoFile window appears, indicating that the data was saved.

Click the OK button.

The data saved to the *.DAT file may be viewed using an editor such as Notepad.  Here a portion of the CITYHALL.DAT file is displayed.

The third and final step involves the setup and running of the modal analysis calculation.  The modal analysis routine is located in STANDARD FOR WINDOWS.  Open the program Standard from the EPMA Software folder on the desktop.

Select (highlight) a standard database that will be used to define the modal phases.  Click the Open button to load this database.

Select Options | Modal Analysis from the menu.

The Modal Analysis window opens.

Start by defining an overall Group, click the New button under Groups.

The ModalGroupNew window opens.  Enter a descriptive name for the group of phases.

Click the OK button.  Default Group and Phase Options are loaded; these will be discussed and modified shortly.

Click the New button under Phases.

The ModalPhaseNew window opens.  Enter the first modal phase.  In this example, the sandstone is composed of mostly quartz with two minor feldspars; an alkali (sodium-potassium) phase and a plagioclase phase along with iron oxides and other trace accessory minerals.  The first modal phase is entered into the text field.

Click the OK button.

Select the Add button under Standards, opening the Add Standards to Run window.

Choose standards to define this modal phase.  These are the phase compositions that the program will use to match against the unknown point analyses.  Try to avoid over-determining the phase.  For example, when defining a sodium-potassium feldspar, select the two end-members (albite and microcline).

The Modal Analysis window would now appear as below.

Continue and enter all phases, defining the phase compositions (standards) to match.  The Alkali Feldspar entry is illustrated below.

Once all of the phases have been identified and standards defined for matching, adjust the Group and Phase Options.

The Minimum Total for Input is the rejection sum for the unknown compositions, sums below this value will not be used in the modal analysis.  Typically 90-95% are good cutoffs.

Select the Do End-Member Calculations option and check the appropriate mineral name under Phase Options to perform end-member calculations as listed.

The Normalize Concentrations For Fit option is used to specify whether the standard and unknown concentrations (above the just defined minimum input total) should be normalized to 100% before the vector fit is calculated.

The Weight Concentrations For Fit option is used to specify if the element concentrations for the standards should be weighted, based on the composition of the element in that phase.  Select this option if the major elements in a phase should have greater influence in determining the vector fit.  Leave unselected, if all concentrations, regardless of their abundance should have equal weight in the vector fit.

The Minimum Vector number (default is 4.0) is basically the tolerance for the match to a defined phase.  If a closer match is desired for one or more phases in the group, decrease the vector value for that phase.  See the User’s Guide and Reference documentation for specific details on the calculation of this vector.

Finally, under Data Files, select the appropriate Input and Output Data File locations.

Click the Start button to initiate the modal analysis calculation on each data point.

After the calculation finishes the ModalStartModal window appears, stating that the output data has been saved to the specified *.OUT file.

Click this OK button.

The modal analysis data may now be viewed in the log window in STANDARD FOR WINDOWS or simply open the newly created *.OUT file.  The output file contains the vector fit, matched phase, end-member calculation (if checked), totals column and composition of each line in the input file.  Lines that do not meet the minimum total requirement are excluded from the output, if those lines are desired either cut and paste the entire output from the main STANDARD FOR WINDOWS log window or capture the entire output by EARLIER selecting the Output | Save To Disk Log menu.

The results of the modal analysis are also tabulated and summarized.  The end summary lists the total number of analyses, the minimum total for a valid composition, number of valid points that sum above the minimum sum, the number of matched points and the percentage of points that were matched.

For each phase, the summary output then lists the phase name, the number of matches for that phase, the percent of matched points, valid points and total matched points for the matches in that phase. This is followed by the average end-member (if selected), weight percent sum and composition for that phase and the standard deviation for each element.

The last page of the just run output file is displayed below.


Line  Vector  Phase     Sum     K2O    Na2O     FeO     CaO   Al2O3   SiO2     MgO    TiO2

268     .05  Quartz  100.30     .03     .02     .10     .10     .08   99.93     .00     .04
269     .04  Quartz   99.50     .02     .01     .12     .00     .12   99.17     .00     .06
270     .00  Quartz   99.66     .00     .00     .02     .00     .00   99.60     .00     .04
271     .06  Quartz   99.97     .02     .01     .20     .00     .12   99.62     .00     .00
272  ------  ------     .21     .04     .00     .00     .02     .03     .12     .01     .00
273     .00  Quartz   99.10     .01     .00     .02     .01     .00   99.04     .00     .02
274     .02  Quartz  100.07     .00     .00     .01     .00     .19   99.87     .00     .00
275     .02  Alkali   99.50     .23   11.50     .32     .03   18.85   68.57     .00     .00
276  ------  ------   36.50    5.25     .13    1.22    1.35    6.90   21.61     .04     .00
277     .00  Quartz   99.68     .00     .00     .01     .00     .00   99.68     .00     .00
278     .04  Quartz   98.30     .07     .00     .00     .03     .15   97.96     .03     .06
279  ------  ------     .13     .00     .00     .05     .02     .01     .06     .00     .00
280     .02  Quartz   99.17     .00     .00     .13     .03     .00   99.00     .01     .00
281     .06  Alkali   98.43   15.22     .24     .07     .01   18.23   64.65     .01     .00
282     .00  Quartz   99.79     .01     .00     .03     .00     .00   99.75     .00     .00
283     .01  Plagioc  98.09     .13     .02     .03   18.91   35.54   43.35     .07     .03
284     .00  Quartz   99.75     .00     .00     .04     .00     .00   99.68     .02     .01
285     .00  Quartz   99.59     .01     .00     .04     .00     .00   99.53     .02     .00
286     .67  Quartz   92.61     .07     .01     .29     .54     .16   91.46     .02     .06
287     .00  Quartz   99.43     .00     .00     .00     .03     .03   99.32     .00     .05
288     .02  Opaques  92.35     .03     .04   88.00     .01     .12     .42     .38    3.36
289  ------  ------   31.13     .23     .01     .15    1.66    1.45   27.55     .08     .00
290     .02  Quartz   99.65     .00     .00     .13     .00     .00   99.52     .00     .00
291     .13  Plagioc  99.23     .87     .08     .03   18.34   36.21   43.59     .09     .01
292  ------  ------   13.55     .04     .27     .07     .30     .19   12.55     .11     .03
293     .03  Plagioc  99.23     .03     .30     .12   18.88   36.27   43.54     .02     .07
294  ------  ------    9.42     .45     .00     .47     .56     .07    7.72     .00     .15
295  ------  ------   35.75     .30     .01     .61     .72   11.33   22.68     .09     .00
296  ------  ------     .86     .00     .00     .01     .01     .00     .85     .00     .00
297     .00  Quartz   99.88     .01     .00     .04     .01     .03   99.78     .00     .01
298     .21  Quartz   93.18     .13     .02     .20     .07     .20   92.48     .01     .07
299     .10  Quartz   98.14     .05     .03     .14     .12     .12   97.63     .03     .02
300     .00  Quartz  100.08     .00     .00     .04     .00     .00  100.05     .00     .00
301     .01  Quartz   99.49     .01     .00     .04     .02     .00   99.34     .02     .06
302     .02  Quartz  100.11     .00     .00     .08     .04     .00   99.91     .02     .07
303     .00  Quartz  100.32     .00     .00     .02     .02     .00  100.28     .00     .01
304     .00  Quartz   99.95     .00     .00     .00     .00     .00   99.95     .00     .00
305     .00  Quartz   99.97     .00     .00     .00     .00     .00   99.92     .00     .05
306     .02  Quartz  100.39     .01     .00     .09     .01     .00  100.21     .00     .06
307  ------  ------    7.62     .79     .00     .12     .33    1.09    5.27     .01     .00
308  ------  ------   15.00     .14     .01     .22     .22     .72   13.64     .04     .01
309  ------  ------    2.08     .06     .01     .14     .23     .36    1.26     .02     .00
310     .01  Plagioc  98.90     .14     .15     .01   19.20    35.63   43.65    .06     .06
311     .00  Quartz  101.36     .02     .00     .00     .02     .00   101.32    .00     .01
312     .02  Plagioc  99.02     .27     .01     .00   19.01   35.78    43.84    .11     .00
313  ------  ------   19.98     .32     .02    1.38    1.25    4.13   12.56     .30     .01
314  ------  ------   44.17     .04     .00     .15     .26   17.47   26.15     .01     .08
315  ------  ------    6.51     .00     .01     .05     .00     .10    6.34     .00     .00
316     .01  Quartz  100.85     .01     .00     .07     .00     .00  100.76     .00     .00
317  ------  ------     .40     .00     .00     .02     .02     .02     .32     .01     .01
318     .01  Quartz  100.95     .03     .00     .05     .02     .00  100.81     .04     .01
319     .00  Quartz  101.53     .00     .00     .03     .01     .00  101.46     .00     .03
320     .56  Quartz   98.51     .06     .00     .01     .33     .71   97.31     .07     .01
321     .01  Quartz  101.04     .02     .00     .07     .00     .00  100.92     .03     .00
322     .01  Quartz  101.03     .02     .00     .05     .02     .01  100.88     .00     .06
323     .00  Quartz  100.46     .00     .00     .00     .00     .00  100.45     .00     .01
324  ------  ------     .13     .00     .01     .02     .00     .00     .08     .00     .02

Results of Modal Analysis
InputFile : C:\Probe Operators\Kremser\cityhall.dat
OutputFile : C:\Probe Obrerators\Kremser\cityhall.out
Date and Time: 12/9/2006 9:42:46 AM
Group Name   : Sandstone
Total Number of points in File :  324
Valid Number of points in File :  240
Match Number of points in File :  237
Minimum Total for Valid Points :    90.00
Percentage of Valid Points :     74.1
Percentage of Match Points :     73.1

Phase   #Match  %Total  %Valid  %Match  AvgVec
Quartz     190    58.6    79.2    80.2     .09
           Sum     K2O    Na2O     FeO     CaO   Al2O3    SiO2     MgO    TiO2
Average: 99.27     .03     .01     .06     .04     .10   98.99     .01     .03
StdDev:   2.00     .08     .03     .08     .08     .21    2.12     .03     .03
Minimum: 90.60     .00     .00     .00     .00     .00   89.53     .00     .00
Maximum:102.77     .89     .34     .50     .54    1.52  101.46     .32     .20

Phase  #Match  %Total  %Valid  %Match  AvgVec
Plagiocl   20     6.2     8.3     8.4     .05
           Sum     K2O    Na2O     FeO     CaO   Al2O3    SiO2     MgO    TiO2
Average: 99.15     .31     .05     .08   18.73   36.01   43.90     .05     .02
StdDev:    .78     .34     .07     .10     .47     .28     .40     .03     .02
Minimum: 97.58     .02     .00     .00   17.95   35.54   43.32     .00     .00
Maximum:100.25    1.31     .30     .39   19.42   36.48   44.88     .11     .07

Phase  #Match  %Total  %Valid  %Match  AvgVec
AlkaliF    21     6.5     8.8     8.9     .06
           Sum     K2O    Na2O     FeO     CaO   Al2O3    SiO2     MgO    TiO2
Average: 99.11   13.19    1.83     .07     .22   18.40   65.33     .02     .03
StdDev:    .61    5.39    4.01     .07     .19     .28    1.50     .03     .04
Minimum: 97.60     .11     .02     .00     .01   18.02   63.91     .00     .00
Maximum:100.05   15.86   11.65     .32     .71   18.85   69.24     .08     .13

Phase  #Match  %Total  %Valid  %Match  AvgVec
Opaques6     1.9     2.5     2.5     .16
          Sum     K2O    Na2O     FeO     CaO   Al2O3    SiO2     MgO    TiO2
Average:91.69     .09     .05   88.86     .18     .39     .94     .43     .75
StdDev:  1.10     .07     .04     .64     .22     .46     .43     .33    1.31
Minimum:90.33     .01     .02   88.00     .00     .12     .42     .22     .03
Maximum:93.13     .18     .11   89.56     .59    1.29    1.38    1.09    3.36

Click the Close button on the Modal Analysis window.

Finish by exiting STANDARD FOR WINDOWS.


Deadtime Calculations

This section describes how to calibrate the deadtime constants for each spectrometer and where to enter them so that PROBE FOR WINDOWS will utilize these factors.

Deadtime (τ) is defined as the time interval (after arrival of a pulse) when the counting system does not respond to additional incoming pulses (Reed, 1993).  The equation normally used to correct for deadtime losses is given as:

(1)

Where: n is the deadtime corrected count rate in counts per second

n' is the measured count rate in counts per second

τ is the deadtime constant in seconds

The time interval when the counting system is dead to additional pulses is defined as τn'.  The live time then, is (1-τn').  The true count rate (n) is proportional to the beam current (i) by a constant factor, designated k.  Thus, equation (1) may be rewritten as:

(2)

A plot of n' / i (cps/nA) versus n' (cps) will yield a straight line with slope of (-kτ). The intercept on the n' / i axis will be the constant, k, and thus the deadtime factor (τ) may be determined.

A second deadtime correction option is also available in Probe for EPMA.  This is a high precision expression for use with very high count rates (Willis, 1993).  This expression differs from the normal equation only when very high count rates (>50K cps) are achieved.  The precision deadtime expression is:

    (3)

The deadtime correction option and type is selected from the Analysis Calculation Options window.  Click Analytical | Analysis Options menu from the main Probe for EPMA log window.  Click the OK button to confirm the selections.

STARTWIN can be used to obtain the x-ray intensities required for the deadtime calculation.  The procedure involves collecting precise beam current and count rate data over a wide range of beam currents. This data set can then be loaded into the supplied Excel template to automatically calculate the deadtime factor for your spectrometers.  Paul Carpenter has put together an excellent but slightly more elaborate Excel template, contact Probe Software, Inc. for further details on obtaining his spreadsheet and related documentation.

To calibrate the deadtime factors for your WDS system use high purity, homogeneous metal standards.  Depending on the microprobe configuration one standard may be employed to collect data on all spectrometers.  Here, a silicon metal standard will be used.

Open the Count Times window and disable both the Use Beam Drift Correction and the Normalize To Counts Per Second options to allow raw intensity data to be collected.  Set an On Peak Count Time that will give a precise measurement of intensities.

Peak each spectrometer to the x-ray line that will be used (Si Kα on TAP and Ti Kα on PET crystals for the JEOL).  Upon completion of the peak center routine, move the spectrometers to the new peak positions.

Prior to collecting data, run PHA scans on each spectrometer for Si Kα, check the pulse height distribution at low and very high beam currents (ideally duplicating the range of beam currents for the deadtime measurements).  At very high count rates (large beam currents), significant pulse pileup and gain shifts do occur. Fully open your pulse height windows, optimize your gain settings to see all the signal over the range of beam currents employed.

Data collection and analysis is straightforward.  Select Output | Open Link To Excel menu from the main STARTWIN log window. Collect three replicate intensity measurements and beam current data.  Each time count rate data is acquired, it will automatically be sent to an Excel spreadsheet along with column labels.  Measure the replicate count intensities at ten different beam currents; ranging from a few nanoamps to several hundred nanoamps.

Create a count time column, prior to the beam current column in the Excel raw data spreadsheet and enter the relevant count times (in this example 10 seconds was used).  The resulting spreadsheet may look similar to the one printed below except you may have data from more than three spectrometers.

Time	Beam	1	2	3
10	3.259	22083	79765	19470
10	3.263	22069	79995	19590
10	3.247	21755	80091	19679
10	7.08	42503	154130	37952
10	7.06	42642	154665	38087
10	7.072	42168	154182	38282
10	14.547	83163	293002	74572
10	14.539	83315	292602	74281
10	14.543	82998	293326	74636
10	29.911	162904	547744	147492
10	29.917	164053	549493	147209
10	29.935	163684	548386	147078
10	50.539	266841	838976	240665
10	50.562	266672	837625	240860
10	50.58	267751	837948	240463
10	80.844	409290	1169741	370608
10	80.856	410142	1169175	370821
10	80.933	409098	1168965	370368
10	103.225	507351	1352944	460976
10	103.229	507971	1354039	460353
10	103.235	508408	1352991	460165
10	153.811	709360	1640458	647802
10	153.828	711086	1640015	647158
10	153.871	710654	1641034	648664
10	199.604	873187	1790253	800511
10	199.545	871582	1788206	799268
10	199.402	873093	1788780	799117
10	248.192	1027320	1878180	945455
10	248.884	1027517	1878519	945061
10	248.947	1026681	1878716	945783

Open the DEADTIME_CALC.XLS file.  Copy and paste count times, beam current information and counts for the first spectrometer into the raw data template starting in cell A26.

By placing data into this template, the program will automatically calculate the following items: the average of three replicate time counts, the average of three replicate beam current measurements, the %RSD on the average beam current, the average of three replicate raw intensity measurements and the %RSD on the average raw intensity measurement.

Next, the counts per second (x-axis) and the counts per second per nanoamp (y-axis) are determined.  A least squares method is then used to calculate a straight line that best fits your data.  The slope and Y-intercept are reported for a straight line fit to all 10 data pairs and also for the last 6.  The latter being a more precise determination of deadtime.

Below is the calculation portion of the Excel template.

The following screen capture illustrates the raw data template.

Calculate the deadtime factor for each spectrometer in turn, by overwriting the last column of count data in the raw data template portion of the Excel spreadsheet.  Simply highlight the data in the Excel linked spreadsheet (from STARTWIN), use the copy function and paste it into the appropriate column.  Edit cells A2 and A24 to update the title of the spreadsheet, for documentation and printout purposes.  Calculations on the new data set will be automatically updated and output.

The deadtime constants are placed into the SCALARS.DAT file (line 13).  Enter a value for each spectrometer (units of microseconds, as output from the Excel spreadsheet).

Deadtime may not be a constant and probably varies with the line energy of the x-ray being measured.  One way to get around this is to place a pulse stretching circuit before the counter timer board to ensure that a forced deadtime is used to mask the actual deadtime range of the spectrometer.  A pulse width (from the pulse stretcher) greater than the worse case deadtime found for the spectrometer is produced.  Using this value will lead to a more accurate deadtime correction at all energies.


Calculation Options

Prior to analyzing collected x-ray data, the user may wish to specify various output calculation options.  These choices may be found by clicking the Calculation Options button in the Analyze! window.

The Calculation Options window opens.

Each of the selected options in the above window will be briefly discussed in conjunction with the data output for the selected sample Un 5 Gore Mt. Garnet Standard.

Example Output Gore Mt. Garnet

Un    5 Gore Mt. Garnet Standard
TakeOff = 40.0  KiloVolt = 15.0  Beam Current = 30.0  Beam Size =    1
(Magnification =        .)                    Beam Mode = Analog  Spot

Std 40
Number of Data Lines:  2               Number of 'Good' Data Lines:  2
First/Last Date-Time: 11/21/1996 10:51:22 AM to 11/21/1996 10:55:35 AM

Average Total Oxygen:       42.298     Average Total Weight%:  100.752
Average Calculated Oxygen:  42.298     Average Atomic Number:   13.652
Average Excess Oxygen:        .000     Average Atomic Weight:   22.871
Average ZAF Iteration:        4.00     Average Quant Iterate:     2.00

Oxygen Calculated by Cation Stoichiometry and Included in the Matrix Correction

Results in Elemental Weight Percents

SPEC:        O
TYPE:     CALC

AVER:   42.298
SDEV:     .027

ELEM:       Mg      Mn      Ca      Al      Fe      Ti      Si
BGDS:      LIN     LIN     LIN     LIN     LIN     LIN     LIN
TIME:    40.00   40.00   40.00   40.00   40.00   40.00   20.00

ELEM:       Mg      Mn      Ca      Al      Fe      Ti      Si   SUM
    22   5.229    .418   5.345  11.932  17.024    .065  18.474 100.765
    23   5.189    .413   5.287  11.896  16.982    .067  18.588 100.740

AVER:    5.209    .415   5.316  11.914  17.003    .066  18.531 100.752
SDEV:     .028    .004    .041    .025    .029    .002    .081
SERR:     .020    .003    .029    .018    .021    .001    .057
%RSD:       .5      .9      .8      .2      .2     2.4      .4
STDS:       40      26      40      40      40      41      40

STKF:    .0330   .3086   .0499   .0841   .1468   .2898   .1370
STCT:   1818.5  1807.9  1356.0  5863.6   993.9  7261.9  2319.7

UNKF:    .0328   .0035   .0502   .0831   .1476   .0006   .1377
UNCT:   1807.9    20.8  1365.0  5792.0   999.3    14.5  2332.6
UNBG:     31.0     7.3    33.8    59.4    11.7    46.8     6.3

ZCOR:   1.5892  1.1716  1.0580  1.4339  1.1517  1.1330  1.3457
KRAW:    .9941   .0115  1.0066   .9878  1.0054   .0020  1.0055
PKBG:    59.27    3.87   41.44   98.54   86.43    1.31  375.63

Results in Oxide Weight Percents

SPEC:        O
TYPE:     CALC

AVER:     .000
SDEV:     .000

ELEM:      MgO     MnO     CaO   Al2O3     FeO    TiO2    SiO2   SUM
    22   8.671    .539   7.478  22.545  21.901    .108  39.522 100.765
    23   8.605    .533   7.398  22.478  21.848    .112  39.767 100.740

AVER:    8.638    .536   7.438  22.512  21.874    .110  39.645 100.752
SDEV:     .047    .005    .057    .048    .038    .003    .174
SERR:     .033    .003    .040    .034    .027    .002    .123
%RSD:       .5      .9      .8      .2      .2     2.4      .4

Results in Atomic Percents

SPEC:        O
TYPE:     CALC

AVER:   60.010
SDEV:     .026

ELEM:       Mg      Mn      Ca      Al      Fe      Ti      Si   SUM
    22   4.884    .173   3.027  10.040   6.921    .031  14.933 100.000
    23   4.846    .170   2.994  10.007   6.902    .032  15.021 100.000

AVER:    4.865    .172   3.011  10.023   6.911    .031  14.977 100.000
SDEV:     .027    .001    .024    .023    .013    .001    .063
SERR:     .019    .001    .017    .017    .009    .001    .044
%RSD:       .6      .9      .8      .2      .2     2.3      .4


Results Based on 24 Atoms of o

SPEC:        O
TYPE:     CALC

AVER:   24.000
SDEV:     .000

ELEM:       Mg      Mn      Ca      Al      Fe      Ti      Si   SUM
    22   1.954    .069   1.211   4.016   2.769    .012   5.974  40.005
    23   1.937    .068   1.197   4.001   2.759    .013   6.006  39.981

AVER:    1.946    .069   1.204   4.009   2.764    .012   5.990  39.993
SDEV:     .012    .001    .010    .011    .007    .000    .022
SERR:     .008    .000    .007    .008    .005    .000    .016
%RSD:       .6      .9      .8      .3      .2     2.3      .4

Garnet Mineral End-Member Calculations (Ca, Mg, Fe, Mn)

           Gro     Pyr     Alm      Sp
    22    20.2    32.6    46.1     1.2
    23    20.1    32.5    46.3     1.1

AVER:     20.1    32.5    46.2     1.1
SDEV:       .1      .0      .1      .0

Detection limit at 99 % Confidence in Elemental Weight Percent (Single Line):

ELEM:       Mg      Mn      Ca      Al      Fe      Ti      Si
    22    .012    .039    .017    .012    .043    .023    .023
    23    .012    .041    .017    .012    .044    .023    .020

AVER:     .012    .040    .017    .012    .044    .023    .021
SDEV:     .000    .001    .000    .000    .001    .000    .002
SERR:     .000    .001    .000    .000    .000    .000    .001

Percent Analytical Error (Single Line):

ELEM:       Mg      Mn      Ca      Al      Fe      Ti      Si
    22      .4     5.1      .4      .2      .5    14.6      .5
    23      .4     5.3      .4      .2      .5    14.4      .5

AVER:       .4     5.2      .4      .2      .5    14.5      .5
SDEV:       .0      .1      .0      .0      .0      .2      .0
SERR:       .0      .1      .0      .0      .0      .1      .0

Range of Homogeneity in +/- Elemental Weight Percent (Average of Sample):

ELEM:       Mg      Mn      Ca      Al      Fe      Ti      Si
  60ci    .025    .004    .035    .023    .025    .001    .077
  80ci    .055    .009    .078    .052    .056    .003    .172
  90ci    .113    .019    .159    .107    .114    .007    .353
  95ci    .228    .038    .321    .214    .230    .014    .710
  99ci   1.142    .189   1.607   1.074   1.151    .069   3.556

Test of Homogeneity at 1.0 % Precision (Average of Sample):

ELEM:       Mg      Mn      Ca      Al      Fe      Ti      Si
  60ci     yes     yes     yes     yes     yes      no     yes
  80ci      no      no      no     yes     yes      no     yes
  90ci      no      no      no     yes     yes      no      no
  95ci      no      no      no      no      no      no      no
  99ci      no      no      no      no      no      no      no


Level of Homogeneity in +/- Percent (Average of Sample):

ELEM:       Mg      Mn      Ca      Al      Fe      Ti      Si
  60ci      .5     1.0      .7      .2      .1     2.3      .4
  80ci     1.1     2.2     1.5      .4      .3     5.1      .9
  90ci     2.2     4.5     3.0      .9      .7    10.4     1.9
  95ci     4.4     9.1     6.0     1.8     1.4    21.0     3.8
  99ci    21.9    45.5    30.2     9.0     6.8   105.2    19.2

Detection Limit in Elemental Weight Percent (Average of Sample):

ELEM:       Mg      Mn      Ca      Al      Fe      Ti      Si
  60ci     ---    .008     ---     ---     ---    .008     ---
  80ci     ---    .017     ---     ---     ---    .018     ---
  90ci     ---    .035     ---     ---     ---    .037     ---
  95ci     ---    .070     ---     ---     ---    .075     ---
  99ci     ---    .349     ---     ---     ---    .374     ---

Projected Detection Limits (99% CI) in Elemental Weight Percent (Average of Sample):

ELEM:       Mg      Mn      Ca      Al      Fe      Ti      Si
TIME:      .63     .63     .63     .63     .63     .63     .31
PROJ:      ---   2.793     ---     ---     ---   2.989     ---
TIME:     1.25    1.25    1.25    1.25    1.25    1.25     .63
PROJ:      ---   1.975     ---     ---     ---   2.113     ---
TIME:     2.50    2.50    2.50    2.50    2.50    2.50    1.25
PROJ:      ---   1.396     ---     ---     ---   1.494     ---
TIME:     5.00    5.00    5.00    5.00    5.00    5.00    2.50
PROJ:      ---    .987     ---     ---     ---   1.057     ---
TIME:    10.00   10.00   10.00   10.00   10.00   10.00    5.00
PROJ:      ---    .698     ---     ---     ---    .747     ---
TIME:    20.00   20.00   20.00   20.00   20.00   20.00   10.00
PROJ:      ---    .494     ---     ---     ---    .528     ---
TIME:    40.00   40.00   40.00   40.00   40.00   40.00   20.00
PROJ:      ---    .349     ---     ---     ---    .374     ---
TIME:    80.00   80.00   80.00   80.00   80.00   80.00   40.00
PROJ:      ---    .247     ---     ---     ---    .264     ---
TIME:   160.00  160.00  160.00  160.00  160.00  160.00   80.00
PROJ:      ---    .175     ---     ---     ---    .187     ---
TIME:   320.00  320.00  320.00  320.00  320.00  320.00  160.00
PROJ:      ---    .123     ---     ---     ---    .132     ---
TIME:   640.00  640.00  640.00  640.00  640.00  640.00  320.00
PROJ:      ---    .087     ---     ---     ---    .093     ---
TIME:  1280.00 1280.00 1280.00 1280.00 1280.00 1280.00  640.00
PROJ:      ---    .062     ---     ---     ---    .066     ---
TIME:  2560.00 2560.00 2560.00 2560.00 2560.00 2560.00 1280.00
PROJ:      ---    .044     ---     ---     ---    .047     ---

Analytical Sensitivity in Elemental Weight Percent (Average of Sample):

ELEM:       Mg      Mn      Ca      Al      Fe      Ti      Si
  60ci    .035    .006    .049    .033    .035    .002    .109
  80ci    .078    .013    .110    .073    .079    .005    .243
  90ci    .160    .027    .225    .151    .161    .010    .499
  95ci    .322    .053    .454    .303    .325    .020   1.004
  99ci   1.615    .267   2.272   1.519   1.628    .098   5.029

Selecting Display Results As Oxides check box permits the user to display the results of an analysis in oxide weight percents based on the cation ratios defined for each element in the Element/Cations dialog window.  Results are also, always reported in elemental weight percents.

The Calculate with Stoichiometric Oxygen button allows the user to calculate oxygen by stoichiometry if oxygen is not an analyzed element in the routine.  If oxygen is either measured or calculated by stoichiometry and the Display Results As Oxides check box is selected, then the program will automatically calculate and report the actual excess or deficit oxygen in the analysis.  This information can be very useful in determining if the selected cation ratios are correct (iron bearing oxides, for example).

All elements to be calculated by stoichiometry, difference or formula basis must be listed in the sample setup.  Add these elements using the Elements/Cations button.  Each must be added as a “not analyzed” element; click any empty row in the element list, type in the element symbol and leave the x-ray line blank.  Analyses can also be output as atomic percents if the Calculate Atomic Percents check box is marked.  This calculation is based on the fraction of the atomic weight of each element and is normalized to a 100% total.

The Formula and Mineral Calculation fields at the base of the Calculation Options window allow the user to compute formulas based on any number of oxygens for oxide runs or any analyzed or specified element in elemental runs.  Further, olivine, feldspar, pyroxene, and two garnet end-member calculations are written into the software.  These formula calculations are based only on atomic weight and do not consider charge balance and site occupancy.  See the appendix sections in An introduction to The Rock-Forming Minerals by Deer, Howie, and Zussman (1992) for details on calculating formulas for hydrous phases.

The user may also select the Calculate Detection Limits and Homogeneity check box.  The calculation of the sample detection limits is based on the standard counts, the unknown background counts, and includes the magnitude of the ZAF correction factor. The calculation is adapted from Scott et al., (1995).  This detection limit calculation is useful in that it can be used even on inhomogenous samples and can be quoted as the detection limit in weight percent for a single analysis line with a confidence of 99% (assuming 3 standard deviations).

Where:

ZAF  is the ZAF correction factor for the sample matrix

IS is the count rate on the analytical (pure element) standard

IB is the background count rate on the unknown sample

t is the counting time on the unknown sample

After this, a rigorous calculation of the analytical error also for single analysis lines, is performed based on the peak and background count rates (Scott et al., 1995).  The results of the calculation are displayed after multiplication by a factor of 100 to give a percent analytical error of the net count rate.  This analytical error result can be compared to the percent relative standard deviation (%RSD) displayed in the analytical calculation.  The analytical error calculation is as follows:

Where:

NP  is the total peak counts

NB  is the total background counts

tP  is the peak count time

tB  is the background count time

A more comprehensive set of calculations for analytical statistics will also be performed.  These statistics are based on equations adapted from Scanning Electron Microscopy and X-Ray Microanalysis, Second Edition by Goldstein, et al., (1992).  All calculations are expressed for various confidence intervals from 60 to 99% confidence.

The calculations are based on the number of data points acquired in the sample and the measured standard deviation for each element.  This is important because although x-ray counts theoretically have a standard deviation equal to square root of the mean, the actual standard deviation is usually larger due to variability of instrument drift, x-ray focusing errors, and x-ray production.

The statistical calculations include:

The range of homogeneity in plus or minus weight percent.

The level of homogeneity in plus or minus percent of the concentration.

The trace element detection limit in weight percent.

The analytical sensitivity in weight percent.

Where:

 is the concentration to be compared with

C is the actual concentration in weight percent of the sample

Cs  is the actual concentration in weight percent of the standard

 is the Student tfor a 1-α confidence and n-1 degrees of freedom

n  is the number of data points acquired

 is the standard deviation of the measured values

is the average number of counts on the unknown

is the continuum background counts on the unknown

is the average number of counts on the standard

is the continuum background counts on the standard

The homogeneity test compares the 99% confidence range of homogeneity value with 1% of the sample concentration for each element.  If the range of homogeneity is less than 1% of the sample concentration then the sample may be considered to be homogenous within 1%.  The detection limit calculation here is intended only for use with homogenous samples since the calculation includes the actual standard deviation of the measured counts. This detection limit can, however, be quoted for the sample average and of course will improve as the number of data points acquired increases.  Note that the homogenous sample detection limit calculation are ignored for those elements which occur as minor or major concentrations (>1%).

Conversely, the analytical sensitivity calculation is ignored for elements whose concentrations are present at less than 1%.

Three other calculation options are available to the user; Element By Difference, Stoichiometry To Calculated Oxygen, and Stoichiometry To Another Element.

When the Element By Difference check box is selected, the user can include an element in the analysis to account for their affect on the other x-ray intensities.  This element must be specified in the sample setup.  Note this method causes the calculation to result in a 100% total.

The Stoichiometry To Calculated Oxygen option is often used in the analysis of carbonate or borate samples in an oxide run. This feature permits the user to analyze just the cations in the sample and have oxygen calculated by stoichiometry and another specified element (usually C in carbonates and B in borates) calculated relative to oxygen.  In the carbonate scenario (CaCO3), carbon is always in the ratio of 1 to 3 to oxygen.  If the user specifies carbon by stoichiometry relative to the stoichiometric element oxygen at 0.333 (1 divided by 3) the correct amount of both carbon and oxygen will be incorporated into the ZAF matrix correction and totals without analyzing for either element.  This method should only be used with phases where the ratio to oxygen is both fixed and known.

The following iron carbonate mineral (siderite) output illustrates oxygen calculated by cation stoichiometry with the element carbon is calculated at 0.333 atoms relative to 1.0 atom of oxygen.

Un    4 Siderite Standard
TakeOff = 40.0  KiloVolt = 15.0  Beam Current = 15.0  Beam Size =   10
(Magnification =        .)                    Beam Mode = Analog  Spot
Number of Data Lines:  3               Number of 'Good' Data Lines:  3
First/Last Date-Time: 07/26/1999 02:45:52 PM to 07/26/1999 02:51:12 PM

Average Total Oxygen:       41.449     Average Total Weight%:  100.042
Average Calculated Oxygen:  41.449     Average Atomic Number:   16.438
Average Excess Oxygen:        .000     Average Atomic Weight:   23.164
Average ZAF Iteration:        8.00     Average Quant Iterate:     2.00

Oxygen Calculated by Cation Stoichiometry and Included in the Matrix Correction
Element C is Calculated  .333 Atoms Relative To 1.0 Atom of Oxygen

Results in Elemental Weight Percents

SPEC:        C       O
TYPE:     STOI    CALC

AVER:   10.356  41.449
SDEV:     .017    .013

ELEM:       Ca      Mg      Mn      Fe
BGDS:      LIN     LIN     LIN     LIN
TIME:    20.00   30.00   40.00   40.00

ELEM:       Ca      Mg      Mn      Fe   SUM
    19    .000    .067   2.235  45.825  99.929
    20    .014    .072   2.375  46.005 100.264
    21    .013    .077   2.323  45.708  99.933

AVER:     .009    .072   2.311  45.846 100.042
SDEV:     .008    .005    .071    .150
SERR:     .004    .003    .041    .086
%RSD:     86.8     6.9     3.1      .3
STDS:      130     131     132     132

STKF:    .3826   .0853   .0202   .4131
STCT:   3040.8  1502.9    43.1   955.3

UNKF:    .0001   .0003   .0205   .4124
UNCT:       .2     6.1    43.6   953.6
UNBG:     11.0    11.0     3.0     5.6

ZCOR:    .9873  2.0734  1.1296  1.1117
KRAW:    .0001   .0041  1.0111   .9982
PKBG:     1.03    1.55   16.50  172.56

Results in Oxide Weight Percents

SPEC:      CO2       O
TYPE:     STOI    CALC

AVER:   37.946    .000
SDEV:     .062    .000

ELEM:      CaO     MgO     MnO     FeO   SUM
    19    .000    .111   2.886  58.954  99.929
    20    .019    .119   3.066  59.185 100.264
    21    .018    .128   3.000  58.803  99.933

AVER:     .012    .119   2.984  58.981 100.042
SDEV:     .011    .008    .091    .192
SERR:     .006    .005    .053    .111
%RSD:     86.8     6.9     3.1      .3

Another interesting example demonstrating this feature is nicely documented in the User’s Guide and Reference documentation (see Stoichiometry to Oxygen section).  There, several trace metals are analyzed for in a stoichiometric Al2O3 matrix without measuring aluminum or oxygen, BUT the correct amount of Al2O3 is added to the matrix correction!

The Stoichiometry To Another Element option gives the user another recalculation method similar to the Stoichiometry To Calculated Oxygen option just discussed.  Here, the user may select any other analyzed or specified element as the stoichiometric basis element.

The example below calculates CO2 on the basis of moles of CaO, rather than by stoichiometry to oxygen.

The setup is shown in the Calculation Options window below.

The resulting carbonate output is seen next.

Un    3 Dolomite Standard
TakeOff = 40.0  KiloVolt = 15.0  Beam Current = 15.0  Beam Size =   10
(Magnification =        .)                    Beam Mode = Analog  Spot
Number of Data Lines:  3               Number of 'Good' Data Lines:  3
First/Last Date-Time: 07/26/1999 02:36:45 PM to 07/26/1999 02:42:40 PM

Average Total Oxygen:       52.237     Average Total Weight%:  100.376
Average Calculated Oxygen:  52.237     Average Atomic Number:   10.882
Average Excess Oxygen:        .000     Average Atomic Weight:   18.446
Average ZAF Iteration:        3.00     Average Quant Iterate:     2.00

Oxygen Calculated by Cation Stoichiometry and Included in the Matrix Correction
Element C is Calculated  2 Atoms Relative To 1.0 Atom of Ca

Results in Elemental Weight Percents

SPEC:        C       O
TYPE:     RELA    CALC

AVER:   13.070  52.237
SDEV:     .042    .111

ELEM:       Ca      Mg      Mn      Fe
BGDS:      LIN     LIN     LIN     LIN
TIME:    20.00   30.00   40.00   40.00

ELEM:       Ca      Mg      Mn      Fe   SUM
    16  21.744  13.226    .044    .033 100.210
    17  21.881  13.159    .023    .030 100.560
    18  21.793  13.246    .000    .028 100.356

AVER:   21.806  13.210    .022    .030 100.376
SDEV:     .069    .045    .022    .003
SERR:     .040    .026    .013    .002
%RSD:       .3      .3    98.3     8.7
STDS:      130     131     132     132

STKF:    .3826   .0853   .0202   .4131
STCT:   3040.8  1502.9    43.1   955.3

UNKF:    .2045   .0847   .0002   .0003
UNCT:   1625.7  1492.4      .3      .6
UNBG:      8.6     9.0     2.1     3.4

ZCOR:   1.0661  1.5597  1.2224  1.2016
KRAW:    .5346   .9930   .0070   .0006
PKBG:   189.50  167.57    1.19    1.18

Results in Oxide Weight Percents

SPEC:      CO2       O
TYPE:     RELA    CALC

AVER:   47.890    .000
SDEV:     .152    .000

ELEM:      CaO     MgO     MnO     FeO   SUM
    16  30.425  21.932    .056    .043 100.210
    17  30.616  21.822    .030    .038 100.560
    18  30.493  21.966    .000    .036 100.356

AVER:   30.511  21.907    .029    .039 100.376
SDEV:     .097    .075    .028    .003
SERR:     .056    .043    .016    .002
%RSD:       .3      .3    98.3     8.7


Linear Calibration - Curve Method

Probe for EPMA offers a sophisticated calibration curve (multi-standard) method for correcting x-ray data.  It is based on a second order polynomial fit to multiple standard intensity data.  This option has been utilized in special situations such as the analysis of trace carbon in steels and when a suitable set of well characterized standards are available.

The example outlined below will document this calibration curve method for the specific analysis of Al, Ti and O in titanium aluminides doped with oxygen.  This data and mdb file was generously supplied from research conducted by Jim Smith at the NASA Glenn Research Center.  These low density, high strength alloys are part of an ongoing study of the transport kinetics of oxygen in these metals in conjunction with the development of superior alloys for aircraft engine gas turbine turn blades.

A series of nine titanium-aluminides (varying Ti/Al ratio) were carefully prepared, each doped with a specific concentration of oxygen, ranging from 0 to 3.21%, thereby bracketing the expected unknowns range of oxygen concentration.  Each standard alloy was analyzed by other techniques to verify the nominal compositions.  The nine standard compositions were then entered into the STANDARD.MDB database and the positions of each standard digitized in Probe for EPMA.

Aluminum, titanium and oxygen were peaked on the appropriate standards and count rate data (five spots each) were acquired on each of the nine standards.  The count rate data was then examined in the Analyze! window to ascertain the precision of the five data points on each standard, deleting any selected lines as deemed appropriate.

After standard collection, the user must select the calibration curve approach as the matrix correction method.  From the main Probe for EPMA log window, select Analytical | ZAF Selections from the menu.

The Matrix Correction Methods window opens.  Select the Calibration Curve (multi-standard) button for the Correction Method.

Click the OK button, returning to the main Probe for EPMA log window.

Next, the user will evaluate each of the three calibration curves.  Open the Plot! window.  Select all nine titanium-aluminide alloy standards to plot.  Note, the check box for Force Black and White Print was selected.  If the user wants to output and archive the graphed data, this option causes the printer to create a clean black and white print, without gray background (see discussion below).  Select a Graph Type. Check the Average Only check box to use the average value of each standard sample.

Finally, plot O Elemental Percents versus O On Peak Counts (P+B).

Click the Output button.

All of the selected standards are analyzed and reported in the main Probe for EPMA log window.

The Plot Graph Data window appears.

The Show Grid Lines check box has been marked to facilitate reading the percent and count values.  This calibration curve may be printed out by clicking the Print button next to the OK button. If the user selects the Force Black and White Print check box in the Plot! window, then the corresponding output will be a black and white print, if not, then the printer will output the above gray background.

The user may evaluate the data using the Zoom Full capabilities (click and drag mouse over region of interest on graph) to expand the scaling.  Here, in the center group, two data points clearly overlap. Placing the mouse cursor over any selected point on the graph returns the x and y values of that position (read above the Zoom Full button).

When finished, click the OK button to return to the Plot! window to next review the other calibration curves.

The output for aluminum is plotted similarly.  Select the Al Elemental Percents versus Al On Peak Counts (P+B), (the counts per second determined on peak) and click the Output button in the Plot! window.  This curve is viewed below.

The graph may be modified by selecting any of the buttons across the top to enter the Graph Control module.

Graph Control (see tabs in display below) can be used to edit numerous graph parameters.  Under the Trends tab, for instance, are the curve fitting options.  Select Curve Fit under Statistical Lines and edit the Curve Fit Type and Order as required.

Click the Apply Now or OK button to see the changes in the Graph Data window.

The modified Plot Graph Data window returns.  Here, a title, different, larger symbols, standard numbers and the second order polynomial curve have been added to the graph below.

Upon creating the previous plots, the selected standards were analyzed, each of the nine standards were reported in the main Probe for EPMA log window along with the second order polynomial fit parameters.

The following output displays the log window output for one of these: St  617 Set   5 TiAl-7 standard.

St  617 Set   5 TiAl-7
TakeOff = 52.5  KiloVolt = 15.0  Beam Current = 30.0  Beam Size =    1
(Magnification =        .)                    Beam Mode = Analog  Spot

Titanium aluminide 3.03 Oxygen
Number of Data Lines:  5               Number of 'Good' Data Lines:  5
First/Last Date-Time: 12/11/2000 04:03:01 PM to 12/11/2000 04:06:28 PM
WARNING- Using Calibration Curve Matrix Corrections

Average Total Oxygen:         .000     Average Total Weight%:  100.651
Average Calculated Oxygen:    .000     Average Atomic Number:   20.177
Average Excess Oxygen:        .000     Average Atomic Weight:   40.571

Results in Elemental Weight Percents

ELEM:       Al      Ti       O
BGDS:      LIN     LIN     LIN
TIME:    10.00   10.00   10.00

ELEM:       Al      Ti       O   SUM
   978  15.452  81.618   2.986 100.056
   979  15.680  81.966   3.114 100.761
   980  15.864  81.880   3.053 100.796
   981  15.788  81.957   3.000 100.744
   982  15.747  82.245   2.908 100.900

AVER:   15.706  81.933   3.012 100.651
SDEV:     .157    .224    .077
SERR:     .070    .100    .035
%RSD:      1.0      .3     2.6

PUBL:   15.730  81.240   3.030 100.000
%VAR:     -.15     .85    -.59
DIFF:    -.024    .693   -.018

UNCT:  13020.8 11807.2   329.4
UNBG:    215.5    29.2   116.0
PKBG:    61.44  406.68    3.84

FIT1:    .4671-363.4302  -.5152
FIT2:    .0013   .0712   .0098
FIT3:  .000000-.000003 .000003
DEV:       2.2      .9     7.1

The coefficients for the second order polynomial are listed last (Fit 1, Fit 2, Fit 3 and Dev).  The three Fit terms represent the intercept, the slope and the second order curvature factor, respectively.  The DEV term represents the total deviation (sum of the residuals) between the calculated curve and the original data.  The smaller the number, the better here.  The software prints a warning line if this correction method is active.

The analysis of unknown samples is straightforward.  Create a new sample and collect the x-ray intensity data on the unknown.  Select the Analyze! window and click Analyze.

An example is printed next.

Un   25 Sample 3-9 LowOx 0-30u 2u increments
TakeOff = 52.5  KiloVolt = 15.0  Beam Current = 30.0  Beam Size =    1
(Magnification =        .)                    Beam Mode = Analog  Spot
Number of Data Lines: 16               Number of 'Good' Data Lines: 16
First/Last Date-Time: 12/11/2000 01:27:41 PM to 12/11/2000 01:40:41 PM
WARNING- Using Calibration Curve Matrix Corrections

Average Total Oxygen:         .000     Average Total Weight%:   99.869
Average Calculated Oxygen:    .000     Average Atomic Number:   19.838
Average Excess Oxygen:        .000     Average Atomic Weight:   39.903

Results in Elemental Weight Percents

ELEM:       Al      Ti       O
BGDS:      LIN     LIN     LIN
TIME:    10.00   10.00   10.00

ELEM:       Al      Ti       O   SUM
   851  21.260  76.350   1.956  99.566
   852  21.236  76.469   2.003  99.708
   853  21.193  76.689   1.957  99.839
   854  21.128  76.174   1.941  99.244
   855  21.222  76.599   1.842  99.663
   856  21.120  76.425   1.755  99.300
   857  21.126  76.494   1.838  99.459
   858  21.125  77.135   1.767 100.027
   859  21.142  76.686   1.791  99.619
   860  21.153  76.835   1.796  99.784
   861  21.139  77.248   1.703 100.089
   862  21.256  77.439   1.685 100.379
   863  21.190  77.360   1.833 100.383
   864  21.216  77.503   1.779 100.499
   865  21.286  77.127   1.643 100.055
   866  21.264  77.541   1.482 100.287

AVER:   21.191  76.880   1.798  99.869
SDEV:     .058    .454    .133
SERR:     .015    .113    .033
%RSD:       .3      .6     7.4

UNCT:  18339.0 11021.6   222.4
UNBG:    218.2    26.6   113.8
PKBG:    85.05  417.09    2.96

FIT1:    .4718-363.4019  -.5152
FIT2:    .0013   .0712   .0098
FIT3:  .000000-.000003 .000003
DEV:       2.2      .9     7.1

The second order polynomial coefficients are always listed last (just above).  Elements calculated by difference or stoichiometry can be calculated along with calibration curve corrected elements.  And if both off-peak and MAN acquired data are present, Probe for EPMA will construct separate sets of calibration curves and fit a second order polynomial expression that is used in the iteration procedure to determine the concentration of the element.

The user may elect to run standards after completing unknown sample acquisition and then correct for any standard intensity drift.  From the main Probe for EPMA log window select Analytical and then choose Analysis Options from the drop-down menu.

The Analysis Calculation Options window appears, remember to check that the Use Automatic Drift Correction on Standard Intensities check box is marked.

In some instances, it may be useful to add to the acquired data set a zero point (off-peak elements only) to improve the polynomial fit. To include a zero point, check the box (prior to analyzing the standards) labeled Use Zero Point For Calibration Curve (Off-Peak Elements Only). This choice is also found in the Analysis Calculation Options window.


Time-Dependent Intensity Correction

Probe for EPMA gives the user two powerful methods for the correction of time-dependent intensities of elements in unknown samples.  By time-dependent we mean a signal intensity that is varying with time in a smooth fashion, and due to the processes of element migration, volatilization, or any other process that results in a time-varying intensity. The term is used to indicate a generic process with perhaps incomplete knowledge of the details of the process involved (i.e., the term volatile is typically incorrect and cannot be confirmed by EPMA measurement alone). Each correction option provides a means to automatically correct data for the time-dependent loss of x-ray intensity due to continuous electron bombardment (heating damage/charge buildup) or carbon contamination that occurs during normal electron microprobe work.  The time-dependent element extrapolation may be applied to any degradation (or enhancement) of the x-ray intensity over time.  Elements most susceptible include sodium, potassium, perhaps sulfur and bound water.  This correction is most useful for samples that are too small to utilize a defocused beam and also allows the operator to run higher than normal sample currents to improve analytical sensitivity.

Two different methods are available for time-dependent element corrections; the Self Calibration TDI Acquisition and the Assigned Calibration TDI Acquisition.  Each will be documented.

In the Self Calibration TDI Acquisition method, the program acquires the on-peak count data for the time-dependent element during the normal sample acquisition process for the unknown sample.  This method works well when samples to be analyzed have widely differing compositions as the calibration is determined with every sample acquisition.

Open Probe for EPMA, and proceed through the normal calibration and standardization routine.  Check suitable standards for accuracy, these should not be time-dependent or beam sensitive.

Move to your first unknown sample.  Open the New Sample window from the Acquire! window and create a new unknown sample.

Click the Special Options button in the Acquire! window.

The TDI Calibration and Quick Wavescan Samples window opens.  Select the Self Calibration TDI Acquisition option. The time-dependent element correction is only applied to the first element analyzed for on each spectrometer.  Enter a number into the TDI Count Time Intervals text field (up to 50 intervals may be used).  The program will use this interval number and the on-peak count time to create a calibration curve.  In this example, Na is counted for 40 seconds on peak and with an interval of 10 entered, the program will automatically count ten 4 second intervals.  Each element listed first on each spectrometer is treated in this way.  The off peak counts are not affected.

Click the OK button to return to the Acquire! window.

Click the Start Standard or Unknown Acquisition button to initiate count data collection on the unknown sample.

After collecting an unknown sample the user may display the time-dependent correction from the Analyze! window.

Click the Standard Assignments button.

The Standard and Interference Assignments window opens.

Select the element row (sodium in this example) to view the time-dependent calibration data.

The Assignment Properties window opens and lists all possible time-dependent element calibration samples.

Highlight the appropriate TDI Element Calibration Sample Assignment sample and click the Display TDI Fit button.

The TDI (Self) Fit Plot window opens.

The drop in sodium count intensity with time appears to fit an exponential function (Nielson and Sigurdsson, 1981).  A plot of the natural log of the intensity data versus time should yield a straight line relationship as seen above.

The extrapolation correction is quantitative in that the correction is based on a self calibration curve acquired during the run.  The correction uses the actual elapsed time for all calculations.  It is applied during the ZAF or Phi-Rho-Z iteration phase of the analysis to correct for changes in the matrix correction due to the extrapolation correction.

The next example will illustrate the Assigned Calibration TDI Acquisition method.  Here, small rhyolitic glass inclusions of similar composition will be analyzed.

Start Probe for EPMA in the normal manner.  Go through the calibration and standardization process, then check standards.  Save the analysis routine as a sample setup.  Next, create a separate sample setup with a subset of elements to which the time-dependent correction will be assigned, in this example silicon and sodium.  Note time-dependent element calibration corrections can only be applied to elements that are the first element collected by each spectrometer.

Locate the sample to obtain the time-dependent element correction acquisition upon.  This should be either the unknown sample or a material similar to the unknowns.

Click the Special Options button in the Acquire! window.

The TDI Calibration and Quick Wavescan Samples window opens.  The default acquisition choice is Normal Acquisition.

Click the Setups button in the TDI Calibration and Quick Wavescan Samples window, opening the Sample Setup Selection dialog box.  Check that the appropriate volatile setup is active.

Click the OK button.

Select the Assigned Calibration TDI Acquisition method.  Enter text into the Volatile Sample Name field. Edit the TDI Count Time Intervals (number of steps in the calibration line) and adjust the Stage X and Y Increment (um) values if required, resulting in the following window.

The time-dependent element effect will be calibrated precisely on the first unknown sample.  One important assumption is that the time-dependent element behavior on the calibration sample is similar to all the unknowns to be analyzed. Therefore, because a consistent time-dependent correction is used, variation in composition represents real differences in composition (or volatilization) not precision of the analyses.  Each element in the method (sample setup) will be acquired one element at a time.  In this example, 10 time intervals are specified and the default on-peak count times for silicon and sodium are 20 and 40 respectively.  Thus the calibration curve for each will be composed of 10 spots of 2 and 4 seconds each.  The program will acquire each element in the sample (at a fresh spot) at a new stage position based on the X and Y increments specified to allow the time-dependent element effect to be calibrated precisely.

Click the OK button to return to the Acquire! window.

The time-dependent sample acquisition will start automatically upon clicking the Start Standard or Unknown Acquisition button, using the sample name entered in the TDI Sample Name field.

The user may now collect unknown data points.  REMEMBER to load in the appropriate sample setup AND under the Special Options button of the Acquire! window, click the Normal Acquisition button.

After collecting an unknown sample the user may assign the time-dependent correction from the Analyze! window.

The time-dependent element calibration can be assigned to any element in a sample provided that it was acquired as the FIRST element on that spectrometer.

Click the Standard Assignments button.  The Standard and Interference Assignments window opens.

To evaluate the correction effect on sodium, click that element’s row to view and edit its’ time-dependent assignment.

The Assignment Properties window opens and lists all possible time-dependent element calibration samples.

Highlight the appropriate TDI Element Calibration Sample Assignment sample and click the Display TDI Fit button.

The TDI (Assigned) Fit Plot window opens.

The drop in sodium count intensity with time appears to fit an exponential function (Nielson and Sigurdsson, 1981).  A plot of the natural log of the intensity data versus time should yield a straight line relationship as seen above.

The extrapolation correction is quantitative in that the correction is based on a calibration curve acquired during the run.  The correction uses the actual elapsed time for all calculations.  It is applied during the ZAF or Phi-Rho-Z iteration phase of the analysis to correct for changes in the matrix correction due to the extrapolation correction.

Along with alkali loss, the operator may notice an increase in count intensity from the non-mobile elements (such as silicon and aluminum) in the sample.  The possible enhancement with time may be corrected for as well.

Close the TDI (Assigned) Fit Plot window, returning to the Assignment Properties window.  Click the OK button.

To evaluate the possible enhancement of intensity with time on silicon in the rhyolitic glass, click on the element row for silicon in the Standard and Interference Assignments window.

The Assignment Properties window opens.  Again, highlight the appropriate TDI Element Calibration Sample Assignment sample and click the Display TDI Fit button.

The TDI (Assigned) Fit Plot window opens displaying the natural log of silicon x-ray intensity versus time.  The enhancement is evident and maybe corrected for quantitatively.

Return to the Analyze! window by clicking the Close button above.

Click the OK button of the Assignment Properties and the Standard and Interference Assignments windows, respectively.


Advanced Interference Corrections

Probe for EPMA permits the user to select a fully quantitative correction for spectral interferences (Donovan et al., 1993).  The user can correct for up to four interfering elements per channel.  The program requires that both the interfered and interfering elements be analyzed for. Further, an interference calibration standard must be acquired that contains a major concentration of the interfering element and none of the interfered element nor any other elements that interfere with the interfered element.

Most interferences between a pair of elements work in one direction. Consider a phase with high concentrations of manganese in the presence of the iron.  Here the Mn Kβ line interferes with the Fe Kα analytical line.  The reverse situation does not cause a problem, iron does not interfere with manganese.  However, there are some cases where both elements interfere with each other!  These dual interfering elements are extremely difficult to correctly quantify.  Fortunately, Probe for EPMA can handle this situation because it’s quantitative interference correction is an iterated solution (see Donovan et al., 1993 for details).

The following example (analyzed at U.C. Berkeley by John Donovan) involves the dual interference of zinc and rhenium in a natural organo-metallic phase.  Both lines interfere with each other (Zn Kβ and Re Lα) and both lines are used for quantitative analysis.  Other elements analyzed for are cesium, iron, sulfur, and selenium.  Oxygen, nitrogen, carbon and hydrogen are also in the samples.  Each is listed in the Element/Cations window for use in the matrix correction routine but are not analyzed.

Solving these interferences requires the analyses of both rhenium and zinc and two interference standards. To correct for the interference on zinc, a standard that contains rhenium but no zinc is required. Likewise, to handle the interference on rhenium, a standard will be needed that contains zinc but no rhenium.

The procedure to specify interferences and the correction thereof was documented in the silicate chapter in the Users Guide to Getting Started manual, cogent details will briefly mentioned here.

To view the various interference assignments, click the Standard Assignment button in the Analyze! window.

The Standard and Interference Assignments window opens.

In John’s routine, Cs Lα is measured on the standard CsBr, likewise Fe Kα is done on a pyrite (FeS2) standard and the Se Kα line is measured on the ZnSe standard.

Zinc, rhenium and sulfur require additional discussion.

Highlighting the zinc element row opens the Assignment Properties window for that element.  The user has the option to display all possible interferences based on the current set of analyzed and interfered elements. Clicking the Calculate Interference button displays these.  The program calculates the interference based on a gaussian peak shape assuming a worst case scenario of 0.1% of the analyzed element and 100% of each of the other analyzed and possibly interfering elements.

As mentioned earlier and seen below, Re Lα interferes with the Zn Kα on the LIF crystal.  To correct for this overlap, a standard with no zinc present (rhenium metal) is employed for the calculation.

The Assignment Properties window for Re Lα is shown next.  Here, Zn Kα interferes with the Re Lα x-ray position.  To make the quantitative correction the standard ZnSe (with no rhenium) is used.

Finally, the Assignment Properties window for S Kα is displayed.  In this unique mineral, Cs Lα overlaps with S Kβ.  Therefore, a cesium standard without sulfur (CsBr) is also required to complete the spectral interference deconvolution!

After setting up the parameters for the analysis session, calibration and standardization was accomplished, then several tiny and complex organometallic phases were probed for their chemistry.

The Analysis Calculation Options window allows the user to enable or disable the interference correction routine to view its effect on quantitative analysis results.  This window is activated from the Analytical | Analysis Options menu in the main Probe for EPMA log window.

Data illustrating the power of the Probe for EPMA interference correction routine is shown below.  The results are for a real unknown sample that has some variability and was rather small to analyze.  The following unknown analysis is without any interference corrections.

Un   10 Zn-ReSCN gr2
TakeOff = 40.0  KiloVolt = 20.0  Beam Current = 20.0  Beam Size =    0
(Magnification =        .)                    Beam Mode = Analog  Spot
Number of Data Lines:  5               Number of 'Good' Data Lines:  3
First/Last Date-Time: 06/15/1998 04:39:29 PM to 06/15/1998 04:55:23 PM

Average Total Oxygen:         .000     Average Total Weight%:  122.912
Average Calculated Oxygen:    .000     Average Atomic Number:   53.080
Average Excess Oxygen:        .000     Average Atomic Weight:   54.153
Average ZAF Iteration:        4.00     Average Quant Iterate:     2.00


Results in Elemental Weight Percents

SPEC:        O       N       C       H
TYPE:     SPEC    SPEC    SPEC    SPEC

AVER:    1.900   5.000   4.200    .200
SDEV:     .000    .000    .000    .000

ELEM:       Cs      Fe      Zn      Re       S      Se
BGDS:      LIN     LIN     LIN     LIN     LIN     LIN
TIME:    10.00   10.00   10.00   10.00   10.00   10.00

ELEM:       Cs      Fe      Zn      Re       S      Se   SUM
    53    .000    .000  19.553  73.669  17.465    .000 121.988
    55    .000    .007  20.556  74.517  16.507    .000 122.886
    56    .000    .019  19.668  74.715  18.159    .000 123.862

AVER:     .000    .009  19.926  74.301  17.377    .000 122.912
SDEV:     .000    .010    .549    .556    .830    .000
SERR:     .000    .006    .317    .321    .479    .000
%RSD:       .1   113.3     2.8      .7     4.8      .1
STDS:      834     730     660     575     730     660

STKF:    .5978   .4297   .5001  1.0000   .4719   .5162
STCT:    627.1  3670.9  3712.9  3672.8  4423.1  1315.9

UNKF:    .0000   .0001   .2100   .6516   .0943   .0000
UNCT:     -2.7      .3  1559.2  2393.0   884.3    -3.2
UNBG:     11.6    28.0    67.3    61.5     6.6    64.0

ZCOR:   1.2016   .9276   .9487  1.1404  1.8421  1.0254
KRAW:   -.0043   .0001   .4199   .6516   .1999  -.0024
PKBG:      .77    1.01   24.18   40.00  135.89     .95

Results Based on 6 Atoms of re

SPEC:        O       N       C       H
TYPE:     SPEC    SPEC    SPEC    SPEC

AVER:    1.786   5.368   5.258   2.983
SDEV:     .013    .040    .039    .022

ELEM:       Cs      Fe      Zn      Re       S      Se   SUM
    53    .000    .000   4.536   6.000   8.260    .000  34.323
    55    .000    .002   4.714   6.000   7.718    .000  33.784
    56    .000    .005   4.499   6.000   8.468    .000  34.281

AVER:     .000    .002   4.583   6.000   8.149    .000  34.129
SDEV:     .000    .003    .115    .000    .387    .000
SERR:     .000    .002    .067    .000    .224    .000
%RSD:       .8   113.2     2.5      .0     4.8      .8

The user immediately realizes that there is a problem since the unknown sums to a total of 123%.

The following is the same unknown but with the iterated interference software applied.

Un   10 Zn-ReSCN gr2
TakeOff = 40.0  KiloVolt = 20.0  Beam Current = 20.0  Beam Size =    0
(Magnification =        .)                    Beam Mode = Analog  Spot
Number of Data Lines:  5               Number of 'Good' Data Lines:  3
First/Last Date-Time: 06/15/1998 04:39:29 PM to 06/15/1998 04:55:23 PM

Average Total Oxygen:         .000     Average Total Weight%:  101.367
Average Calculated Oxygen:    .000     Average Atomic Number:   54.332
Average Excess Oxygen:        .000     Average Atomic Weight:   50.124
Average ZAF Iteration:        4.00     Average Quant Iterate:    13.00


Results in Elemental Weight Percents

SPEC:        O       N       C       H
TYPE:     SPEC    SPEC    SPEC    SPEC

AVER:    1.900   5.000   4.200    .200
SDEV:     .000    .000    .000    .000

ELEM:       Cs      Fe      Zn      Re       S      Se
BGDS:      LIN     LIN     LIN     LIN     LIN     LIN
TIME:    10.00   10.00   10.00   10.00   10.00   10.00

ELEM:       Cs      Fe      Zn      Re       S      Se   SUM
    53    .000    .000   6.237  65.851  17.508    .000 100.896
    55    .000    .007   7.365  65.342  16.513    .000 100.527
    56    .000    .020   6.100  67.048  18.210    .000 102.678

AVER:     .000    .009   6.568  66.080  17.410    .000 101.367
SDEV:     .000    .010    .694    .876    .853    .000
SERR:     .000    .006    .401    .506    .492    .000
%RSD:       .1   113.3    10.6     1.3     4.9      .0
STDS:      834     730     660     575     730     660

STKF:    .5978   .4297   .5001  1.0000   .4719   .5162
STCT:    627.1  3670.9  3712.9  3672.8  4423.1  1315.9

UNKF:    .0000   .0001   .0691   .5781   .0943   .0000
UNCT:     -2.7      .3   512.7  2123.2   884.3    -3.2
UNBG:     11.6    28.0    67.3    61.5     6.6    64.0

ZCOR:   1.2162   .9395   .9511  1.1431  1.8455  1.0235
KRAW:   -.0043   .0001   .1381   .5781   .1999  -.0024
PKBG:      .77    1.01    8.61   35.59  135.89     .95
INT%:      .00     .00  -67.17  -11.27     .00     .00

Results Based on 6 Atoms of re

SPEC:        O       N       C       H
TYPE:     SPEC    SPEC    SPEC    SPEC

AVER:    2.008   6.036   5.913   3.355
SDEV:     .026    .080    .078    .044

ELEM:       Cs      Fe      Zn      Re       S      Se   SUM
    53    .000    .000   1.619   6.000   9.264    .000  34.252
    55    .000    .002   1.926   6.000   8.805    .000  34.239
    56    .000    .006   1.555   6.000   9.463    .000  34.083

AVER:     .000    .003   1.700   6.000   9.177    .000  34.191
SDEV:     .000    .003    .199    .000    .337    .000
SERR:     .000    .002    .115    .000    .195    .000
%RSD:      1.2   112.3    11.7      .0     3.7     1.3

The total now is acceptable, slightly over 100%.

Finally to demonstrate with standard samples (of known composition), both rhenium metal and the ZnSe standard will be reproduced without the benefit of the interference correction.

St  575 Set   1 rhenium metal
TakeOff = 40.0  KiloVolt = 20.0  Beam Current = 20.0  Beam Size =    0
(Magnification =        .)                    Beam Mode = Analog  Spot
Number of Data Lines:  5               Number of 'Good' Data Lines:  5
First/Last Date-Time: 06/15/1998 01:40:14 PM to 06/15/1998 01:48:18 PM

Average Total Oxygen:         .000     Average Total Weight%:  124.489
Average Calculated Oxygen:    .000     Average Atomic Number:   67.420
Average Excess Oxygen:        .000     Average Atomic Weight:  142.028
Average ZAF Iteration:        3.00     Average Quant Iterate:     2.00

Results in Elemental Weight Percents

ELEM:       Cs      Fe      Zn      Re       S      Se
BGDS:      LIN     LIN     LIN     LIN     LIN     LIN
TIME:    10.00   10.00   10.00   10.00   10.00   10.00

ELEM:       Cs      Fe      Zn      Re       S      Se   SUM
     1    .000    .000  21.083 103.240    .022    .000 124.344
     2    .161    .005  20.061 102.988    .008    .000 123.223
     3    .000    .000  21.644 103.274    .000    .176 125.093
     4    .000    .023  21.565 103.834    .000    .156 125.578
     5    .000    .000  19.873 104.122    .000    .213 124.207

AVER:     .000    .000  20.845 103.492    .000    .109 124.489
SDEV:     .072    .010    .833    .469    .009    .102
SERR:     .032    .004    .372    .210    .004    .045
%RSD:    223.6   177.2     4.0      .5   159.1    93.2

PUBL:    n.a.    n.a.    n.a.  100.000   n.a.    n.a.  100.000
%VAR:      .00     .00     .00    3.49     .00     .00
DIFF:     .000    .000    .000   3.492    .000    .000
STDS:      834     730     660     575     730     660

STKF:    .5978   .4297   .5001  1.0000   .4719   .5162
STCT:    628.0  3665.4  3726.7  3679.3  4367.5  1316.2

UNKF:    .0003   .0001   .2405  1.0000   .0000   .0012
UNCT:       .0    -1.1  1792.2  3679.3     -.4     1.8
UNBG:     11.6    41.9    95.5    93.5     8.4    93.4

ZCOR:   1.2008   .8713   .8667  1.0349  2.1623   .9212
KRAW:    .0000  -.0003   .4809  1.0000  -.0001   .0013
PKBG:     1.00     .97   19.77   40.37     .95    1.02


St  660 Set   1 ZnSe (synthetic)
TakeOff = 40.0  KiloVolt = 20.0  Beam Current = 20.0  Beam Size =    0
(Magnification =        .)                    Beam Mode = Analog  Spot
Number of Data Lines:  5               Number of 'Good' Data Lines:  5
First/Last Date-Time: 06/15/1998 01:50:41 PM to 06/15/1998 01:58:46 PM

Average Total Oxygen:         .000     Average Total Weight%:  151.911
Average Calculated Oxygen:    .000     Average Atomic Number:   47.924
Average Excess Oxygen:        .000     Average Atomic Weight:   92.977
Average ZAF Iteration:        3.00     Average Quant Iterate:     2.00

Results in Elemental Weight Percents

ELEM:       Cs      Fe      Zn      Re       S      Se
BGDS:      LIN     LIN     LIN     LIN     LIN     LIN
TIME:    10.00   10.00   10.00   10.00   10.00   10.00

ELEM:       Cs      Fe      Zn      Re       S      Se   SUM
     6    .000    .000  44.441  55.414    .019  51.417 151.290
     7    .000    .014  44.022  56.029    .036  51.668 151.768
     8    .000    .000  44.551  56.249    .035  51.099 151.934
     9    .000    .015  44.700  56.069    .037  50.974 151.795
    10    .000    .000  45.281  56.025    .052  51.409 152.767

AVER:     .000    .000  44.599  55.957    .036  51.314 151.911
SDEV:     .000    .008    .457    .317    .012    .277
SERR:     .000    .003    .204    .142    .005    .124
%RSD:       .0   137.1     1.0      .6    33.5      .5

PUBL:    n.a.    n.a.   45.290   n.a.    n.a.   54.710 100.000
%VAR:      .00     .00   -1.53     .00     .00   -6.21
DIFF:     .000    .000   -.691    .000    .000  -3.396
STDS:      834     730     660     575     730     660

STKF:    .5978   .4297   .5001  1.0000   .4719   .5162
STCT:    628.0  3665.4  3726.6  3679.0  4367.5  1316.2

UNKF:    .0000   .0001   .5002   .5193   .0002   .5162
UNCT:    -23.1    -1.4  3726.7  1910.6     1.7  1316.2
UNBG:     29.3    23.0    48.3    47.7     5.6    47.5

ZCOR:   1.1332   .8826   .8917  1.0775  1.9149   .9940
KRAW:   -.0367  -.0004  1.0000   .5193   .0004  1.0000
PKBG:      .21     .95   78.18   41.17    1.31   28.85

The rhenium standard displays an apparent rhenium concentration of 103% and a zinc total of an additional 20%.  The ZnSe is even more interesting in that the total is approaching 152% with 44% zinc, 56% rhenium and 51% selenium.  The true composition is 45% zinc, and 55% selenium with no rhenium!  The normal matrix correction comes close with respect to both zinc and selenium abundances but also reports a whopping 56% rhenium concentration.

Below, both standards are rerun with the interference corrections applied.

St  575 Set   1 rhenium metal
TakeOff = 40.0  KiloVolt = 20.0  Beam Current = 20.0  Beam Size =    0
(Magnification =        .)                    Beam Mode = Analog  Spot
Number of Data Lines:  5               Number of 'Good' Data Lines:  5
First/Last Date-Time: 06/15/1998 01:40:14 PM to 06/15/1998 01:48:18 PM

Average Total Oxygen:         .000     Average Total Weight%:  100.138
Average Calculated Oxygen:    .000     Average Atomic Number:   74.736
Average Excess Oxygen:        .000     Average Atomic Weight:  184.273
Average ZAF Iteration:        2.00     Average Quant Iterate:    13.80

Results in Elemental Weight Percents

ELEM:       Cs      Fe      Zn      Re       S      Se
BGDS:      LIN     LIN     LIN     LIN     LIN     LIN
TIME:    10.00   10.00   10.00   10.00   10.00   10.00

ELEM:       Cs      Fe      Zn      Re       S      Se   SUM
     1    .000    .000    .375  99.332    .023    .000  99.730
     2    .164    .005    .000  99.625    .007    .000  99.801
     3    .000    .000   1.096  98.532    .000    .168  99.796
     4    .000    .023    .847  99.364    .000    .149 100.383
     5    .000    .000    .000 100.776    .000    .203 100.979

AVER:     .000    .000    .463  99.526    .000    .104 100.138
SDEV:     .074    .010    .496    .810    .010    .097
SERR:     .033    .004    .222    .362    .004    .043
%RSD:    223.6   177.2   107.0      .8   164.6    93.2

PUBL:    n.a.    n.a.    n.a.  100.000   n.a.    n.a.  100.000
%VAR:      .00     .00     .00    -.47     .00     .00
DIFF:     .000    .000    .000   -.474    .000    .000
STDS:      834     730     660     575     730     660

STKF:    .5978   .4297   .5001  1.0000   .4719   .5162
STCT:    628.0  3665.4  3726.7  3679.3  4367.5  1316.2

UNKF:    .0003   .0001   .0055   .9942   .0000   .0012
UNCT:       .0    -1.1    10.3  3657.8     -.4     1.8
UNBG:     11.6    41.9    95.5    93.5     8.4    93.4

ZCOR:   1.2302   .8692   .8409  1.0011  2.3003   .8800
KRAW:    .0000  -.0003   .0028   .9942  -.0001   .0013
PKBG:     1.00     .97    1.11   40.14     .95    1.02
INT%:      .00     .00  -99.57    -.59   -2.86     .00

St  660 Set   1 ZnSe (synthetic)
TakeOff = 40.0  KiloVolt = 20.0  Beam Current = 20.0  Beam Size =    0
(Magnification =        .)                    Beam Mode = Analog  Spot
Number of Data Lines:  5               Number of 'Good' Data Lines:  5
First/Last Date-Time: 06/15/1998 01:50:41 PM to 06/15/1998 01:58:46 PM

Average Total Oxygen:         .000     Average Total Weight%:  100.261
Average Calculated Oxygen:    .000     Average Atomic Number:   32.317
Average Excess Oxygen:        .000     Average Atomic Weight:   72.276
Average ZAF Iteration:        3.00     Average Quant Iterate:    14.80

Results in Elemental Weight Percents

ELEM:       Cs      Fe      Zn      Re       S      Se
BGDS:      LIN     LIN     LIN     LIN     LIN     LIN
TIME:    10.00   10.00   10.00   10.00   10.00   10.00

ELEM:       Cs      Fe      Zn      Re       S      Se   SUM
     6    .000    .000  45.120    .000    .017  54.799  99.936
     7    .000    .014  44.451   1.080    .032  54.982 100.559
     8    .000    .000  45.152    .477    .032  54.459 100.120
     9    .000    .015  45.415    .000    .033  54.368  99.831
    10    .000    .000  46.003    .000    .047  54.807 100.857

AVER:     .000    .000  45.228    .311    .032  54.683 100.261
SDEV:     .000    .008    .560    .477    .011    .259
SERR:     .000    .003    .251    .213    .005    .116
%RSD:       .0   137.1     1.2   153.1    33.5      .5

PUBL:    n.a.    n.a.   45.290   n.a.    n.a.   54.710 100.000
%VAR:      .00     .00    -.14     .00     .00    -.05
DIFF:     .000    .000   -.062    .000    .000   -.027
STDS:      834     730     660     575     730     660

STKF:    .5978   .4297   .5001  1.0000   .4719   .5162
STCT:    628.0  3665.4  3726.6  3679.0  4367.5  1316.2

UNKF:    .0000   .0001   .4995   .0028   .0002   .5162
UNCT:    -23.1    -1.4  3721.5     2.3     1.7  1316.2
UNBG:     29.3    23.0    48.3    47.7     5.6    47.5

ZCOR:   1.0893   .8842   .9055  1.1054  1.7229  1.0593
KRAW:   -.0367  -.0004   .9986   .0006   .0004  1.0000
PKBG:      .21     .95   78.08    1.05    1.31   28.85
INT%:      .00     .00    -.14  -99.88     .00     .00

Now, the apparent zinc in the rhenium metal standard is gone and the average total sums correctly near 100%. The ZnSe standard is perfect, matching the published standard composition for both zinc and selenium, and effectively removing all of the apparent 56% rhenium.


Light Element Analysis - Empirical APFs

Quantitative analysis of light elements (beryllium to fluorine) is difficult to do correctly with the electron microprobe.  Numerous issues impede the analysis of light elements (see Appendix B in the User’s Guide and Reference documentation as well as Goldstein et al., 1992 for further discussion).  A few brief comments will be made here, as an introduction to this section.

Typically, for x-ray analysis in this energy range (0.1 to 0.7 keV), a large absorption correction is necessary.  This large correction in conjunction with the fact that the mass absorption coefficients for the low energy x-rays are very large and not very well known (see Appendix C of the User’s Guide and Reference documentation) place a severe demand on the established ZAF and Phi-Rho-Z models for light element matrix corrections. Some reduction in this large x-ray absorption factor is possible by analyzing at higher take-off angles and at lower acceleration voltages.  The former aids by shortening the path length for absorption in the sample while the latter involves a decrease in the electron beam penetration so that x-rays are generated closer to the surface and can escape to be detected.

Low count rates for these light elements are often found. This is due to the low fluorescent yields from the Kα x-ray lines and the inefficient nature of WDS counting systems.  Count rates can be increased by increasing the beam current substantially but this may then lead to deadtime problems for metal lines that interfere with the light element lines of interest.  The use of new layered dispersive element (LDE) synthetic multi-layer crystals with large d-spacings can also improve light element peak count rates.

There is also a strong possibility of interferences from higher order metal lines such as titanium, chromium, manganese, iron, nickel, zirconium, niobium, and molybdenum with the light element lines.  These interferences are often severe for minor and trace level measurements.  It is critical to eliminate peak overlaps and understand the background intensity around the light element peak position in both the sample and standard.  The new LDE multi-layer crystals also help here by strongly suppressing these higher order reflections.

Finally, chemical bonding effects can result in wavelength shifts, increases or decreases in the relative intensities of various lines and alteration of the shape of the analytical line.  These effects are most significant for the lightest (lowest energy) elements.  Polarization phenomena and crystallographic orientation may also cause variations in peak shape and intensity especially for boron.  Therefore, to measure the intensity of the light elements, one measures the integral intensity (area) under the characteristic peak rather than just the peak intensity.

Bastin and Heijliers (1984, 1991) pioneered the area-peak factor (APF) analysis method.  The APF is defined as the ratio between the integral intensity (area) k-ratio from the sample and the standard and the peak intensity k-ratio from the same sample and the standard.  This factor is only valid for a given compound with respect to a given standard and a specific spectrometer setup. These integral measurements can be made by scanning the spectrometer in small steps across the characteristic peak with the wavescan feature in Probe for EPMA.  After acquiring the peak shape profiles for a primary standard and a number of secondary standards and correcting for the background and removal of interfering peaks, the APF can be calculated as:

Where :

is the integrated intensity of the secondary standard

is the peak intensity of the primary standard

is the peak intensity of the secondary standard

is the integrated intensity of the primary standard

After an APF has been determined, future measurements of that compound can be accomplished by measuring only the peak intensity in the sample and standard.  Then, multiplication of the peak k-ratio with the appropriate APF will yield the correct integral k-ratio.

Probe for EPMA allows the user to select an APF correction for use in correcting x-ray intensities for peak shift and shape changes between the standard and the multi-element unknown.  This is critical when the user is analyzing the Kα lines of the light elements (boron, carbon, nitrogen and oxygen).  This correction may also be of use for other elements such as S Kα that also exhibits peak shift and shape changes when comparing sulfate and sulfide peaks.

The power of this correction will be documented in the analysis of oxygen using several simple oxide standards.  Open a new Probe for EPMA run.  Here oxygen Kα will be analyzed using MgO as the calibration standard.  The spectrometer is equipped with a standard P-10 gas flow counter and a LDE (59.8Å) W/Si reflecting crystal.  The other two elements to be determined are magnesium on MgO and silicon on SiO2 .  Peak the three elements and acquire standard samples for each.

Create a new unknown sample and collect data on the SiO2 standard.  Analyze the sample from the Analyze! window.

A low total for the analysis is found.  The nominal composition for the SiO2 standard is silicon: 46.74 and oxygen: 53.26.  Here, the discrepancy in the total rests primarily with the oxygen concentration.  The low oxygen concentration is independent of the matrix correction (and mass absorption coefficient) and can only be corrected for by using the appropriate APFs.

Select Analytical | Analysis Options from the main Probe for EPMA log window to open the Analysis Calculation Options window.  Click the Use Empirical APF Values check box to activate this option.

Click the OK button.

Next, select Analytical | Empirical APFs from the main Probe for EPMA log window to open the Add Empirical APFs (area peak factors) to Run window.

Find the appropriate correction.  Two choices are given for oxygen in the presence of SiO2, Bastin’s value of 1.04440 using Fe2O3 as the calibration standard and Donovan’s correction factor of 1.070 when using MgO as the calibration standard.  Although the values seen in the text field below are relative to Fe2O3, the APF for MgO relative to Fe2O3 is also 1.000, which means that these correction factors apply equally well relative to MgO.  Hence, the use of MgO as the calibration standard.

The APF correction values are defined in the EMPAPF.DAT file in the Probe for EPMA directory.  The file contains some 80 values that may or may not be applicable, depending on the analyzing crystals and standards available.  It is strongly recommended that the user measure the integral intensities and peak intensities and calculate the APFs for your particular spectrometer setup.  And if necessary, edit the EMPAPR.DAT file using any ASCII text editor such as NotePad to insert their own measurements.  See the User’s Guide and Reference documentation for editing format and details.

Click the Add To Run >> button to place the selected empirical APF into the run.

Click the OK button.

Re-analyze the unknown SiO2 sample.  The total is now closer to 100% and a more reasonable oxygen concentration is calculated.

The APF values selected or entered are always measured relative to some standard sample.  For example, if measuring oxygen Kα and using either MgO or Fe2O3 as the primary standard for oxygen, then any oxygen Kα APF values used must be those measured relative to either MgO or Fe2O3.  For the same reason, if using APF values for a particular light element and the user decides to re-assign the standard for that element, the APF values for that element must also be re-normalized to reflect the standard re-assignment.

Thus, it is most efficient to always use the same standard for each light element analyzed.  Typically (in order to utilize the APF values in the supplied EMPAPF.DAT file) these will be:

The APF correction in Probe for EPMA will allow the user to enter empirical APF values in each run.  The user may enter one or more APF factors for each emitting element although they are generally applied to soft x-ray lines.  The APF for each absorber will be summed according to it's weight fraction in the composition and applied to the emitting element counts during the ZAF or Phi-Rho-Z iteration.


References

Bastin G.F., and Heijliers H.J.M., (1984)  Quantitative Electron-Probe Microanalysis of Carbon in Binary Carbides., Microbeam Analysis (Editors: A.D. Romig, Jr. and J.I. Goldstein) pp 291-294.

Bastin G.F., and Heijliers H.J.M., (1991)  Quantitative Electron-Probe Microanalysis of Ultra-Light Elements (Boron-Oxygen)., Electron Probe Quantitation (Editors: K.F.J. Heinrich and D.E. Newbury), New York: Plenum Press pp 145-161.

Deer W.A., Howie R.A., and Zussman J., (1992)  An introduction to The Rock-Forming Minerals, Second Edition, New York: Wiley.

Donovan J.J., Snyder D.A., and Rivers M.L., (1993)  An Improved Interference Correction for Trace Element Analysis. Microbeam Analysis 2 23-28.

Goldstein J.I., Newbury D.E., Echlin P., Joy D.C., Romig A.D. Jr, Lyman C.E., Fiori C., and Lifshin E., (1992)  Scanning Electron Microscopy and X-Ray Microanalysis, Second Edition, New York: Plenum.

Nielsen C.H., and Sigurdsson H., (1981)  Quantitative methods for electron microprobe analysis of sodium in natural and synthetic glasses. Am. Mineral., 66 547-552.

Reed S.J.B., (1993)  Electron-Microprobe Analysis, Second Edition, New York: Wiley.

Scott V.D., Love G.,and Reed S.J.B. (1995)  Quantitative Electron-Probe Microanalysis, Second Edition, London: Ellis Horwood.

Willis J.P., (1993)  Course on Theory and Practice of XRF Spectrometry, University of Western Ontario, Department of Geology, Course Notes.


WDS Background Offsets



Kakanui Hornblende WDS Scans

Kakanui Hornblende is an excellent microprobe standard for routine use in the microanalysis of silicates by WDS since it contains, like many hornblende amphiboles, a number of elements at easily detected concentrations. That is, Kakanui hornblende is a good proxy for the location of both the analytical peaks of interest as well as potential interferences for typical silicates. Kakanui hornblende can be used for WDS spectrometer peaking, the determination of background offsets which can be used for a wide range of silicate minerals as long as other elements are not present which could cause peak interferences, and can also be used as a either a primarly or secondary calibration standard.

In general it is highly recommended that you acquire full-range WDS wavelength scans on all primary standards that are used in your laboratory, or in the absence of such data, obtain WDS scans that can be used to inspect peak regions for potential interferences.

Note that for the analysis of other phases having compositions that are atypical, such as REE-bearing or other accessory phases, one must spend considerable time acquiring WDS wavelength scans on both standards and samples in order to adequately prepare for background selection and subsequent quantitative analysis.

Here are wavelength scans obtained on a grain of Kakanui hornblende using the WU JEOL JXA-8200 microprobe.

TAP Full Range Scan



TAP Na Scan

Nominal background offsets for Na:


TAP Mg Scan

Nominal background offsets for Mg:


TAP Al Scan

Nominal background offsets for Al:


TAP Si Scan

Nominal background offsets for Si:


PET Full Range Scan



PET P Scan

Nominal background offsets for P:


PET S Scan

Nominal background offsets for S:


PET Cl Scan

Nominal background offsets for Cl:


PET K Scan

Nominal background offsets for K:


PET Ca Scan

Nominal background offsets for Ca:


LiF Full Range Scan



LiF Ti Scan

Nominal background offsets for Ti:


LiF Cr Scan

Nominal background offsets for Cr:


LiF Mn Scan

Nominal background offsets for Mn:


LiF Fe Scan

Nominal background offsets for Fe:


Startwin Program


Launch the Startwin program.


This is the main window of the Startwin Program:

This is the Count Acquisition window:

The Startwin program can be used for the following utility functions:

These capabilities are outlined in the following sections.

See the Startwin help documentation for further information.


Beam and Detector Stability

Testing beam (drift) stability is an important step prior to acquiring any quantitative data.  The following step-by-step procedure illustrates how to monitor and plot beam current with time.  On JEOL microprobes the FARADAY module may also be run for measurement of the beam (see the User’s Guide and Reference documentation for further details).

From the STARTWIN log window, select Modes from the menu bar and click on Cycle Counters from the menu choices.  The Measure Faraday menu should also be selected.



Next, click the Count Times button in the Count Acquisition window.


This opens the Count Times dialog box.  Choose an On Peak and Cycle Time count time.  The On Peak time is the time the scalers will count and the Cycle Time is the interval of time between successive measurements.  Thus, the sum of both numbers is the time between measurements of the beam.  Finally, disable the beam drift correction, confirm that the Use Beam Drift Correction box is unchecked.


Click the OK button returning to the Count Acquisition window.


Click the Start Count button to initiate a continuous cycle of beam current measurements.  In this example, a twenty second scaler count will be done followed by a twenty-second countdown and then a Faraday current measurement.  This process repeats until the user cancels the loop.


When the user has acquired a suitable number of beam current measurements, click the Cancel button in the Automation Status bar located at the bottom right side of the STARTWIN log window to stop the acquisition cycle.


The STARTWIN log window will contain the beam current data acquired so far (in this example, reported in nanoamps).  The other three columns represent counts in counts per second acquired by each spectrometer with the beam blanked (essentially detector/electronics noise).



Evaluating the trend between beam current and time may best be viewed in graphical format rather than in looking at a long series of numbers.  Use the mouse to select the data set to graph.  Then, select Output from the menu bar and click Plot Count Data (Selected in Log Window) from the drop-down menu choices.



This opens the Display Data window.


While all data columns were selected by the mouse operation previously, the user may plot a single column of data by clicking the column label of the desired data and then clicking the Graph Selected button.


Below, Beam Counts versus Measurement Number (time) are graphed and the overall beam stability with time may be judged.  Here, the beam measurements fluxuate in the second decimal place (between 30.40 and 30.49).

The numeric value of any point on the graph maybe read by placing the mouse cursor over the data point and viewing it’s value in the two windows directly above the Close button (bottom right).


Click the Close button to return to the STARTWIN log window.


Determination of WDS Spectrometer Alignment

It is important to confirm that all wavelength-dispersive spectrometers (WDS) on the microprobe are individually aligned with respect to the Rowland circle, and further that all WDS are mutually aligned so that all Rowland circles are coincident at the sample focus position. Two components of alignment are necessary:


Determination of WDS Spectrometer Deadtime

Currently done using Excel spreadsheet with Probe for EPMA Remote Automation Interface.


Determination of Pulse Height Analysis Parameters

Pulse Height Analysis

It is important to properly discriminate between the X-ray pulses produced by the sample and baseline noise which is also produced by the detector system. Remember that unlike an EDS pulse processing system, the WDS pulse processing electronics do not have digital electronics that automatically process the pulses and ignore the noise pulses. The WDS system requires that you properly set the detector voltage (hereafter referred to as the bias), gain, baseline, and window for X-ray counting. The procedure established on the Caltech JEOL 733 microprobe was to vary the detector bias in order to produce a 4V voltage pulse on a single-channel analyzer (SCA) plot. This was to reduce the detector deadtime dependence on X-ray energy. More practically, this approach allows the analyst to use a consistent baseline and window setting for all elements, but does require that specific detector bias values are used for the element being analyzed.

The proper PHA settings for WDS analysis require that the effective gain is sufficient to cleanly separate the X-ray pulses from the baseline noise of the detector system. All WDS counting systems are operated at a detector voltage, typically referred to as the bias, which is typically 1500-1900 volts, and results in the detector operating in the proportional counter range, where a photon produces a signal that is proportional to the X-ray energy.

The effect of increasing the detector bias is to increase the effective resolution of the X-ray pulse from the baseline noise, so that on an SCA plot the X-ray pulse moves to a higher voltage value. The effect of increasing the detector gain is to also increase the resolution and also moves the pulse to a higher SCA voltage. In practice, there are equivalent sets of bias and gain settings that should result in identical counting performance as measured by P/B and total count rate. Given the sets of equivalent bias and gain values, it is prudent to use a higher gain coupled with a lower detector bias in order to maximize the life of the detector. This implicitly assumes that modern counting systems do not exhibit increased noise at higher gain settings, which appears to be the case. You should demonstrate to your own satisfaction that these relationships are observed on your system.

Two different approaches are used for JEOL and Cameca microprobes which result in similar discrimination of pulses from baseline noise. On the JEOL microprobe, the detector gain is chosen from integer values of (typically) 16x, 32x, 64x, or 128x, and the bias is adjusted relative to the chosen gain value in order to achieve the proper PHA settings.

On the Cameca microprobe, the gain is continuously variable and the procedure is to choose a detector bias and adjust the gain value again in order to achieve the proper PHA settings. Experience with WDS counting systems has revealed that there is a dependance of the deadtime of the counting electronics on X-ray energy, which can be minimized by always applying the same effective gain to the X-rays being measured. This is the reason for adjusting the bias and gain in order to produce a consistent single channel analyzer (SCA) plot of the observed voltage distribution returned from the counting electronics. On a JEOL microprobe the bias and gain are set to obtain a ~4V SCA pulse, and on a Cameca microprobe the bias and gain are set to obtain a ~2.5V SCA pulse distribution. These relationships are summarized below:

JEOL Microprobe PHA Method

For a given spectrometer, generally two gain values can be used that cover all X-ray energies for the crystals on that spectrometer:

Cameca Microprobe PHA Method

After the peak position has been determined, it is important to verify that the Pulse Height Analysis (PHA) settings are correct for this element. The spectrometer peak position can be determined with PHA settings that are only approximately correct since the peak intensity is usually obvious. The reverse approach is not practical as adjusting the PHA settings requires that the spectrometer is already tuned to the peak position.


PHA Parameters: Bias Scans (JEOL)

Move Motors window showing current spectrometer positions before move operation:

Spectrometer 3 is selected for Cu Kα using the LiF crystal:

Move Motors window showing spectrometer positions after move operation (drive to Cu peaks all spectrometers):

Peak/Scan Options selected and display of peak positions before update:

Peak/Scan Options display of peak positions after update (spectrometer positions have been updated):

All spectrometers have been peaked and peaking results are displayed:

PHA button is clicked to open Pulse Height Analysis Parameters Window:

A Bias Scan is initiated and the results are shown for all spectrometers:

NEED PLOT

Summary of bias scan for all spectrometers on Cu at 32x gain:

Summary of bias scan for spectrometer 5 on Cu at 32x gain:

SCA scan acquired on all spectrometers using 32x gain and bias values determined from bias scan. Note that now the correct bias has been used and all SCA scans show a pulse at ~4V:

Summary of SCA scan acquired on all spectrometers using 32x gain and bias values determined from bias scan. Note the broad pulse distribution of Cu Lα compared to the sharper distribution for Cu Kα. Light element and low energy X-ray pulses will generally be smeared out on an SCA plot, and the determination of the baseline voltage setting is more important:

Long count SCA scan acquired on all spectrometers using 32x gain and bias values determined from bias scan. Note the broad distribution of pulses, this is the reason for using integral count mode:


PHA Parameters: Gain Scans (Cameca)

This section awaits PHA scans obtained on Cameca instruments...by Cameca operators...you know who you are.

For Cameca instruments one uses a fixed bias value for all elements measured on that detector, and uses a specific gain value for each element since the gain can be varied over a significant range. So for Cameca setups one obtains gain scans that are specific to an element x-ray energy in order to obtain a PHA scan with a pulse at the desired voltage.

Because the Cameca uses P-10 Ar-methane gas for all counters, it is necessary to perform PHA scans on standards containing K, Ca, Sc, Ti, etc. for K-lines, and on In, Sn, Sb, etc. for L-lines. These PHA plots will reveal the presence of the Ar escape peak which must be either included or excluded as necessary for measurement.


PHA Parameters: SCA Scans

The term PHA stands for "pulse height analysis" and the term SCA stands for "single channel analyzer" (the terms PHA and SCA seem to be used interchangeably). EDS counting systems have a sophisticated pulse processing system using digital electronics that are capable of very high throughput and discrimination of X-ray pulses from one another and from background noise. Because WDS detector counting systems do not have the sophisticated pulse discrimination capabilities that EDS systems have, it is necessary for the operator to select the detector bias and/or gain as well as the baseline and window for pulse processing. As has been discussed, a bias scan is used on JEOL instruments to determine the bias that is used for a specific gain value in order to obtain an SCA scan with the pulse at ~4V. For the Cameca a gain scan is used for a specific bias in order to obtain an SCA scan at ~2V.

SCA Scans on P-10 Flow Counters

All Cameca detectors and the JEOL flow-proportional counters use P-10 Ar-methane gas mixture. For measurements using these flow-proportional counters, it is necessary to perform SCA scans to determine the proper baseline and window settings for routine measurement. The baseline setting is to eliminate baseline detector/electronics noise from measurement. A detector that exhibits a significant count rate with the electron beam turned off is likely passing noise counts due to a baseline that is too low. It is recommended to use the integral window setting rather than differential mode in order to accomodate pulse pileup at higher count rates. The importance of PHA energy discrimination has diminished since modern WDS spectrometers have excellent X-ray resolution so that PHA overlaps are not typically observed. While any higher order X-ray pulse that is diffracted at the same spectrometer position as the line of interest requires PHA discimination via specific baseline and window settings, this is not typical.

It is critically important to determine the baseline setting especially for light elements as the X-ray pulse energy is similar to that of the baseline noise. SCA scans should be acquired for light element measurement for each analytical session

SCA Scans on P-10 Flow Counters -- Ar Escape Peaks

X-rays that have an energy just above the Ar Kα excitation energy (K-edge) will fluoresce an Ar Kα X-ray which escapes from the detector since it is not strongly absorbed by the Ar detector gas. The Ar excitation energy and X-rays which produce an Ar escape peak are:

  • Ar excitation energy is 3.2 keV. X-rays with energies close to and above this energy will be strongly absorbed and fluoresce an Ar Kα X-ray.
  • K-lines producing an Ar escape peak: K (3.313 keV), Ca (3.691 keV), Sc (4.09 keV), Ti (4.51 keV), Cr (5.41 keV) (and on a Cameca, Mn, Fe, etc.).
  • L-lines producing an Ar escape peak: In (3.287 keV), Sn (3.444 keV), Sb (3.605 keV), Te (3.769 keV), etc.
  • No M alpha lines have sufficient energy to produce an Ar escape peak.
  • The escape peak for K Kα is at lowest voltage with progressive movement of the escape peak into the main pulse distribution for higher energy X-rays.
  • This means that a higher baseline is required for elements producing an escape peak at low voltage.
  • On a JEOL flow-proportional spectrometer using a PET crystal, Cr Kα is the most energetic K-line that can be practically accessed, and Sm Lα is the most energetic L-line that can be accessed.
  • On Cameca instruments one can observe the Ar escape peaks as discussed here but also for Mn, Fe, etc. since all spectrometers use P-10 flow counters.

Note that for X-rays having energies less than the Ar excitation energy, no escape peak is generated (e.g., for all K-lines up to Ar Kα, all L-lines up to Cd Lα, and all Mα lines). Also, for X-rays with energies significantly above the Ar excitation energy the escape peak is buried in the main pulse distribution and is not observed.

The best policy is to understand the pulse distribution and to either include or exclude the escape peak by proper selection of the PHA baseline.

Example SCA Scans Using P-10 Flow Counter on a JEOL 8200 with Ar Escape Peaks

The following plots show SCA plots for the K-lines of K, Ca, Sc, Ti, V, and Cr, and were obtained by placing the spectrometer at the peak position, acquiring a bias scan in order to determine the bias necessary to obtain a ~4V pulse on a SCA scan, then acquiring an SCA scan. In each case a typical microprobe standard was used (microcline for K, wollastonite for Ca, etc.). All data were acquired at 15kV and varying probe currents as required for necessary count rates. The SCA scans were obtained at 0.1 V steps with 5 sec per step.

All SCA scans superimposed to show progressive shift in Ar escape peak position with increasing X-ray energy.

SCA scan for K (JEOL P-10 counter, PET crystal, microcline)

SCA scan for Ca (JEOL P-10 counter, PET crystal, wollastonite)

SCA scan for Sc (JEOL P-10 counter, PET crystal, Sc metal)

SCA scan for Ti (JEOL P-10 counter, PET crystal, TiO2)

SCA scan for V (JEOL P-10 counter, PET crystal, V metal)

SCA scan for Cr (JEOL P-10 counter, PET crystal, Cr2O3)

SCA Scans on JEOL Sealed Counters -- Xe Escape Peaks

JEOL instruments have sealed Xe detectors that are used with PET and LiF crystals for the analysis of more energetic X-rays. In the same fashion as for Ar escape peaks, X-rays that have an energy just above the Xe Lα excitation energy (LIII-edge) will fluoresce a Xe Lα X-ray which escapes from the detector since it is not absorbed by the Xe detector gas. Note, however, that the size of the Xe escape peak is smaller than the Ar escape peak due to the higher Z of Xe and inherent stronger attenuation of X-rays as a result. The Xe excitation energy and X-rays which produce an Xe escape peak are:

  • Xe Lα excitation energy is 4.782 keV, this is the Xe LIII edge energy. X-rays with energies close to and above this energy will be strongly absorbed and fluoresce an Xe Lα X-ray.
  • K-lines producing a Xe escape peak: V (4.952 keV), Cr (5.414 keV), Mn (5.898 keV), Fe (6.403 keV), etc.
  • L-lines producing a Xe escape peak: Ce (4.840 keV), Pr (5.034 keV), Nd (5.230 keV), (Pm), Sm (5.636 keV), etc.
  • No M alpha lines have sufficient energy to produce a Xe escape peak.
  • The escape peak for V Kα is at lowest voltage with progressive movement of the escape peak into the main pulse distribution for higher energy X-rays.
  • This means that a higher baseline is required for elements producing an escape peak at low voltage.

Again, the best policy is to understand the pulse distribution and to either include or exclude the escape peak by proper selection of the PHA baseline.

Example SCA Scans Using Sealed Xe Counter on a JEOL 8200 with Xe Escape Peaks

The following plots show SCA scans for K-lines of V, Cr, Mn, Fe, Co, Ni, Ge, and L-line of Pb. They were obtained by placing the spectrometer at the peak position, acquiring a bias scan in order to determine the bias necessary to obtain a ~4V pulse on a SCA scan, then acquiring an SCA scan. In each case a typical microprobe standard was used (V metal for V, Cr2O3 for Cr, etc.). All data were acquired at 15kV and varying probe currents as required for necessary count rates. The SCA scans were obtained at 0.1 V steps with 5 sec per step.

All SCA scans superimposed to show progressive shift in Xe escape peak position with increasing X-ray energy.

SCA scan for V (JEOL Xe counter, LiF crystal, V metal)

SCA scan for Cr (JEOL Xe counter, LiF crystal, Cr2O3)

SCA scan for Mn (JEOL Xe counter, LiF crystal, Mn metal)

SCA scan for Fe (JEOL Xe counter, LiF crystal, Hematite)

SCA scan for Co (JEOL Xe counter, LiF crystal, Co metal)

SCA scan for Ni (JEOL Xe counter, LiF crystal, Ni olivine)

SCA scan for Cu (JEOL Xe counter, LiF crystal, NIST SRM 478 Brass )

SCA scan for Zn (JEOL Xe counter, LiF crystal, Zn metal)

SCA scan for Ge (JEOL Xe counter, LiF crystal, Ge metal)

SCA scan for Pb Lα (JEOL Xe counter, LiF crystal, PbS)



Return to Microanalysis Facility


Page last updated on: Copyright © Dan Kremser, HTML Paul Carpenter