EPSC Microanalysis Facility
Microprobe: JEOL 8200 Imaging System

Procedures for Routine Tasks on WU JXA-8200 Electron Microprobe

JEOL JXA-8200 Imaging

The Jeol 8200 imaging system is computer controlled using the OS-9 computer system and is most conveniently run using the OS-9 monitor.

The operation of the Jeol imaging system is covered in the JEOL Control and Display document, paper copy only in the lab.

The following guide can be used to perform routine imaging tasks, but requires that you understand where the menus and buttons are located as described in the Jeol OS-9 Digital Imaging System manual.

Setting Up for Digital Imaging and Photography

The following sequence should be used or verified before you can acquire digital images.

  1. Samples already in the microprobe and vacuum has recovered so that electron beam is stable. Vacuum level should be ~ 1x10-3 Pascal in sample chamber.
  2. High voltage turned on, confirmed from green color square in uppler left corner of OS-9 monitor or pink colored button for HV in EOS Monitor windowkl.
  3. Filament saturated (Experienced users only, use EOS monitor to set saturation).
  4. Column aligned if necessary.
  5. Probe current for an imaging run is set using EOS Monitor CL lens button.
    There are two probe current ranges to consider, trading or balancing spatial resolution vs. signal quality:
    • Increased spatial resolution at decreased backscattered-electron sensitivity. Use 5-10 nA for high magnification and good resolution requirements.
    • Increased BSE sensitivity at decreased spatial resolution. Use 25nA for compatibility with analytical runs that use 25nA probe current. Use 10-40nA for good backscattered-electron signal (but note beam sensitive materials require lower probe current).
  6. Choose a feature that has both x and y components of contrast, i.e., a set of cracks or pits with detail.
  7. Set optical focus using joystick arrows and reflected light monitor with reflected light turned on.
  8. Focus and stigmation checked at selected probe current

Vacuum levels during and after sample exchange:

Abnormal specimen chamber vacuum levels:

Electron gun vacuum levels:

JEOL 8200 Column Front Panel

The front panel of the microprobe column has the hardware interface for the vacuum system:

JEOL 8200 Column Front Panel: Vacuum System and Pump-down Controls

The front panel has three sections and a fourth section on the side panel of the column. The left panel has a diagram of the vacuum logic system which shows the valves that are open which are illuminated green. A valve that is closed is not illuminated.
Below the vacuum valve diagram are two push buttons that are used for venting the gun (Gun Vent) and the entire microprobe column (Spec Vent). The gun vent push button is used to vent the electron gun when the filament is to be changed, and only the gun is vented. The spec vent push button is used to vent the entire microprobe for maintenance purposes.

These vent buttons are not used for sample exchange or any of the routine procedures.

JEOL 8200 Column Front Panel: Defect Light Panel, Filament Mode, and OPE Pwr Switch

The center portion of the panel has defect lights that are illuminated when a defect condition is detected for:

Below the defect panel are two toggle switches. The LB6/W switch is used to select which filament type is in the electron gun, LaB6 or W (tungsten); we normally use a tungsten filament but run the instrument in LaB6 mode. When in LaB6 mode the SIP ion pump is turned on and the vacuum system runs in a different configuration with the gun isolated from the turbo pump system. The vacuum display will be different than when running in the W mode.

The OPE Pwr switch supplies power to all of the 8200 computer boards in the column and also turns on the high voltage supply. It must be on for the microprobe to be able to generate an electron beam. There are times when the computer loses communication and recycling of the OPE Power switch is necessary.

JEOL 8200 Column Front Panel: Transmitted Light Periscope and Reset Switches for SC and OPE

The right hand side of the front panel has a switch for the transmitted light periscope to be inserted or retracted. Our microprobe does not have this attachment.

The SC Reset and OPE Reset switches are in principle used to perform a hardware reset of the system command (SC) board and the operations power (OPE) systems. Generally, pressing these will guarantee that the OPE Power will have to be cycled anyway, so their use is not recommended, or rather, their use does not result in a solution to the computer lockup problem.

JEOL 8200 Column Side Panel: Accelerating Voltage Hard Switch, HV Ready Lamp, Master Power Switch

The side panel has the Accel Voltage push button located at the top. This push button is a hardware switch for the high voltage power and must be pushed in and illuminated in order for the microprobe system to turn on the high voltage supply via the software. Below this is the HV Ready indicator which is illuminated only when the microprobe has pumped to high vacuum and furthermore to the minimum vacuum level required for operation of the high voltage supply. The following is a summary of the requirements for producing an electron beam with the microprobe:

Below the HV Ready lamp is the Over Emission indicator which generally indicates that too high of a gun bias has been used and that a dendrite or whisker has grown from the filament to thw whenelt (grid cap). This means that the gun has to be cleaned.

The Power switch is at the bottom of the panel and is used to turn the entire microprobe system on and off. The switch on our system is a three position switch with off in the lower position, on in the middle position, and start in the upper position. This switch is used for the following purposes:

JEOL 8200 Vacuum System: Sequencing Details

This section covers the details of valve sequencing for the JEOL 8200 during pump down. The valve configuration can be viewed in two places, the front panel display on the column, and the vacuum application (with "pressure" check box selected) in the JEOL software. The Probe for Windows Vacuum application can also display the vacuum levels without the valve information.

This is the normal high vacuum mode for the WU 8200 with the filament mode set to LaB6 mode, with the SIP ion pump turned on.

This display indicates that the instrument is at high vacuum and is at "High Voltage Ready" so that the accelerating voltage can be (or currently is) turned on.

The valves shown on this display are as follows:

The following chart shows the valve sequencing during pump-down:

End of vacuum document.

Return to procedures